510(k) Data Aggregation

    K Number
    K220372

    Validate with FDA (Live)

    Date Cleared
    2022-05-17

    (97 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Heel Stick Safety Lancet is intended for the collection of capillary blood from the heel of newborn, preemie, and toddler. The lancet has equipped with safety protection features.

    Device Description

    The heel stick safety lancet is offered in various depth size (0.65mm, 0.85mm, 1.00mm, 1.50mm) and width size (1.50 mm, 1.75mm, 2.50 mm, 2.80 mm). It consists of blade base, blade, which is welded with the blade base, spring, top head, bottom, and pushing button. The top head and bottom are intended to provide physical protection to the blade, and PET blister along with the tyvek paper cover the primary sterile barrier system for the device. The product is individually primarily packaged and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf-life of the product is 5 years. Heel stick safety lancet is intended to be used by professionals. The product is intended for prescription (Rx) only.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the "Heel Stick Safety Lancet (HHZ-II, HHZ-III)" and does not involve an AI or Machine Learning (ML) enabled device. Therefore, many of the typical acceptance criteria and study components requested in the prompt, such as data provenance, expert adjudication, MRMC studies, standalone algorithm performance, and training set details, are not applicable.

    The acceptance criteria provided for this device concern its physical and performance characteristics as a manual surgical instrument, not an AI model.

    Here's an attempt to answer the prompt based on the provided document, highlighting where the requested information is not relevant due to the nature of the device:

    Device: Heel Stick Safety Lancet (HHZ-II, HHZ-III)
    Device Type: Manual Surgical Instrument For General Use (Lancet)

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemAcceptance CriteriaReported Device Performance
    AppearanceThe product color shall be correct, with the surface without burr and scratchesMeet the requirement
    CleannessThe product shall be cleanedMeet the requirement
    Dimension (HHZ-II)length: 32.0mm±1.0mm; width: 11.5mm±0.5mmMeet the requirement
    Dimension (HHZ-III)length: 36.65mm±0.5mm; width: 11.6±0.5mmMeet the requirement
    Firmness (Blade and blade base connection)Blade and blade base shall be firmly connected.Meet the requirement
    Launch performanceLaunch performance should be good, launch button press smoothly, no jamMeet the requirement
    Launch depth and Launch widthLaunch depth and launch width shall meet the requirements.Meet the requirement
    Force to activate the safety feature4 - 10NMeet the requirement
    Testing access to the sharp in safe modeThe blade shall not touch the sphere.Meet the requirement

    Biocompatibility Testing Acceptance and Results:

    ItemAcceptance Criteria (Test Results)Reported Device Performance
    In Vitro CytotoxicityNo CytotoxicityNo Cytotoxicity
    Skin SensitizationNo Skin sensitizationNo Skin sensitization
    Intracutaneous reactivityNo irritationNo irritation
    Acute Systemic ToxicityNo Acute Systemic ToxicityNo Acute Systemic Toxicity
    PyrogenicityNo thermogenic reactionNo thermogenic reaction

    2. Sample Size Used for the Test Set and Data Provenance

    For the Simulated Clinical Use study to evaluate the safety mechanism:

    • Sample Size: 600 device samples each for the HHZ-II and HHZ-III models (implicitly, 1200 samples total, though not explicitly stated as combined).
    • Data Provenance: Not specified, but generally, bench testing and simulated clinical use for a medical device cleared via 510(k) are conducted by the manufacturer or their contracted laboratories to demonstrate compliance with standards. It's a prospective test of the device. Country of origin not specified for the test data, but the manufacturer is in China.

    For the Non-Clinical (Bench) Testing (Appearance, Cleanness, Dimension, etc.):

    • Sample Size: Not explicitly stated for each test, but standard quality control and performance testing typically involve a defined sample size from production batches.
    • Data Provenance: Not specified, likely internal testing by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    This information is not applicable as the device is a manual lancet and the studies are performance/bench/biocompatibility tests, not studies involving expert adjudication of AI outputs or clinical interpretations. The "ground truth" for these tests is based on measurable physical properties and specified standards (e.g., ISO standards).

    4. Adjudication Method for the Test Set

    This is not applicable for the type of device and testing performed. Adjudication methods like "2+1" or "3+1" are typical for clinical studies involving human interpretation (e.g., for image reading in AI studies), not for mechanical device performance or biocompatibility testing. The "ground truth" (acceptance criteria) for the bench tests are against physical measurements and functional performance as per engineering specifications and relevant ISO standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted image interpretation or diagnostic tools to assess the impact of AI on human reader performance. The Heel Stick Safety Lancet is a manual instrument.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone algorithm performance study was not done. This is not an AI/algorithm-based device. The "simulated clinical use" mentioned is a physical test of the device's safety mechanism, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests conducted on the Heel Stick Safety Lancet consisted of:

    • Pre-defined engineering specifications and measurable physical properties: For criteria like appearance, cleanness, dimensions, firmness, launch performance, launch depth/width, activating force, and sharp access in safe mode.
    • Compliance with international standards: Specifically, ISO 23908 for evaluating the safety mechanism and ISO 10993 series for biocompatibility.

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical, manual medical device, not an AI/ML product that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no AI/ML training set for this device.