    K213582

    Validate with FDA (Live)

    Date Cleared
    2022-06-30

    (230 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epicardial Access System is intended to access the epicardial surface of the heart via a subxiphoid approach.

    Device Description

    The subject Epicardial Access System represents device modifications made to the Epicardial Steerable Guiding Sheath and Epicardial Introducer and Stylet components of the legally marketed Epicardial Access System (K191546). The Epicardial Access Needle component remains unchanged. The System is comprised of the following devices, which are single-use and supplied sterile to the user:

    • Epicardial Access Needle
    • Epicardial Steerable Guiding Sheath containing:
      • Steerable Sheath
      • Dilator
      • J-tip Guidewire
    • Epicardial Introducer and Stylet
      The subject device is designed to provide access to the epicardial surface of the heart via a subxiphoid approach. It is used in percutaneous access procedures primarily by Electrophysiologists and Interventional Cardiologists trained in the subxiphoid approach. Procedures using the devices are performed in fully equipped catheter labs with imaging equipment, including fluoroscopy under sterile technique.
    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Epicardial Access System), which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing a new product validation with acceptance criteria based on clinical outcomes or performance metrics like accuracy, sensitivity, or specificity, as would be typical for AI/ML-based devices.

    Therefore, the supplied text does not contain the information requested for acceptance criteria and a study proving a device meets those criteria, specifically concerning data provenance, clinical performance metrics, expert adjudication, or MRMC studies. The document primarily details non-clinical performance testing (mechanical, general physical, biocompatibility, sterilization, pyrogen, and packaging verification) to show the modified device performs similarly to its predicate.

    It describes hardware modifications to an existing system, not the development or validation of an AI/ML algorithm.

    Here's a breakdown of why each requested point cannot be addressed from the given text:

    1. A table of acceptance criteria and the reported device performance:

      • Does not exist in the document. The document describes various non-clinical tests performed (mechanical, physical, biocompatibility, etc.) but does not list pre-defined quantitative acceptance criteria (e.g., success rates, tolerance levels, specific metric thresholds) for these tests, nor does it provide the reported performance values from these tests beyond a general statement that "Epicardial Access System met all test requirements."
      • This type of detailed performance data and acceptance criteria would typically be found in internal test reports, not in an FDA 510(k) summary focused on substantial equivalence.
    2. Sample sizes used for the test set and the data provenance:

      • Not applicable/Not present. The performance testing described is primarily non-clinical (benchtop mechanical, physical, etc.). There is no "test set" in the sense of a clinical dataset with patient data.
      • There is no mention of "data provenance" (country of origin, retrospective/prospective) because no clinical data or clinical test sets are used or described for device acceptance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not present. Since there is no clinical test set requiring human expert ground truth establishment, this information is not provided.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not present. No clinical test set.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable/Not present. This document discusses a physical medical device (catheter introducer system), not an AI/ML device. Therefore, no MRMC study or AI assistance evaluation was performed.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable/Not present. Not an AI/ML algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable/Not present. The "ground truth" for the non-clinical tests would be defined by engineering specifications and established test methods, not clinical outcomes or expert consensus.
    8. The sample size for the training set:

      • Not applicable/Not present. This is not an AI/ML device; therefore, there is no training set mentioned.
    9. How the ground truth for the training set was established:

      • Not applicable/Not present. No training set.

    In summary, the provided FDA document pertains to a hardware medical device (Epicardial Access System) undergoing modifications and seeking 510(k) clearance by demonstrating substantial equivalence through non-clinical performance and engineering tests, not a software/AI device validated with clinical performance metrics against ground truth data.