510(k) Data Aggregation

    K Number
    K222733

    Validate with FDA (Live)

    Date Cleared
    2022-12-11

    (93 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
    Carmustine (BCNU)3.3 mg/ml (3,300 ppm)23.7 Minutes
    Cisplatin1.0 mg/ml(1,000 ppm)240 Minutes
    Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)240 Minutes
    Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)240 Minutes
    Doxorubicin HCl2.0 mg/ml(2,000 ppm)240 Minutes
    Etoposide20.0 mg/ml(20,000 ppm)240 Minutes
    Fluorouracil50.0 mg/ml(50,000 ppm)240 Minutes
    Methotrexate25 mg/ml(25,000 ppm)240 Minutes
    Mitomycin C0.5 mg/ml(500 ppm)240 Minutes
    Paclitaxel6.0 mg/ml(6,000 ppm)240 Minutes
    Thio Tepa10.0 mg/ml(10,000 ppm)45.7 Minutes
    Vincristine Sulfate1.0 mg/ml(1,000 ppm)240 Minutes

    Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.7 Minutes Thio-Tepa 10.0 mg/ml 45.7 Minutes

    Caution: Testing showed an average breakthrough time of 45.7 minutes with Thio-Tepa

    WARNING: Do not use with Carmustine

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered non-sterile and in five sizes: extra-small, small, medium, large, and extra-large.

    AI/ML Overview

    Below is a summary of the acceptance criteria and the study conducted for the Disposable Nitrile Examination Glove, as detailed in the provided FDA 510(k) submission.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength (mm): XS/S: ≥220; M/L/XL: ≥230Width (mm): XS: 70±10; S: 80±10; M: 95±10; L: 110±10; XL: 120±10Thickness (mm): Finger: ≥0.05, Palm: ≥0.05Length: >230 mm for all sizesWidth: XS: 75-80; S: 85-87; M: 95-98; L: 105-107; XL: 115-117Thickness (mm): XS Finger: 0.100.11, Palm: 0.060.10; S Finger: 0.090.12, Palm: 0.070.09; M Finger: 0.080.12, Palm: 0.070.09; L Finger: 0.090.12, Palm: 0.060.09; XL Finger: 0.110.12, Palm: 0.060.09. Result: Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.5XS: 0/125 leaks; S: 1/125 leaks; M: 0/125 leaks; L: 2/125 leaks; XL: 2/125 leaks Result: Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mgXS: 0.08 mg; S: 0.12 mg; M: 0.09 mg; L: 0.10 mg; XL: 0.11 mg. Result: Pass
    ASTM D412 (Physical Properties)Physical properties (Tensile Strength & Elongation)Before Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥500%After Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥400%Before Aging: Tensile Strength: XS:1821, S:1825, M:1826, L:1624, XL:1625; Ultimate Elongation: XS: 520-536, S:518517, M:515575, L:529561, XL:523578 After Aging: Tensile Strength: XS:1521, S:1621, M:1524, L:1623, XL:1624; Ultimate Elongation: XS:500525, S:501522, M:501550, L:501551, XL:503~569. Result: Pass for all
    ASTM D6978Chemotherapy Drugs Permeation (Minimum Breakthrough Time)Specified minimum breakthrough times for each chemotherapy drug (implied from the predicate device's performance, but the standard does not state a universal minimum)Carmustine (BCNU): 23.7 MinutesCisplatin: > 240 MinutesCyclophosphamide (Cytoxan): > 240 MinutesDacarbazine (DTIC): > 240 MinutesDoxorubicin HCl: > 240 MinutesEtoposide: > 240 MinutesFluorouracil: > 240 MinutesMethotrexate: > 240 MinutesMitomycin C: > 240 MinutesPaclitaxel: > 240 MinutesThio Tepa: 45.7 MinutesVincristine Sulfate: > 240 Minutes
    ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells. Result: Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant. Result: Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. Result: Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document provides sample sizes for specific tests:

    • ASTM D5151 (Watertightness): 125 gloves per size (XS, S, M, L, XL) were tested.
    • The specific sample sizes for other tests like physical dimensions (ASTM D6319, ASTM D412), powder content (ASTM D6124), chemotherapy drug permeation (ASTM D6978), and biocompatibility (ISO 10993-5, ISO 10993-10) are not explicitly stated in numerical terms (e.g., "n=X"), but are implied to be sufficient to meet the requirements of the respective ASTM and ISO standards.
    • Data Provenance: The studies were conducted by the device manufacturer, Inner Mongolia Boming Medical Supplies Co., Ltd., which is located in China. The data would be considered prospective as it involves new testing on the subject device to demonstrate its performance against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable to this device. The "ground truth" for the performance of examination gloves is established by objective, standardized laboratory tests (e.g., ASTM and ISO methods) rather than expert clinical judgment or consensus. These are physical and chemical properties measured according to specific protocols.

    4. Adjudication Method for the Test Set:

    This is not applicable to this device. As mentioned above, the evaluation relies on objective measurements from standardized test methods, not subjective assessments requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human readers interpret medical images or other complex data, and their performance is compared with and without AI assistance. Examination gloves are a Class I device with performance assessed via bench testing, not human interpretation.

    6. Standalone (Algorithm Only) Performance:

    No, a standalone (algorithm only) performance study was not done. This device is a physical product (examination glove), not an AI algorithm or software. Its performance is evaluated through material and physical property testing, not algorithmic output.

    7. Type of Ground Truth Used:

    The ground truth used in these studies is based on objective, quantitative measurements obtained through validated and standardized laboratory test methods as defined by:

    • ASTM International (formerly American Society for Testing and Materials) standards: ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, and ASTM D6978.
    • ISO (International Organization for Standardization) standards: ISO 10993-5 and ISO 10993-10.
      These standards define the methods for measuring physical properties, chemical resistance, and biocompatibility, and the "ground truth" is the result of these measurements compared against the specified acceptance criteria within the standard.

    8. Sample Size for the Training Set:

    This information is not applicable. The device is not an AI/ML algorithm that requires a "training set" of data. The manufacturing process of examination gloves is based on established engineering principles and quality control, rather than machine learning models.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no "training set" for this device.