Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The blue colored nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device. The Nitrile Patient Examination Gloves Blue Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The tested chemotherapy drug are as follows: Bleomycin Sulfate 15 mg/ml >240 min., Carboplatin 10 mg/ml >240 min., Carmustine (BCNU) 3.3 mg/ml 17.2 min., Cisplatin 1.0 mg/ml >240 min., Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min., Cytarabine HCI 100 mg/ml >240 min., Dacarbazine (DTIC)10.0 mg/ml >240 min., Daunorubicin 5.0 mg/ml >240 min., Docetaxel 10.0 mg/ml >240 min, Doxorubicin HCI 2.0 mg/ml >240 min., Etoposide (Toposar) 20.0 mg/ml >240 min., Fluorouracil 50.0 mg/ml >240 min., Gemcitabine 38 mg/ml>240 min., Idarubicin 1 mg/ml >240 min., Ifosfamide 50.0 mg/ml >240 min., Irinotecan 20.0 mg/ml >240 min., Mechlorethamine HCI 1.0 mg/ml>240 min., Melphalan 5 mg/ml >240 min., Methotrexate 25 mg/ml >240 min., Mitromycin C. 0.5 mg/ml >240, Mitoxantrone 2.0 mg/ml >240 min., Paclitaxel (Taxol) 6.0 mg/ml >240 min., Thiotepa 10.0 mg/ml 13.9 min., Vincristine Sulfate 1.0 mg/ml >240 min. Warning: Do not use with Carmustine and Thiotepa.

    The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device. The Nitrile Patient Examination Gloves Blue Violet Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The tested chemotherapy drug are as follows: Carmustine (BCNU) 3.3 mg/ml 65.3 min., Cisplatin 1.0 mg/ml >240 min., Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min., Dacarbazine (DTIC)10.0 mg/ml >240 min., Doxorubicin HCI 2.0 mg/ml >240 min., Etoposide (Toposar) 20.0 mg/ml >240 min., Fluorouracil 50.0 mg/ml >240 min., Methotrexate 25 mg/ml >240 min., Paclitaxel (Taxol) 6.0 mg/ml >240 min., Thiotepa 10.0 mg/ml 58.3 min. Warning: Do not use with Carmustine and Thiotepa.

    The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device. The Nitrile Patient Examination Gloves Black Colored were tested for use with Fentanyl Citrate per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Fentanyl Citrate 100mcg/2ml >240 minutes.

    Device Description

    Power-Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are offered in six sizes, x-small (6.5"), small (7"), medium (8"), large (8.5"), X-large (9"), XXL (9.5"). Three colors are available for all size, includes blue, blue violet and black. The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019).

    AI/ML Overview

    The provided document describes the acceptance criteria and study results for "Nitrile Patient Examination Gloves" (various colors) tested for use with chemotherapy drugs or Fentanyl Citrate. This is a medical device, and the evaluation is based on non-clinical performance standards rather than an AI/ML algorithm.

    Therefore, many of the requested fields related to AI/ML studies (sample size for test set, data provenance, number of experts, adjudication method, MRMC comparative effectiveness, standalone performance, ground truth for training set, training set size) are not applicable to this submission.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemMethodology / StandardAcceptance CriteriaReported Device Performance
    Physical Dimensions
    Palm widthASTM D3767-03(2020)XS (70±10mm)S (80±10mm)M (95±10mm)L (110±10mm)XL (120±10mm)XXL (≥ 120mm)Pass
    LengthASTM D3767-03(2020)XS (220mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)XXL (230mm min)Pass
    ThicknessASTM D3767-03(2020)Finger: 0.11mmPalm: 0.05mmPass
    Integrity
    Freedom from holesASTM D5151-19Freedom free holeAQL 2.5Pass
    Physical Properties (before aging)
    Tensile StrengthASTM D412-16≥ 14MpaPass
    Ultimate ElongationASTM D412-16≥ 500%Pass
    Physical Properties (after aging)
    Tensile StrengthASTM D573-04(2019)≥ 14MpaPass
    Ultimate ElongationASTM D573-04(2019)≥ 400%Pass
    Chemical Content
    Residual Powder ContentASTM D6124-06(2017)≤2mg per glovePass
    Biocompatibility
    CytotoxicityISO 10993-5Non-cytotoxicUnder conditions of the study, device extract is cytotoxic.
    Acute Systemic ToxicityISO 10993-11Non-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
    IrritationISO 10993-10Non-irritatingUnder the conditions of the study, not an irritant/ Pass
    SensitizationISO 10993-10Non-sensitizingUnder conditions of the study, not a sensitizer./ Pass
    Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2019)Min. permeation breakthrough time as specified for each drugAchieved or exceeded specified times for individual drugs (see detailed lists in pages 2, 4, 5, 7, 8, 11, 12).
    Fentanyl Citrate PermeationASTM D6978-05 (Reapproved 2019)Min. permeation breakthrough time as specified (e.g., >240 min)Achieved >240 minutes for Fentanyl Citrate (see detailed lists in pages 5, 8, 12).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for each test, but standard ASTM and ISO methods generally specify sample sizes. For example, AQL 2.5 for "freedom from holes" implies a sampling plan. Since these are performance tests on manufactured goods, the "test set" would be a batch or set of gloves sampled from production.
    • Data Provenance: The document does not specify a country of origin for the testing data. The studies are described as "Non clinical tests" performed according to established international standards (ASTM, ISO). They are prospective in the sense that they are performed on the device to be marketed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This is not applicable as the "ground truth" for these tests is defined by objective physical and chemical measurements according to standardized protocols (ASTM, ISO), not expert interpretation.

    4. Adjudication method for the test set

    • Not applicable, as the evaluation is based on objective measurements against defined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a submission for physical medical gloves, not an AI/ML-driven device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a submission for physical medical gloves, not an AI/ML-driven device.

    7. The type of ground truth used

    • The "ground truth" for these tests is the direct measurement of physical properties (e.g., dimensions, tensile strength, elongation), chemical properties (e.g., residual powder content), and barrier properties (permeation breakthrough time for chemotherapy drugs and Fentanyl Citrate) against established, objective industry standards (ASTM, ISO). For biocompatibility, it's the observed biological response to the glove material under controlled test conditions.

    8. The sample size for the training set

    • Not applicable. This is a submission for physical medical gloves, not an AI/ML-driven device.

    9. How the ground truth for the training set was established

    • Not applicable. This is a submission for physical medical gloves, not an AI/ML-driven device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1