Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212402

    Validate with FDA (Live)

    Date Cleared
    2021-12-16

    (136 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MIVI Q Distal Access Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    The MIVI Q Distal Access Catheter (Q Catheter) is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q Catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The Q Docking Station may be used to facilitate insertion and extraction of the Q Catheter through a hemostasis valve attached to the 8F guide catheter/6F guide sheath. The pin vise may be used to advance the catheter.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the MIVI Q Distal Access Catheter:

    The document provided is a 510(k) Summary for the MIVI Q Distal Access Catheter, which is a premarket notification to the FDA for a medical device seeking substantial equivalence to a predicate device. This type of submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate, rather than proving novel effectiveness.

    Given the nature of a 510(k) submission for a catheter, the "acceptance criteria" and "device performance" in this context refer to the ability of the device to meet specified engineering and biocompatibility standards, and to perform its intended function without significant differences from the predicate. There is no clinical study in the traditional sense (e.g., patient trials with efficacy endpoints) described for this device, as it's a substantially equivalent determination based on non-clinical testing.

    Here's the information extracted and organized according to your request:


    Acceptance Criteria and Device Performance for MIVI Q Distal Access Catheter

    1. Table of Acceptance Criteria and Reported Device Performance

    For Bench Testing (Non-Clinical Performance):

    TestAcceptance CriteriaReported Device Performance
    Dimensional VerificationThe device must meet dimensional specifications.Pass
    System Introduction (Simulated Use)The device must function as intended.Pass
    Liquid Leakage under PressureThe device must hold a hydrostatic pressure.Pass
    Tensile StrengthTensile strength pull force minimum must be met.Pass

    For Biocompatibility Testing:

    TestAcceptance Criteria (Implied)Reported Device Performance
    CytotoxicityNon-toxic to cells (determined by specific extract and cell line).PASS
    Irritation or Intracutaneous ReactivityNo significant irritation or reactivity in vivo (animal model).PASS
    SensitizationNo significant sensitization reaction (animal model).PASS
    Acute Systemic ToxicityNo acute systemic toxicity (animal model).PASS
    Pyrogenicity (Material-mediated)Non-pyrogenic (no fever-inducing substances in vivo).PASS
    HemolysisNo significant hemolysis (destruction of red blood cells) in vitro.PASS
    Unactivated Partial Thromboplastin TimeNo significant impact on blood clotting time (in vitro).PASS
    Complement ActivationNo significant complement activation (immune response) in vitro.PASS

    For Sterilization and Shelf Life:

    TestAcceptance Criteria (Implied)Reported Device Performance
    SterilizationConfirmed sterility of the final packaging configuration and method.Met specification
    Shelf Life and Expiration DatingMaintain performance and integrity for the specified period.3 years (same as predicate)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes used for each individual non-clinical (bench and biocompatibility) test. Such details are typically found in the full test reports, not in the summary document.

    Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting (in vitro or in animal models), not on human subjects. Therefore, there is no country of origin for human patient data. The provenance of the data is the laboratory where the tests were performed. The studies are prospective in the sense that they were designed and executed to test the characteristics of the MIVI Q Distal Access Catheter.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not applicable to the provided document. The "ground truth" in this context is established by objective measurements and standardized test methods (e.g., ISO standards for biocompatibility, engineering specifications for bench tests), not by expert consensus on clinical data. No human "experts" were used to establish ground truth for this type of non-clinical testing.


    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly for interpreting imaging or clinical outcomes where there might be inter-reader variability. For non-clinical bench and biocompatibility tests, results are typically objective (e.g., a tensile strength value, a cytotoxicity rating) and do not require expert adjudication in this manner.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The MIVI Q Distal Access Catheter is a physical medical device (a catheter), not an AI algorithm, and therefore does not involve human readers interpreting output from an AI system. The submission is for substantial equivalence of a physical device.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. As explained above, this device is a physical catheter, not an AI algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • Bench Testing: Ground truth is defined by engineering specifications and objective physical measurements (e.g., specified dimensions, force thresholds, pressure integrity).
    • Biocompatibility Testing: Ground truth is established by standardized biological assays and observation in animal models and in vitro systems, as per recognized international standards (e.g., ISO 10993 series). The results are objective measurements against defined pass/fail criteria.

    8. The sample size for the training set

    This is not applicable. There is no AI algorithm involved that requires a training set. The "study" here consists of non-clinical performance and biocompatibility testing of the physical catheter.


    9. How the ground truth for the training set was established

    This is not applicable, as there is no AI algorithm or training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1