Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222358

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-20

    (138 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    0 - 21
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators:

    As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

    INDIGO Aspiration Tubing:

    As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump:

    The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

    Device Description

    The INDIGO® Aspiration System is comprised of the several devices:

    • INDIGO Aspiration Catheter
    • Penumbra Aspiration Pump
    • INDIGO Aspiration Pump Canister
    • INDIGO Aspiration Tubing
    • INDIGO Separator

    The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, and introducer. The INDIGO Separator may be provided with an introducer and torque device.

    AI/ML Overview

    The provided text is a compilation of FDA documents related to the 510(k) clearance for the Penumbra, Inc. Indigo® Aspiration System – Lightning® Flash. It describes the device, its indications for use, and a comparison to predicate devices, along with summaries of non-clinical data (biocompatibility and bench-top testing).

    Based on the information provided, here's an analysis regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a quantitative manner as one might expect for an AI/ML device. Instead, it describes general compliance and qualitative outcomes for various tests.

    The reported performance is described in a narrative format, indicating that tests "met all acceptance criteria" and "performed similarly to the predicate devices."

    Acceptance Criteria CategoryReported Device Performance
    Biocompatibility- Cytotoxicity: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
    - Sensitization: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
    - Irritation: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
    - Systemic Toxicity (Acute Systemic Injection, Material Mediated Pyrogen): Passed (Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
    - Hemocompatibility (Thrombogenicity, Prothrombin Time (PT), Partial Thromboplastin Time (PTT), Complement Activation, Direct/Indirect hemolysis): Passed (Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
    Conclusion: "The results from the testing performed showed the subject devices to be biocompatible."
    Design Verification (Bench-Top) Testing- Dimensional/Visual Inspection: Met acceptance criteria (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter and Select Catheter)
    - Performance/Simulated Use Testing: Met acceptance criteria (Lightning Flash Apiration Tubing, Flash Aspiration Catheter and Select Catheter)
    - Tensile Testing: Met acceptance criteria (Lightning Flash Apiration Tubing)
    - Friction Testing: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
    - Torsion Testing: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
    - Vacuum Test: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
    - Indigo Aspiration System Compatibility: Met acceptance criteria
    - Valve Sense Testing: Met acceptance criteria
    - Coating Integrity Testing: Met acceptance criteria
    - Particulate Testing: Met acceptance criteria
    - Hub Air Aspiration: Met acceptance criteria
    - Catheter Pressure: Met acceptance criteria
    - Hub / Shaft Tensile Strength: Met acceptance criteria
    - Catheter Shaft Tensile: Met acceptance criteria
    - Elongation to Failure: Met acceptance criteria
    - Corrosion: Met acceptance criteria
    Conclusion: "The in vitro bench-top tests demonstrated that the subject Lightning Flash met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Lightning Flash devices function as intended and does not raise any new questions of safety and effectiveness compared to the predicate devices."
    Electrical Safety/EMC Testing- Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and ISO 10079-1: Complied
    Software Verification and Validation- Compliance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices": Provided and tested (Minor Level of Concern)

    2. Sample size used for the test set and the data provenance:

    The document does not provide details on the specific sample sizes used for each non-clinical (bench-top) test or biocompatibility test. It mentions "testing was conducted" and "tests were performed" without quantifying the number of units or replicates involved.

    Data provenance (e.g., country of origin, retrospective/prospective) is not applicable as the studies described are non-clinical bench-top tests and biocompatibility assessments, not clinical studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is an embolectomy catheter system, which is a physical medical device, not an AI/ML diagnostic or prognostic tool that requires expert interpretation to establish ground truth for a test set. The "ground truth" for its performance is established through measurable physical and biological parameters in laboratory settings (e.g., tensile strength, aspiration efficacy in simulated models, biocompatibility of materials).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable for the reasons stated in point 3. The studies are technical verification and validation, not studies requiring adjudication of image interpretations or clinical outcomes by multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. An MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-driven diagnostic tools, not for a mechanical medical device like an aspiration system. The document explicitly states: "No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This concept is not applicable as the Indigo® Aspiration System – Lightning® Flash is a mechanical medical device, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is based on well-established engineering standards, material science principles, and biological safety assessments. This includes:

    • Quantitative measurements from bench-top tests (e.g., dimensions, tensile strength, vacuum pressure).
    • Adherence to recognized standards (e.g., ISO, ASTM for materials and performance, IEC for electrical safety).
    • Biocompatibility testing results (e.g., lack of cytotoxicity, sensitization).
    • Functional performance in simulated use models (e.g., removal of simulated thrombi).

    8. The sample size for the training set:

    Not applicable. This device is not an AI/ML product that undergoes a "training set" for model development. The development process involves engineering design, prototyping, and testing of physical units.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no "training set" in the context of this mechanical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1