Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222897

    Validate with FDA (Live)

    Device Name
    Enlight 2100
    Manufacturer
    Date Cleared
    2023-03-07

    (165 days)

    Product Code
    Regulation Number
    868.1505
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's thorax.

    ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring.

    It is intended for mechanically ventilated adult and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 - 134 cm.

    ENLIGHT 2100 does not measure regional ventilation of the lungs

    Device Description

    ENLIGHT 2100 is a Ventilatory electrical impedance tomograph that uses several electrodes (usually between 16 and 32) placed around the patient's thorax to assess regional impedance variation in a lung slice (tomography). It provides a relative measurement, so it only provides information on variations in local impedance. ENLIGHT 2100 estimates Local Impedance Variation, occurring in a cross section of the thorax during a respiratory cycle, and which are linearly related to Variations in Regional Air Content within the lung.

    AI/ML Overview

    The ENLIGHT 2100 is a ventilatory electrical impedance tomograph that also provides respiratory parameters based on spirometric monitoring. It is intended for mechanically ventilated adult and pediatric patients in a hospital setting. The device provides information on impedance variation from a cross-section of a patient's thorax as an adjunctive tool to support the user's assessment of variations in regional air content. It does not measure regional ventilation of the lungs. The device relies on a primary predicate (ENLIGHT 2100 - K211135) for its Electrical Impedance Tomography (EIT) data and a secondary predicate (Philips NM3 Respiratory Profile Monitor with VentAssist – K103578) for its spirometric monitoring capabilities. The data provided focuses on non-clinical/bench testing to demonstrate substantial equivalence to these predicates.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ENLIGHT 2100 are based on its performance characteristics in comparison to its predicate devices for both EIT parameters and spirometric parameters. The study conducted was non-clinical bench testing.

    Table 1: Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (from Subject Device's "Performance Characteristics")Reported Device Performance (from Subject Device's "Performance Characteristics" and "Explanation of Differences")
    EIT Parameters
    Distribution RatiosRange: 0 – 100%Anterior, Posterior, Left, Right Distribution Ratio: Uncertainty of +/- 10 p.p. (Same as primary predicate). The explanation states the predicate didn't present the numeric parameter, but because the hardware and EIT algorithm are the same, the performance is equivalent.
    Tidal Variation Z (TVz)Range: 20% to 500%Tidal Impedance Variation (TVz): Uncertainty of +/- 10% of reading. (Same as primary predicate). Calculates and displays: a) maximum impedance variation for the respiratory cycle in which the reference is positioned, b) the maximum impedance variation for the respiratory cycle in which the cursor is positioned, and c) the relationship between the maximum impedance variation of these two timepoints.
    Tidal Variation RateAdult: 5 to 50 bpm, Pediatric: 10 to 140 bpmTidal Variation Rate: ±2.0 bpm if ≤ 60 bpm, ±5.0 bpm if > 60 bpm. (Same as primary predicate). The device calculates and displays the Tidal Variation Rate, considering the last minute (number of oscillations identified in the last minute).
    Spirometric Parameters----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Tidal VolumeRange: 40 to 2500 mlAccuracy: Max Error is less than or equal to 11.71mL, Max Relative Error is less than or equal to 4.97%. Comparison to NM3: Absolute accuracy related to NM3 is less than 6.27mL, and relative accuracy related to NM3 is less than 4.33%.
    Respiratory RateRange: 5 to 150 breath/minAccuracy: Max Error is less than or equal to 0.70 bpm. Comparison to NM3: Absolute accuracy related to NM3 is less than 0.39 bpm.
    Positive End Expiratory Pressure (PEEP)Range: 1.0 - 50.0 cmH2OAccuracy: Max Error is less than or equal to 1.0cmH2O, Max Relative Error is less than or equal to 2.47%. Comparison to NM3: Absolute accuracy related to NM3 is less than 0.73 cmH2O.
    Peak Inspiratory Pressure (PIP)Range: 1.0-120.0 cmH2OAccuracy: Max Error is less than or equal to 0.44 cmH2O. Comparison to NM3: Relative accuracy related to NM3 is less than 4.88%.
    ResistanceRange: 5 - 40 cmH2O/L/sAccuracy: Bias: 0 cmH2O/L/s, Std Dev: 3 cmH2O/L/s. Comparison to NM3: Absolute mean accuracy related to NM3 is less than 6 cmH2O/L/s.
    ComplianceRange: 3 - 80 ml/cmH2OAccuracy: Bias: -1 mL/cmH2O, Std. Dev: 5 mL/cmH20. Comparison to NM3: Absolute mean accuracy related to NM3 is less than 6.97 mL/cmH2O.
    Plateau PressureRange: 10.0-90.0 cmH2OAccuracy: Bias: 0.1 cmH2O, Std Dev: 1.1 cmH2O. Comparison to NM3: Absolute mean accuracy related to NM3 is less than 1.47 cmH2O.

    The "Explanation of Differences" column in the provided tables typically serves as the primary source for the reported device performance and the proof that it meets the acceptance criteria (i.e., demonstrating substantial equivalence to the predicates based on the non-clinical testing). For the EIT parameters, the performance is reported as "Same" as the predicate, with the rationale that the hardware and EIT algorithm are identical. For the spirometric parameters, specific accuracy metrics are provided for the subject device and its accuracy relative to the secondary predicate (NM3).

    2. Sample size used for the test set and data provenance

    The document explicitly states that "Bench Testing - We have performance tests to check the automatic calculation of the parameters shown at the Trend Screen." and lists "Non-clinical testing" and "Bench Test" as the method. Therefore, the data provenance is bench test data. No information is provided regarding the specific sample size (e.g., number of test points, simulated cases, or repetitions) used for this bench testing. The data does not specify the country of origin, but given the sponsor's location (Brazil) and the nature of bench testing, it would likely be laboratory/engineering data. It is inherently prospective as it involves controlled testing to confirm performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The study described is non-clinical bench testing. It involves evaluating the device's numerical calculations and performance against specified ranges and accuracy criteria, likely using simulated physiological signals or validated reference equipment. Therefore, no human experts were used to establish ground truth in the way they would be for image interpretation or diagnosis. The ground truth for this type of testing is established by the specifications of the signals generated or the reference standards of the testing equipment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Given that this was a non-clinical bench study focused on numerical calculation accuracy and signal acquisition performance, no human adjudication method was employed. The "ground truth" was inherently defined by the test setup and reference measurements, not by human interpretation or consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The device (ENLIGHT 2100) functions as a sensor and data display unit for physiological parameters, not an AI-assisted diagnostic tool that would directly assist human readers in interpreting complex medical images or data where a reader study would be applicable. The document describes it as an "adjunctive tool to other clinical information," implying it provides data for clinicians to interpret, but not in a way that necessitates an AI-assistance reader study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone (algorithm only) performance evaluation was implicitly done through the bench testing. The document states, "We have performance tests to check the automatic calculation of the parameters shown at the Trend Screen." This testing evaluates the device's ability to accurately measure and calculate the specified EIT and spirometric parameters within defined ranges and accuracies, independent of human interaction or interpretation beyond setting up the test and recording results.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for this non-clinical bench testing was established through engineered inputs/simulations and/or comparisons to a validated reference standard. For example, for spirometric parameters like Tidal Volume or Respiratory Rate, the system would be fed precisely controlled and measured airflow/pressure signals, and the device's output would be compared to the known input values of these signals. For EIT parameters, the ground truth would similarly come from controlled electrical impedance variations generated under laboratory conditions.

    8. The sample size for the training set

    The document describes non-clinical bench testing for the purpose of demonstrating substantial equivalence. It does not mention any machine learning or AI components that would require a separate training set. Therefore, information regarding a training set sample size is not applicable or provided in this context. The device's algorithms are likely based on established physiological and electrical impedance principles, not trained on a large dataset of patient measurements requiring specific "training set" ground truth establishment.

    9. How the ground truth for the training set was established

    As there is no mention of a training set or machine learning/AI model training, the question of how its ground truth was established is not applicable based on the provided information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1