510(k) Data Aggregation

    K Number
    K213051

    Validate with FDA (Live)

    Date Cleared
    2021-12-18

    (87 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in six sizes: XS, S, M, L, XL, XXL. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs). It details the device's technical characteristics, intended use, and performance testing results to demonstrate substantial equivalence to a predicate device.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core performance evaluation for these gloves relates to their physical properties, freedom from holes, powder content, biocompatibility, and most importantly, resistance to permeation by chemotherapy drugs.

    Test MethodPerformance CharacteristicAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions Test
    Length (mm) - XS≥220> 240/Pass
    Length (mm) - S/M/L/XL/XXL≥230> 240/Pass
    Width (mm) - XS70 ± 1067-72/Pass
    Width (mm) - S80 ± 1076-83/Pass
    Width (mm) - M95 ± 1091-99/Pass
    Width (mm) - L110 ± 10106-112/Pass
    Width (mm) - XL120 ± 10116-124/Pass
    Width (mm) - XXL130 ± 10127-134/Pass
    Thickness (mm) - Finger≥0.050.06-0.10/Pass
    Thickness (mm) - Palm≥0.050.07-0.10/Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass (This implies 0 holes out of 125 gloves tested, meeting or exceeding AQL 2.5)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 ≤ 2.0mg0.02mg/Pass
    ASTM D412Physical Properties (Tensile Strength & Elongation)
    Before Aging:
    Tensile Strength≥14MPa16.4-17.8MPa/Pass
    Ultimate Elongation≥500%533-551%/Pass
    After Aging:
    Tensile Strength≥14MPa15.8-16.8MPa/Pass
    Ultimate Elongation≥400%520-554%/Pass
    ISO 10993-5Biocompatibility - CytotoxicityNon- In Vitro CytotoxicityUnder conditions of the study, device extract is not cytotoxic. /Pass
    ISO 10993-10Biocompatibility - IrritationNon-irritatingUnder the conditions of the study, not an irritant. / Pass
    ISO 10993-10Biocompatibility - SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass
    **ASTM D6978-05 (Reapproved 2019) **Chemotherapy Drug Permeation (Minimum Breakthrough Detection Time)
    Carboplatin (10.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Carmustine (BCNU) (3.3 mg/ml)N/A (Comparative to predicate)35.2 Minutes
    Cyclophosphamide (Cytoxan) (20.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Doxorubicin (2.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Etoposide (20.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Fluorouracil (50.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Methotrexate (25.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    Paclitaxel (6.0 mg/ml)N/A (Comparative to predicate)> 240 Minutes
    ThioTepa (10.0 mg/ml)N/A (Comparative to predicate)76.3 Minutes

    Note on Chemotherapy Permeation Acceptance Criteria: The document primarily presents these as reported values and compares them to the predicate device in Table 1 (pages 6-7). The acceptance criteria for these would implicitly be that the breakthrough times are considered sufficient for safe use, or comparable/better than the predicate for equivalent drugs, with a clear warning for drugs with low permeation times. The "Warning: Please do not use with Carmustine (BCNU) and ThioTepa" indicates that the low breakthrough times for these specific drugs are accepted with a cautionary label, rather than requiring an explicit "minimum time" acceptance criterion for regulatory clearance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the total sample size for all tests performed. However, it does provide a specific sample size for one critical test:

    • Watertightness Test (ASTM D5151): 125 gloves were tested (indicated by "0/125/Pass").
    • For other tests (e.g., physical dimensions, tensile strength, powder content, biocompatibility, chemotherapy permeation), the specific number of gloves or samples tested is not explicitly stated in the provided summary, although the ASTM standards themselves would specify statistical sampling plans.
    • Data Provenance: The tests were conducted by the manufacturer, Yunnan Huazhiyuan Medical Technology Co., Ltd. The location is China. The studies are retrospective in the sense that they are conducted on manufactured devices to demonstrate compliance with standards for the 510(k) submission, rather than being part of a prospective clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable to this type of device and submission. The "ground truth" for the performance of these medical gloves is established through adherence to recognized international and national standards (ASTM, ISO) for physical properties, chemical resistance, and biocompatibility, as measured by laboratory equipment and procedures, not by expert human interpretation or consensus. No human "experts" are interpreting images or clinical data for this non-diagnostic device.

    4. Adjudication Method for the Test Set

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly in imaging or diagnostic device evaluations where multiple readers interpret data and discrepancies need to be resolved. For laboratory bench testing against physical and chemical standards, there is no human interpretation needing adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    This section is not applicable. MRMC studies are specific to diagnostic devices, especially those involving image interpretation by multiple human readers, to assess the impact of an AI algorithm on reader performance. This device is a medical glove, a Class I non-diagnostic device, and its performance is evaluated through bench testing against established standards, not human reader studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This section is not applicable. This device is not an algorithm or AI-based product. Its performance is entirely standalone in the sense that its physical and chemical properties are measured directly from the device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based entirely on laboratory bench testing results measured against predefined acceptance criteria established by recognized international standards (ASTM and ISO).

    • Physical properties: Measured dimensions, tensile strength, elongation.
    • Freedom from holes: Tested using a water-tightness test.
    • Powder content: Measured residual powder.
    • Biocompatibility: In vitro cytotoxicity, irritation, and sensitization tests.
    • Chemotherapy drug permeation: Measured breakthrough times using specific chemical assays as per ASTM D6978-05.

    8. The Sample Size for the Training Set

    This section is not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" for the device's manufacturing process would be based on quality control and manufacturing specifications, not a data training set.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable as it pertains to AI/ML devices.