510(k) Data Aggregation

    K Number
    K212303

    Validate with FDA (Live)

    Device Name
    T2
    Date Cleared
    2021-09-21

    (60 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    T2 is digital X-Ray imaging equipment for dental professionals that convert X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.

    Device Description

    T2 is a digital X-ray CT, panoramic, and Cephalo imaging system device composed of X-ray generator, X-ray controller, X-ray supporter, image processing unit (sensor), PC, and software. The apparatus attached to the equipment column is a structure that can be rotated 360 by the system control unit. This system control unit actuates the motor control, X-ray generator, and image processing unit (sensor). The height controlling unit controls the column and adjusts the height of the equipment. The X-ray generator and image processing unit (sensor) are attached to the rotating apparatus. When the rotating apparatus starts the rotation, X-ray is irradiated from the X-ray generator (generating unit). This X-ray irradiation penetrates the subject and reaches the image processing unit (sensor), and then is converted into electric signals to secure imagery information. Inside the imaging section of the image processing unit (sensor), real time X-ray input is converted into electric signals and consecutively combined, resulting in imagery information. The combined panoramic imagery information is then sent to the PC and saved in patient management software.

    AI/ML Overview

    The provided document is a 510(k) summary for the T2 imaging device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results with acceptance criteria.

    Therefore, many of the requested details regarding acceptance criteria for clinical performance, sample sizes, expert qualifications, and ground truth establishment from a clinical trial perspective are not available in this document. The document explicitly states: "No clinical studies are submitted."

    Here's a breakdown of the available information based on the prompt's questions:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific acceptance criteria or reported device performance in terms of clinical accuracy (e.g., sensitivity, specificity) against a clinical ground truth. Instead, the substantial equivalence hinges on non-clinical performance tests demonstrating compliance with relevant standards and technical specifications being similar to the predicate device.

    The closest to "acceptance criteria" are the performance specifications listed in the "Substantial Equivalence Matrix" (pages 4-5) which compare the proposed device (T2) to the predicate device (T1-CS, T1-C). These are primarily technical specifications rather than diagnostic performance metrics.

    FeatureAcceptance Criteria (Predicate's Performance)Reported Device Performance (T2)Outcome
    Device NameT1-CS, T1-CT2Different
    510(k) No.K183475ProposedDifferent
    ManufacturerOsstem Implant Co., Ltd.Osstem Implant Co., Ltd.Same
    Indications for UseSimilar description for imaging oral anatomySimilar description for imaging oral anatomySame
    Performance SpecificationPanoramic, cephalometric and computed tomographyPanoramic, cephalometric and computed tomographySame
    Input Voltage100-240 VAC100-240VACSame
    Tube Voltage60-100 kV60-95KVSimilar (included within range of predicated product)
    Tube Current5-16 mA5-11mASimilar (included within range of predicated product)
    Focal Spot Size0.5 mm0.5mmSame
    Exposure TimeMax. 22 sMax. 22sSame
    2D Image Viewing ProgramModel Name: 2D01, 2D02Model Name: 2D02Similar (included within predicated product)
    3D Image Viewing ProgramModel Name: 3D01, 3D02Model Name: 3D02Similar (included within predicated product)
    Anatomical SitesMaxillofacialMaxillofacialSame
    Image Sensor (Computed Tomography)1515DXT, SEN11515DXT, SEN1Same
    Image Sensor (Panoramic)1515DXT, SEN11515DXT, SEN1Same
    Image Sensor (Cephalometric)XID-C24DCXID-C24DCSame
    Size of Imaging Volume (CT)1515DXT: Max. 15x9, SEN1: Max 15x10 (Stitching Mode: Max 15 X 15)1515DXT: Max. 15x9, SEN1: Max 15x10Similar (included within range of predicated product)
    Pixel Resolution (CT)1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm (-)Same (The listed values for T2 are the high-resolution modes of predicate)
    Pixel Resolution (Panoramic)1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mm1515DXT: 3.94 lp/mm, SEN1: 4.0 lp/mmSame
    Pixel Resolution (Cephalometric)XID-C24DC: 5 lp/mmXID-C24DC: 5 lp/mmSame
    Pixel Size (CT)1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μm1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μmSimilar (included within range of predicated product)
    Pixel Size (Panoramic)1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μm1515DXT: 127 μm * 127 μm, SEN1: 125 μm * 125 μmSame
    Pixel Size (Cephalometric)XID-C24DC: 100 μm * 100 μmXID-C24DC: 100 μm * 100 μmSame
    Active Area (CT)1515DXT: 146 x 146 mm, SEN1: 160 x 128 mm1515DXT: 146 x 146 mm, SEN1: 160 x 128 mmSame
    Active Area (Pano)1515DXT: 146 x 8.13 mm, SEN1: 160 x 8 mm1515DXT: 146 x 8.13 mm, SEN1: 160 x 8 mmSame
    Active Area (Ceph)1.8 x 150 mm1.8 x 150 mmSame

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as no clinical studies were submitted. Non-clinical performance tests were conducted, but details on the "test set" for those are not provided in terms of origin or type (e.g., phantom data).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical studies were submitted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical studies were submitted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The T2 device is an X-ray imaging system, not an AI-assisted diagnostic tool. No clinical studies, let alone MRMC studies, were reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The T2 device is an X-ray imaging system, not an AI algorithm. Its performance is related to image acquisition and reconstruction, not autonomous diagnostic interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical studies were submitted. For the non-clinical performance, the "ground truth" would be established by physical measurements and adherence to technical standards.

    8. The sample size for the training set

    Not applicable. This device is a hardware imaging system, not a machine learning algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable. This device is a hardware imaging system, not a machine learning algorithm.