    Intended Use

    The PureZero LIMON Nitrile Powder-Free Exam Gloves with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid claim and PureZero LIMON-XTRA Nitrile Powder-Free Exam Gloves with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid claim is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:
    Cisplatin (1 mg/ml)
    Cyclophosphamide (20 mg/ml)
    Dacarbazine (10 mg/ml)
    Doxorubicin HCI (2 mg/ml)
    Etoposide (20 mg/ml)
    Fluorouracil (50 mg/ml)
    Ifosfamide (50 mg/ml)
    Methotrexate (25 mg/ml)
    Mitomycin C (0.5 mg/ml)
    Mitoxantrone HCI (2 mg/ml)
    Paclitaxel (6 mg/ml)
    Vincrinstine Sulfate (1 mg/ml)
    Fentanyl Citrate Injection (100 mcg/2 mL)
    Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 90 minutes:
    ThioTEPA (10 mg/ml), breakthrough detected at 87.6 minutes
    Carmustine (3.3 mg/ml), breakthrough detected at 34.3 minutes
    Warning: Not for Use With: Carmustine, ThioTEPA

    Device Description

    The subject device is a green colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, patient examination gloves.

    AI/ML Overview

    This document, K220541, is a 510(k) Premarket Notification from the FDA regarding "PureZero LIMON Nitrile Powder-Free Exam Gloves" and "PureZero LIMON-XTRA Nitrile Powder-Free Exam Gloves." It focuses on the substantial equivalence of these gloves to a predicate device and their resistance to permeation by various substances.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Important Note: This document describes the testing performed for medical gloves to demonstrate their chemical resistance and physical properties, rather than an AI/ML powered medical device. Therefore, questions 2, 3, 4, 5, 8, and 9, which apply specifically to AI/ML device studies (test set sample size, data provenance, expert adjudication, MRMC studies, training set details), are not directly applicable to this document's content.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents several key performance metrics for the gloves, tested against specific ASTM and ISO standards.

    TestStandard(s)Acceptance CriteriaReported Device Performance (Subject Device)
    Resistance to Permeation by Chemotherapy Drugs, Fentanyl Citrate, and Gastric AcidASTM D6978-05For "NO Breakthrough detected up to 240 minutes": Cisplatin (1 mg/ml), Cyclophosphamide (20 mg/ml), Dacarbazine (10 mg/ml), Doxorubicin HCl (2 mg/ml), Etoposide (20 mg/ml), Fluorouracil (50 mg/ml), Ifosfamide (50 mg/ml), Methotrexate (25 mg/ml), Mitomycin C (0.5 mg/ml), Mitoxantrone HCl (2 mg/ml), Paclitaxel (6 mg/ml), Vincristine Sulfate (1 mg/ml), Fentanyl Citrate Injection (100 mcg/2 mL), Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution. For "breakthrough detected in less than 90 minutes": ThioTEPA (10 mg/ml), Carmustine (3.3 mg/ml). (Note: Specific breakthrough times would be the acceptance criteria for these, implied to be under 90 minutes, with a warning not to use for these specific chemicals).PureZero LIMON / LIMON-XTRA Gloves:NO breakthrough detected up to 240 minutes for: Cisplatin (1 mg/ml), Cyclophosphamide (20 mg/ml), Dacarbazine (10 mg/ml), Doxorubicin HCI (2 mg/ml), Etoposide (20 mg/ml), Fluorouracil (50 mg/ml), Ifosfamide (50 mg/ml), Methotrexate (25 mg/ml), Mitomycin C (0.5 mg/ml), Mitoxantrone HCI (2 mg/ml), Paclitaxel (6 mg/ml), Vincrinstine Sulfate (1 mg/ml), Fentanyl Citrate Injection (100 mcg/2 mL), Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution. • Breakthrough detected in less than 90 minutes for: ThioTEPA (10 mg/ml) at 87.6 minutes, Carmustine (3.3 mg/ml) at 34.3 minutes. • Warning: Not for Use With: Carmustine or ThioTEPA.
    Detection of Holes (Leakage)ASTM D5151-06, ASTM D63192.5% AQL (Acceptable Quality Level) (or implied to meet this general glove standard), No leakage.Testing shows it meets the 2.5% AQL requirement in the standards for leakage. The device meets the acceptance criteria of the standard.
    Residual PowderASTM D6124-06, ASTM D6319< 2 mg maximum powder per glove.Residual powder is an average of 0.4 mg/glove, which is within the powder-free limit of < 2 mg, meeting acceptance criteria.
    Physical Dimensions and Properties (Tensile Strength, Elongation)ASTM D6319-10Dimensions: • Length: 235-250 mm (PureZero LIMON), 295-325 mm (PureZero LIMON-XTRA)• Palm Width: XSmall: 60-80 mm, Small: 70-90 mm, Med: 85-105 mm, Large: 100-120 mm, X-Large: 110-130 mm• Finger thickness: 0.10-0.19 mm• Palm thickness: 0.10-0.16 mm• Cuff thickness: 0.10-0.13 mmPhysical Properties:• Before Aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥500%• After Aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥400%The physical dimensions of the subject device are within the limits of the standard. The physical properties met the requirements for tensile strength before and after aging. (Results: Meets requirements for dimensions and physical properties).
    Biocompatibility - Systemic ToxicityISO 10993-11No animals treated with test extracts exhibit greater reaction than control animals.No evidence of systemic toxicity. Meets acceptance criteria. The test article was considered non-toxic.
    Biocompatibility - Irritation (Skin)ISO 10993-10Primary Irritation Index ≤ 2.0.Under the conditions of the study, the device is not an irritant. Meets acceptance criteria. The test article was considered nonirritating.
    Biocompatibility - Sensitization (Skin)ISO 10993-10Grade < 1.Under the conditions of the study, the device is not a sensitizer. Meets acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each specific test (e.g., number of gloves tested for permeation, number of gloves for physical properties). However, it refers to standard testing methods (ASTM and ISO), which typically specify minimum sample sizes for statistical validity. The provenance of the data is not mentioned in terms of country of origin or whether it was retrospective or prospective. Given the nature of physical product testing, it would be prospective laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The device is a physical barrier (glove), not an AI/ML diagnostic or image interpretation tool. Ground truth is established through standardized laboratory tests measuring physical and chemical properties as per ASTM and ISO guidelines, not through expert human interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods are relevant for subjective interpretations (e.g., medical image reads), not for objective physical and chemical testing of a medical device like a glove.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is not an AI-powered device, nor is it a device that directly assists human readers in medical interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. This device is a physical product (glove), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device's performance is established by standardized laboratory testing procedures and industry consensus standards (ASTM and ISO). These standards define the methodologies and acceptable performance thresholds for properties like chemical permeation, physical dimensions, tensile strength, and biocompatibility. For example:

    • Chemical Permeation: Breakthrough time is objectively measured using analytical techniques.
    • Hole Detection: Based on a quantitative water leak test.
    • Physical Properties: Measured using calibrated equipment (e.g., tensile tester for strength and elongation, calipers for dimensions).
    • Biocompatibility: In-vitro or in-vivo biological assays with defined endpoints.

    8. The sample size for the training set

    This question is not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. This is not an AI/ML device that requires a training set.