Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212847

    Validate with FDA (Live)

    Device Name
    Penta Glove
    Date Cleared
    2022-01-06

    (121 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for the "Penta Glove," a Blue Nitrile Examination Glove. It is a 510(k) submission, meaning it aims to demonstrate substantial equivalence to a legally marketed predicate device rather than undergoing a de novo approval process.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodPurposeAcceptance CriteriaReported Device PerformanceRemarks
    ASTM D6319-2019 (Length)To determine the length of the glovesMin 230 mm for all sizesX-Small: 246 mm, Small: 246 mm, Medium: 248 mm, Large: 248 mm, X-Large: 250 mmMeets criteria
    ASTM D6319-2019 (Width)To determine the width of the glovesX-small: 70±10 mm, Small: 80±10 mm, Medium: 95±10 mm, Large: 110±10 mm, X-Large: 120±10 mmX-small: 68 mm, Small: 80 mm, Medium: 92 mm, Large: 105 mm, X-Large: 115 mmMeets criteria (all reported values are within the +/-10mm range of the nominal size criteria)
    ASTM D6319-2019 (Thickness)To determine the thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm min for all sizesX-Small: 0.12mm (Palm), 0.20mm (Finger); Small: 0.12mm (Palm), 0.20mm (Finger); Medium: 0.12mm (Palm), 0.20mm (Finger); Large: 0.12mm (Palm), 0.20mm (Finger); X-Large: 0.12mm (Palm), 0.20mm (Finger)Meets criteria
    ASTM D6319-2019 (Physical Properties - Tensile Strength)To Determine the physical properties - Tensile strengthBefore Ageing: 14 Mpa Min for all sizes; After Ageing: 14 Mpa Min for all sizesBefore Ageing: X-Small 18.45 Mpa, Small 18.54 Mpa, Medium 18.62 Mpa, Large 18.67 Mpa, X-Large 18.72 Mpa; After Ageing: X-Small 17.54 Mpa, Small 17.56 Mpa, Medium 17.68 Mpa, Large 17.75 Mpa, X-Large 17.80 MpaMeets criteria (all reported values are above 14 Mpa)
    ASTM D6319-2019 (Physical Properties - Ultimate Elongation)To Determine the physical properties - Ultimate ElongationBefore Ageing: 500% Min for all sizes; After Ageing: 400% Min for all sizesBefore Ageing: X-Small 686%, Small 690%, Medium 694%, Large 702%, X-Large 705%; After Ageing: X-Small 659%, Small 665%, Medium 668%, Large 670%, X-Large 674%Meets criteria (all reported values are above the minimums)
    ASTM D5151-2019 (Water Tight / Holes)To determine the holes in the glovesAQL 1.5Gloves Passes AQL 1.5Meets criteria
    ASTM D6124-06 (Reapproved 2017) (Residual Powder)To determine the residual powder in the gloves2 Mg/Glove MaxX-small: 0.21mg/glove, Small: 0.21mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/gloveMeets criteria
    ISO 10993-10 (Primary Skin Irritation)To determine the potential of the material to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritantMeets criteria
    ISO 10993-10 (Dermal Sensitization)To determine the skin sensitization potential of the material in Guinea PigUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizerMeets criteria
    ISO 10993-5:2009 (In vitro Cytotoxicity)To evaluate the in vitro cytotoxic potential of the test item extracts in L-929 mouse fibroblasts cellsUnder the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.Does Not Meet Criteria (Reported as "cytotoxic" against a "non cytotoxic" acceptance criteria, though in the previous F. Summary of Technological Characteristics, it was listed as "Same" as predicate with predicate also being "cytotoxic"). This discrepancy needs clarification.
    ISO 10993-11:2017 (Acute Systemic Toxicity)To determine the acute systemic toxicity potential of the test item extracts in Swiss Albino miceUnder the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernMeets criteria
    USP 42 <85> (Bacterial Endotoxin)To determine the Bacterial Endotoxin limit in the gloveNMT 20 EU/pair of gloves<20 EU/pair of glovesMeets criteria

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes for each individual test. However, it references ASTM and ISO standards for the testing. These standards typically outline the required sample sizes for performing such tests (e.g., number of gloves per batch for AQL, number of animals for biocompatibility).

    • Data Provenance: The testing appears to be conducted by the manufacturer or a contracted lab to meet international standards (ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993 series, USP 42). The submitter, Pentavest Holdings Sdn Bhd, is located in Malaysia. Thus, the data provenance is likely from Malaysia or other regions where these standard tests are performed. The data is prospective for the device being submitted, as these tests would have been performed on samples of the Penta Glove specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the document describes the testing of a medical device (examination gloves) based on objective, standardized measurements and laboratory tests (e.g., tensile strength, dimensions, chemical residue, cytotoxicity, etc.). These tests rely on equipment and technical procedures to generate quantifiable results, not expert consensus or interpretation of subjective images/data. Therefore, there's no "ground truth" established by experts in the typical sense of clinical or diagnostic studies.

    4. Adjudication Method for the Test Set:

    This is not applicable. As mentioned above, the tests are objective measurements and laboratory analyses, not human interpretation requiring adjudication. Results are compared against predefined numerical or qualitative acceptance criteria specified in the relevant standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This is not applicable. The device is an examination glove, a physical medical device. It is not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is not applicable for the same reason as point 5. The device is a physical glove, not an algorithm.

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is based on objective, quantitative measurements and established international standards (ASTM, ISO, USP). For example:

    • Dimensions: Measured values compared to specified minimums/ranges.
    • Physical Properties: Measured tensile strength and elongation compared to minimum requirements.
    • Water Tightness: Tested against an Acceptable Quality Level (AQL) specification (e.g., 1.5).
    • Residual Powder: Measured mass of powder per glove compared to a maximum limit.
    • Biocompatibility: In-vitro or in-vivo biological responses (e.g., irritation, sensitization, cytotoxicity, systemic toxicity, endotoxin levels) are observed and compared against predefined criteria for biological safety (e.g., "not an irritant," "not a sensitizer," "non cytotoxic," "do not pose a systemic toxicity concern," "NMT 20 EU/pair").

    8. The Sample Size for the Training Set:

    This is not applicable. The Penta Glove is a physical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reason as point 8.