510(k) Data Aggregation

    K Number
    K230365

    Validate with FDA (Live)

    Device Name
    Sonio Detect
    Manufacturer
    Date Cleared
    2023-07-25

    (165 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sonio Detect is intended to analyze fetal ultrasound images and clips using machine learning techniques to automatically detect views, detect anatomical structures within the views and verify quality criteria of the views.
    The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

    Device Description

    Sonio Detect is a Software as a Service SaaS solution that aims at helping sonographers, OB/GYNs, MFMs and Fetal surgeons (all three designated as healthcare professionals i.e. HCP) to perform their routine fetal ultrasound examinations in real-time. Sonio Detect can be used by Healthcare Professionals HCPs during fetal ultrasound exams for Trimester 1, Trimester 2 and Trimester 3 of the fetus (Gestational Age: from 11 weeks to 37 weeks). The software is intended to assist HCPs in assuring during and after their examination that the examination is complete and all images were collected according to their protocol.
    Sonio Detect receives fetal ultrasound images and clips from the ultrasound machine, that are submitted through the edge software by the performing healthcare professional, in real-time and performs the following:

    • . Automatically detect views;
    • Automatically detect anatomical structures within the supported views; .
    • Automatically verify quality criteria of the supported views by checking whether they . conform to standardized quality criteria.
      Quality criteria are related to:
    • the presence or absence of an anatomical structure; ●
    • the zoom level for some views.
      Sonio Detect then automatically associates the image to its detected view. It also highlights in yellow the view and/or the corresponding quality criteria if there are unverified items : quality criteria not verified or view not detected.
      The end user can interact with the software to override the Sonio Detect's outputs (reassign the image to another view or unassign it or assign it if it was not assigned, change the status of a quality criteria from verified to unverified or from unverified to verified). The user has the ability to review and edit/override the matching at any time during or at the end of the exam.
    AI/ML Overview

    Sonio Detect Acceptance Criteria and Study Details

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for Sonio Detect are implicitly defined by the reported performance metrics, which the FDA has deemed sufficient for substantial equivalence. The reported performance is presented as sensitivities, specificities, and proportions of correctly read annotations.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    3D Fetal Ultrasound Image Detection SensitivityHigh sensitivity0.980 (95% Wilson's CI: 0.930, 0.994)
    Doppler Fetal Ultrasound Image Detection SensitivityHigh sensitivity0.963 (95% Wilson's CI: 0.908, 0.985)
    Fetal Ultrasound Views Detection Proportion CorrectHigh proportion0.923 (95% Wilson's CI: 0.905, 0.938)
    T1 Fetal Ultrasound Views Detection SensitivityHigh sensitivity0.942 (Point estimate)
    T2/T3 Fetal Ultrasound Views Detection SensitivityHigh sensitivity0.919 (Point estimate)
    T2/T3 Fetal Brain Anatomical Structure Detection SensitivityHigh sensitivity0.857 (Point estimate)
    T2/T3 Fetal Brain Anatomical Structure Detection SpecificityHigh specificity0.963 (Point estimate)
    T2/T3 Fetal Heart Anatomical Structure Detection SensitivityHigh sensitivity0.900 (Point estimate)
    T2/T3 Fetal Heart Anatomical Structure Detection SpecificityHigh specificity0.982 (Point estimate)
    Zoom Level Verification Sensitivity (Brain Views)High sensitivity0.952 (95% Wilson's CI: 0.909-0.976)
    Zoom Level Verification Specificity (Brain Views)High specificity0.906 (95% Wilson's CI: 0.758-0.968)

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 17,885 fetal ultrasound images.
    • Data Provenance: The data was collected from 7 clinical sites in the United States, France, and Germany. This indicates a multi-national dataset. The data was retrospective as it was "independent of the data used during model development (training/fine tuning/internal validation) and establishment of device operating points."

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    The document does not explicitly state the number of experts used or their specific qualifications (e.g., years of experience) for establishing the ground truth of the test set. However, it indicates that the device automatically detects fetal ultrasound views "through reading of annotations on images." This implies that human experts (presumably sonographers, OB/GYNs, MFMs, or Fetal surgeons, as these are the intended users) provided the initial annotations that served as the ground truth.

    4. Adjudication Method (Test Set)

    The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for establishing the ground truth of the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed. The document explicitly states: "Clinical Study: Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." Therefore, there is no reported effect size of human readers improving with AI vs. without AI assistance.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The document states: "Sonio conducted a standalone performance testing on a dataset of 17885 fetal ultrasound images..." This indicates the algorithm's performance was evaluated without human intervention in the loop during the assessment of the test set.

    7. Type of Ground Truth Used (Test Set)

    The ground truth for the test set was established through "reading of annotations on images." This suggests the ground truth was based on expert annotations or labeling of the ultrasound images, likely by the qualified healthcare professionals who generated the initial data.

    8. Sample Size for Training Set

    The document does not explicitly state the sample size for the training set. It refers to "data used during model development (training/fine tuning/internal validation)" but does not provide a specific number of images or cases for this phase.

    9. How Ground Truth for Training Set was Established

    The method for establishing the ground truth for the training set is not explicitly detailed. However, given that the test set's ground truth was based on annotations, it is highly probable that the training set's ground truth was established through a similar process of expert annotation or labeling of the fetal ultrasound images and clips.

    Ask a Question

    Ask a specific question about this device