    Intended Use

    Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Examination Gloves are made from nitrile butadiene rubber that covers the hand up to the wrist. They are cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. The device is available in four sizes, which are S, M, L, XL. It could be selected by the user depending on the size of hand. All sizes share the same blue color. The gloves are non-sterile, powder free and are for single use only.

    AI/ML Overview

    The provided document is a 510(k) summary for Nitrile Examination Gloves (Model: ZMG1351). It details the device's characteristics, comparison to a predicate device, and results from non-clinical testing. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    Dimension - ASTM D6319-19Evaluate the glove palm widthS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmS: 8486mmM: 9698mmL: 104106mmXL: 113115mm
    Evaluate the glove lengthS: Min 220mmM: Min 230mmL: Min 230mmXL: Min 230mmS: 241251mmM: 236249mmL: 237245mmXL: 240249mm
    Evaluate the glove finger thicknessS, M, L, XL: Min 0.05mmS: 0.1110.137mmM: 0.1240.150mmL: 0.1120.133mmXL: 0.1170.137mm
    Evaluate the glove palm thicknessS, M, L, XL: Min 0.05mmS: 0.0800.097mmM: 0.0870.098mmL: 0.0740.086mmXL: 0.0770.094mm
    Physical Properties - ASTM D6319-19Evaluate the glove physical properties before and after agingBefore Aging:- Tensile Strength: Min 14Mpa- Ultimate Elongation: Min 500%After Aging:- Tensile Strength: Min 14Mpa- Ultimate Elongation: Min 400%Before Aging:- Tensile Strength: 19.927.4Mpa- Ultimate Elongation: 509553%After Aging:- Tensile Strength: 20.135.1Mpa- Ultimate Elongation: 454504%
    Freedom from Holes - ASTM D6319-19Detect the holes that allow water leakageDo not show droplet, stream, or other type of water leakage (Pass at AQL 2.5, as per comparison table)Performance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
    Powder Residue - ASTM D6319-19Evaluate the residue powder≤2mg/glovePerformance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
    Skin Irritation - ISO 10993-10:2010Evaluated for the potential to cause skin irritationUnder the condition of the test, not an irritant.Performance reported as meeting the acceptance criteria (based on the "Test Results" column not explicitly stating values but confirming compliance with acceptance criteria for all tests)
    Sensitization - ISO 10993-10:2010Evaluated for the potential to cause skin sensitizationUnder the condition of the test, not a sensitizer.No-sensitization (Performance reported as meeting the acceptance criteria as per "No- sensitization" in "Test Results" column and confirmation of compliance for all tests)
    Acute Systemic Toxicity - ISO 10993-11:2017Evaluated for acute systemic toxicityUnder the condition of the test, not a systemic toxicity.No-acute systemic toxicity (Performance reported as meeting the acceptance criteria as per "No- acute systemic toxicity" in "Test Results" column and confirmation of compliance for all tests)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes used for each test. It simply states that "Bench tests were conducted to verify that the proposed device met all design specifications." The data provenance is not explicitly mentioned, but given the manufacturer (Wuhan Zonsen Medical Products Co.,Ltd) and contactor (Shanghai CV Technology Co., Ltd.) locations, it is highly likely the testing and data originated from China. The study appears to be retrospective in the sense that these are design verification tests performed on manufactured gloves.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the document describes physical and biocompatibility testing of medical gloves, not an AI or diagnostic device that requires expert interpretation to establish a ground truth. The "ground truth" for these tests is based on established international standards (ASTM and ISO).

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as above. The tests performed are objective measurements against predefined criteria in established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, not for basic medical devices like examination gloves.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical product (nitrile examination gloves), not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the acceptance criteria and testing of the Nitrile Examination Gloves is based on established industry standards and protocols:

    • ASTM D6319-19: Standard Specification for Nitrile Examination Gloves for Medical Application.
    • ISO 10993-10: Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
    • ISO 10993-11: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
      The tests measure physical properties and biological responses against scientifically defined and accepted limits.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is not an AI or machine learning device. The tests are for product verification and validation.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above.