Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231149

    Validate with FDA (Live)

    Device Name
    qXR-CTR
    Date Cleared
    2023-09-22

    (154 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.

    qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.

    The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.

    The system is suitable for adults > 22 years of age.

    The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.

    Device Description

    The qXR-CTR is a non-invasive software medical device designed to be installed on the computer with specific system requirements. It is a radiological computer-assisted software system that automatically analyzes DICOM chest X-ray images in PA view and outputs the cardiac diameter, thoracic diameter, and CTR through an artificial intelligence algorithm. The structured report includes a preview of the compressed chest X-ray image with the automatically derived CTR result and annotation line, indicating the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity.

    AI/ML Overview

    Acceptance Criteria and Study Details for qXR-CTR

    1. Table of Acceptance Criteria and Reported Device Performance

    MeasureAcceptance Criteria (Predicate)Reported Device Performance (qXR-CTR)
    Cardiac Diameter RMSE8.81 mm7.55 (6.95, 8.34) mm
    Thoracic Diameter RMSE14.4 mm5.43 (4.95, 6.11) mm
    Cardiac Diameter MAENot Reported5.66 (5.0) mm
    Thoracic Diameter MAENot Reported4.04 (3.63) mm
    CTR0.03 (0.02, 0.03)0.02 (0.02)

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 435 scans.
    • Data Provenance: Retrospectively collected chest X-rays from various parts of the US.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    • Number of Experts: 3
    • Qualifications: ABR (American Board of Radiology) thoracic radiologists with a minimum of 10 years of experience.

    4. Adjudication Method for Test Set

    The adjudication method is not explicitly stated, but with 3 experts establishing ground truth, a consensus method (e.g., majority vote or discussion to agreement) is implied. It's not explicitly stated as 2+1, 3+1, or none.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The document describes a standalone performance evaluation of the qXR-CTR against ground truth established by experts.

    6. Standalone Performance

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The reported performance metrics (RMSE, MAE) are intrinsic to the device's measurements compared to expert-established ground truth.

    7. Type of Ground Truth Used

    Expert consensus. The ground truth was established by three ABR thoracic radiologists.

    8. Sample Size for Training Set

    The sample size for the training set is not provided in the given text.

    9. How Ground Truth for Training Set was Established

    The method for establishing ground truth for the training set is not described in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1