Found 1 results

510(k) Data Aggregation

    K Number
    K222798

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-16

    (91 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rejoni Intrauterine Catheter is for the delivery of contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes.

    When used for HSG. the Rejoni Intrauterine Catheter can be used for evaluation of tubal patency.

    The Rejoni Intrauterine Catheter is used to access the uterine cavity for the delivery of saline for Sonohysterography (SHG).

    Device Description

    The Rejoni Intrauterine Catheter is a sterile, single lumen uterine catheter that facilitates access to the uterus to deliver fluids during Hysterosalpingography (HSG) or Sonohysterography (SHG). The Rejoni Intrauterine Catheter consists of an inner shaft and an additional support sheath over the proximal portion of the catheter shaft to provide additional stiffness. A repositionable stopper ("acom") is connected to the support sheath, which is placed on the catheter shaft, and can be repositioned from 0 cm up to 10 cm from the distal tip of the catheter. The stopper maintains placement of the catheter against the external cervical os during the procedure, and a luer lock-style adapter on the proximal end is available for compatibility with syringes. The distal end contains a straight through hole for delivery of fluids during the procedure.

    AI/ML Overview

    The Rejoni Intrauterine Catheter is a sterile, single-lumen uterine catheter intended for delivering contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes. It can also be used for evaluation of tubal patency during HSG.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategorySpecific Test / CriterionAcceptance CriteriaReported Device Performance
    Mechanical TestingPerformance - MechanicalIn accordance with device performance specificationsPass
    Performance - FunctionalIn accordance with device performance specificationsPass
    Dimensional VerificationMet design specificationsPass
    Visual AssessmentMet design specificationsPass
    Tensile Testing (luer to inner shaft; acorn to outer shaft)Met design specificationsPass
    Kink TestingMet design specificationsPass
    Leak TestingMet design specificationsPass
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Non-cytotoxicPass
    Sensitization (ISO 10993-10:2021)Non-sensitizingPass
    Intracutaneous Reactivity (ISO 10993-10:2021)Non-irritatingPass
    SterilitySterility Assurance Level (SAL)10$^{-6}$10$^{-6}$ (validated)
    Sterilization Process ValidationConforms to ISO 11137-1, -2, -3Validated
    Shelf LifeDevice performance over shelf lifeMaintains performance as per mechanical testingPass
    Sterile barrier integrity over shelf lifeMaintains sterile barrierPass
    Package integrity after accelerated aging (ASTM 1980-16)No gross leaks, adequate seal strength, visual integrityPass
    Package integrity after simulated shipping (ISTA-3A:2018, ASTM D4169-16)No gross leaks, adequate seal strength, visual integrityPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test (e.g., how many devices were used for tensile testing, biocompatibility, etc.). However, it indicates these were pre-clinical tests (lab-based) rather than clinical studies with patient data. Therefore, data provenance such as country of origin of data or retrospective/prospective characteristics is not applicable in this context.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The described studies are pre-clinical performance and safety tests (mechanical, biocompatibility, sterility, shelf life) rather than studies that require expert-established ground truth on diagnostic accuracy, for example. The "ground truth" for these tests is defined by established industry standards (ISO, ASTM) and the device's own design specifications.

    4. Adjudication Method:

    Not applicable. As noted above, these are pre-clinical tests following defined protocols and standards, not studies requiring expert adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The device is a physical catheter, not an AI-assisted diagnostic tool requiring evaluation of human reader improvement.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    No, a standalone (algorithm only) study was not done. This device is a manual instrument and does not involve an algorithm.

    7. Type of Ground Truth Used:

    The ground truth for these tests is implicitly or explicitly defined by:

    • Design Specifications: For mechanical and functional performance, the device's own engineering and design requirements serve as the ground truth.
    • International Standards (ISO, ASTM): For biocompatibility, sterility, and packaging/shelf life, the ground truth is established by the specified ISO and ASTM standards and their acceptance criteria (e.g., non-cytotoxic, SAL of 10$^{-6}$).

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.