510(k) Data Aggregation

    K Number
    K212812

    Validate with FDA (Live)

    Device Name
    Surgical Gown
    Date Cleared
    2022-11-16

    (439 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).

    Device Description

    The Sterile Level 4 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The proposed devices are single use and EO sterilized. The Sterile Level 4 Surgical Gowns are available in five models including S、 M、L、XL and XXL. The proposed surgical are constructed of a SMS nonwoven material (SMS+Film covering material+Velcro tape+Cuff material) and it is a kind of reinforced surgical gown and blue color.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

    The document is a 510(k) Summary for a "Surgical Gown" (K212812). It focuses on demonstrating substantial equivalence to a predicate device (K121152) through non-clinical performance tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The table below summarizes the acceptance criteria and the claimed performance of the subject device, as presented in the "Comparison of technological characteristics with the Predicate" section and the "Test Method" table.

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    AATCC 127Hydrostatic pressure>50 cmPass
    AATCC 42Impact penetration≤1gPass
    ASTM D 5034Tensile StrengthLength(lbf): ≥37.6Width(lbf): ≥24.2Pass
    ASTM D 5733Tearing strengthLength yarns torn(lbf):14.3Width yarns torn(lbf):8.1Pass
    ASTM D 1683Seam strengthArm Opening sleeves: 23.5 lbf (FTS), 24.2lbf (FR)Shoulder: 37.0 lbf (FR)Pass
    16 CFR Part 1610Flammability of TextilesClass 1Class 1
    ASTM F1671Bacteriophage Phi-X174 penetrationNo penetrationPass
    Biocompatibility (ISO 10993-5:2009)CytotoxicityISO 10993-5 (non-cytotoxic)Passed (non-cytotoxic)
    Biocompatibility (ISO 10993-10:2002)IrritationISO 10993-10 (non-irritating)Passed (non-irritating)
    Biocompatibility (ISO 10993-10:2002)SensitizationISO 10993-10 (non-sensitizing)Passed (non-sensitizing)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual non-clinical test. It mentions that "Non clinical tests were conducted to verify that the proposed device met all design specifications." Without further detail, it is assumed that appropriate sample sizes, as typically required by the referenced ASTM and ISO standards for material testing, were used.

    The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. However, given that the submitter is "Xianning Marveldas Protective Articles CO.,Ltd." from China and the tests are for a new device submission, these would be prospective tests performed by the manufacturer, likely at their own facilities or contracted testing labs, to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable to this submission. The "ground truth" for the performance tests of a surgical gown is established by the specified acceptance criteria of recognized international standards (ASTM, AATCC, ISO, 16 CFR). These standards define objective, measurable physical and chemical properties, rather than subjective interpretations by human experts.

    4. Adjudication Method for the Test Set

    This section is not applicable. The tests are objective measurements against defined standards. There is no mention of "adjudication" in the context of expert review for establishing ground truth, as the ground truth is the numerical outcome of the physical/chemical tests and comparison against the specified acceptance criteria.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This section is not applicable. This submission is for a medical device (surgical gown) that is evaluated based on its physical properties and biocompatibility, not on a diagnostic or therapeutic function requiring human-in-the-loop performance or clinical effectiveness studies with human readers. The document explicitly states: "No clinical study is included in this submission."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This section is not applicable. The device is a surgical gown, not an AI algorithm or a device with an algorithmic component. The performance evaluated is the physical and biological characteristics of the material.

    7. The Type of Ground Truth Used

    The ground truth used for these non-clinical performance tests is based on established international and national standards and their defined acceptance criteria. These standards specify the methodology and the expected performance thresholds for various properties of surgical apparel, such as:

    • Physical Property Standards: AATCC 127 (Hydrostatic pressure), AATCC 42 (Impact penetration), ASTM D 5034 (Tensile Strength), ASTM D 5733 (Tearing strength), ASTM D 1683 (Seam strength), 16 CFR Part 1610 (Flammability of Textiles), ASTM F1671 (Bacteriophage Phi-X174 penetration).
    • Biocompatibility Standards: ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization).

    For each test, the "ground truth" is whether the device's measured performance meets or exceeds the specified numerical or categorical acceptance criteria set forth in these standards.

    8. The Sample Size for the Training Set

    This section is not applicable. There is no "training set" in the context of a surgical gown submission; this term is typically relevant for machine learning or AI algorithm development. The device's characteristics are evaluated directly through testing, not through a model trained on data.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reasons as point 8.