    Intended Use

    SKORE Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

    Device Description

    The subject SKORE Male Natural Rubber Latex Condoms are made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane and used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted infections. The subject device in this submission consists of SKORE Smooth, Dotted &Thin Condoms, SKORE Bulbous Condoms, and SKORE Benzocaine Condoms.

    AI/ML Overview

    The provided text describes the 510(k) summary for SKORE Male Natural Rubber Latex Condoms. It details the device, its intended use, comparison to a predicate device, and performance testing data.

    Here's the requested information structured for acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedAcceptance Criteria (Standard)Reported Device Performance
    Biocompatibility:ISO 10993-1:2017 (Biological Evaluation of Medical Devices)Demonstrates that the subject devices are biocompatible.
    - Acute Systemic ToxicityISO 10993-11:2017Met specifications.
    - In vitro CytotoxicityISO 10993-5:2009Met specifications.
    - Skin SensitizationISO 10993-10:2021Met specifications.
    - Vaginal Irritation TestISO 10993-23:2021Met specifications.
    Mechanical Performance TestingISO 4074:2015 (Natural rubber latex male condoms - Requirements and test methods) & ASTM D3492-16 (Standard Specification for Rubber Contraceptives)The subject devices meet the standard specifications.
    Air Burst Test Pressure> 1.0 kPa (as per predicate device comparison)> 1.0 kPa (as per predicate device comparison)
    Air Burst Test Volume18 dm³ (as per predicate device comparison)18 dm³ (as per predicate device comparison)
    Shelf Life5 Years (per ISO 4074:2015 & 21 CFR 801.435)Supports a shelf-life of five years as per the requirements.

    2. Sample size used for the test set and the data provenance

    • Sample Size:
      • For Mechanical Performance Testing: "Three lots of subject devices, SKORE Male Natural Rubber Latex Condoms" were tested.
      • For Biocompatibility and Shelf Life: The sample sizes are not explicitly stated, but the tests were performed on "SKORE Male Natural Rubber Latex Condoms."
    • Data Provenance: Not explicitly stated (e.g., country of origin). The testing seems to be conducted by the manufacturer, TTK Healthcare Ltd., which is based in India. The studies are prospective in nature, as they are conducted to demonstrate compliance of the device before market entry.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to the evaluation of condoms. The ground truth for condom performance is established through adherence to recognized international and national standards (e.g., ISO 4074, ASTM D3492) and laboratory testing, rather than expert consensus on interpretive data.

    4. Adjudication method for the test set

    Not applicable. The tests are objective and quantitative, based on physical and chemical properties measured against established standards. There is no subjective interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical product (condom), not an AI-based diagnostic or assistive technology for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device. Performance is determined through physical and biological testing.

    7. The type of ground truth used

    The ground truth used for this device is based on established international and national standards (ISO 4074:2015, ASTM D3492-16, ISO 10993 series, 21 CFR 801.435) for material properties (natural rubber latex), physical integrity (air burst pressure/volume), dimensions, biocompatibility, and shelf-life.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.