510(k) Data Aggregation

    K Number
    K230051

    Validate with FDA (Live)

    Date Cleared
    2023-02-03

    (28 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 14F iSLEEVE™ Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of cardiovascular devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.5 mm diameter.

    Device Description

    The 14F iSLEEVE Introducer Set is a sterile, single-use introducer catheter that provides percutaneous access to the femoral artery for introduction and removal of cardiovascular devices into the vascular system. The 14F iSLEEVE Introducer Set is composed of a dilator and an introducer sheath with a three-way stopcock.

    The distal end of the introducer sheath is passively expandable which allows for transient sheath expansion when the compatible valve system is introduced through it. Once expanded, this region of the introducer sheath is radially compliant, which allows it to expand and collapse as needed during device delivery and removal. This reduces the time the access vessel is expanded. A hydrophilic coating is applied to the working end of the iSLEEVE introducer sheath which increases the lubricity to aid in delivery when activated.

    A product mandrel and flushing tube are also supplied with the iSLEEVE Introducer Set.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the 14F iSLEEVE Introducer Set. It focuses on demonstrating substantial equivalence to a predicate device, specifically after a modification involving an alternative hydrophilic coating.

    However, the document does not contain the level of detail typically found in a clinical study report or a robust performance study for an AI/ML medical device, which would include the specific acceptance criteria tables, sample sizes for test and training sets, expert qualifications, and adjudication methods for ground truth, and MRMC study details.

    The provided information pertains to the acceptance criteria and study that proves the device meets the acceptance criteria for a medical device with a coating change, not an AI/ML algorithm.

    Based on the provided text, here is an attempt to address your questions, while highlighting the limitations due to the nature of the document:

    1. A table of acceptance criteria and the reported device performance

    The document mentions that "Testing demonstrated that the modified 14F iSLEEVE Introducer Set met all previously established acceptance criteria." However, it does not provide a specific table of acceptance criteria with corresponding performance numbers. It lists the types of non-clinical tests performed:

    Test CategoryAcceptance Criteria (Implied)Reported Performance (Implied)
    Coating Material and LengthMet specificationsMet specifications
    Coating Lubricity and Durability (L&D)Met specificationsMet specifications
    Introducer System Advancement ForceMet specificationsMet specifications
    Maximum Junction ODMet specificationsMet specifications
    Particulate EvaluationMet specificationsMet specifications
    Coating IntegrityMet specificationsMet specifications
    Visual AppearanceMet specificationsMet specifications
    Biocompatibility (Cytotoxicity)Met specificationsMet specifications
    Biocompatibility (Irritation)Met specificationsMet specifications
    Biocompatibility (Sensitization)Met specificationsMet specifications
    Biocompatibility (Material Mediated Pyrogenicity)Met specificationsMet specifications
    Biocompatibility (Acute Systemic Toxicity)Met specificationsMet specifications
    Biocompatibility (Hemocompatibility)Met specificationsMet specifications
    Biocompatibility (Genotoxicity)Met specificationsMet specifications
    Animal Study: Performance of Insertion and RemovalFunctionally equivalent to predicateFunctionally equivalent to predicate

    Note: The "Reported Performance (Implied)" column is based on the statement "Testing demonstrated that the modified 14F iSLEEVE Introducer Set met all previously established acceptance criteria" and "Results confirm that the modified device remains biocompatible for its intended use." Specific quantitative results are not provided in this summary.

    2. Sample sizes used for the test set and the data provenance

    • Test Set Sample Size: Not explicitly stated for each non-clinical test. The document refers to "the modified device" meeting criteria. For the animal study, it mentions a "double-blinded animal study" in the "porcine model," implying a sample size of animals, but the exact number of animals or devices tested is not provided.
    • Data Provenance:
      • Non-clinical testing: Performed internally by the manufacturer (Boston Scientific Corporation). The location of the testing is not specified, but it's presumed to be in a controlled laboratory environment.
      • Animal Study: Conducted in a "porcine model." No information on the country of origin or whether it was retrospective or prospective is given, though animal studies for device testing are typically prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Non-clinical tests: Ground truth is established by standard engineering and biocompatibility testing methodologies and validated equipment, not typically by human experts in the same way as medical image analysis.
    • Animal Study: "Physicians were asked to separately evaluate the performance of the insertion and removal of the predicate and alternative hydrophilic coated device."
      • Number of experts: Not specified (e.g., "physicians" could be one or multiple).
      • Qualifications of experts: Only identified as "Physicians." No details on their specialty, experience level, or specific training are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Non-clinical tests: Not applicable in the context of human expert adjudication.
    • Animal Study: The phrase "Physicians were asked to separately evaluate" suggests individual assessments, but it does not specify an adjudication method (e.g., consensus, majority vote) if multiple physicians were involved. It implies a comparative evaluation between the modified and predicate device's performance characteristics during insertion and removal.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was performed. This document describes the testing for a physical medical device (introducer sheath) with a coating change, not an AI/ML algorithm. Therefore, the concept of human readers improving with AI assistance is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This relates to a physical medical device, not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Non-clinical tests: Ground truth is established through physical measurements, chemical analyses, and standardized biological assays, adhering to relevant standards (e.g., ISO 10993-1 for biocompatibility).
    • Animal Study: Ground truth was based on the "functional performance" assessed by "physicians" during the insertion and removal of the devices in the porcine model. This would involve qualitative and potentially some quantitative assessment of ease of use, resistance, and other performance characteristics directly observed during the procedure. It's a form of expert observation/assessment in a simulated use environment.

    8. The sample size for the training set

    • Not applicable. This device is not an AI/ML algorithm. It is a manufactured physical good, and as such, there is no "training set" in the context of machine learning. The design and manufacturing process would involve internal development and verification.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.