Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211319

    Validate with FDA (Live)

    Date Cleared
    2021-08-19

    (111 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Purism Non-sterile Powder Free Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The Purism Non-sterile Powder Free Nitrile Examination Gloves are a non-sterile, powder free, single use, disposable item. The gloves are made from nitrile compound and are blue in color. The gloves are ambidextrous and offered in Small, Medium, Large and X-Large sizes.

    AI/ML Overview

    The provided document describes the FDA's 510(k) clearance for the Purism Non-sterile Powder Free Nitrile Examination Gloves. It outlines the non-clinical tests performed to demonstrate that the device meets acceptance criteria.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device PerformancePass/Fail
    ASTM D6319-19 (Dimensions)To determine glove lengthMin Length (mm): Size S: 230, Size M: 230, Size L: 230, Size XL: 230Avg Min Length (mm): Size S: 236, Size M: 239, Size L: 241, Size XL: 238Pass
    To determine glove palm widthPalm Width (mm): Size S: 80±10, Size M: 95±10, Size L: 110±10, Size XL: ≥110Avg Palm Width (mm): Size S: 84, Size M: 95, Size L: 103, Size XL: 111Pass
    To determine glove finger thicknessFinger Thickness (mm): Size S: 0.100±0.02, Size M: 0.100±0.02, Size L: 0.100±0.02, Size XL: 0.100±0.02Avg Finger Thickness (mm): Size S: 0.096, Size M: 0.111, Size L: 0.108, Size XL: 0.111Pass
    To determine glove palm thicknessPalm Thickness (mm): Size S: 0.060±0.02, Size M: 0.060±0.02, Size L: 0.060±0.02, Size XL: 0.060±0.02Avg Palm Thickness (mm): Size S: 0.058, Size M: 0.066, Size L: 0.074, Size XL: 0.080Pass
    ASTM D6319-19 (Physical Properties)To determine glove physical properties before and after ageingBefore Ageing: Tensile Strength min 14 MPa, Ultimate Elongation min 500%Before Ageing: Tensile Strength min 18.0 MPa, Ultimate Elongation min 694%Pass
    After Ageing: Tensile Strength min 14 MPa, Ultimate Elongation min 400%After Ageing: Tensile Strength min 18.2 MPa, Ultimate Elongation min 664%Pass
    ASTM D5151-19To detect holes and check for water leaksAs per ISO 2859-1. GI Level II, Lot Size 3000, Sample Size 125, AQL 1.5, Ac=5, Re=6Pass at AQL 1.5Pass
    ASTM D6124To determine residual powder≤2 mg/glove0.06 mg/glovePass
    ISO 10993-10:2010 (Primary Skin Irritation)To determine if device is a skin irritantThe device must be a non-irritantUnder the study conditions not an irritantPass
    ISO 10993-10:2010 (Dermal Sensitization)To determine if device is a dermal sensitizerThe device must be a non-sensitizerUnder the study conditions not a sensitizerPass
    ISO 10993-5:2009 (In vitro Cytotoxicity)To determine if device extract is cytotoxicThe device must be non-cytotoxicUnder the study conditions cytotoxicFail
    ISO 10993-11:2017 (Acute Systemic Toxicity)To determine if device induces systemic toxicityThe device must not induce systemic toxicityUnder the study conditions did not induce systemic toxicityPass

    Note on Cytotoxicity: While the device showed cytotoxic properties in the in vitro test, the document states that additional acute systemic toxicity testing (ISO 10993-11:2017) supported that the device was not systemically toxic. The primary skin irritation and dermal sensitization tests also indicated the device was not an irritant or skin sensitizer. This suggests that the manufacturer provided sufficient evidence to mitigate concerns arising from the in vitro cytotoxicity result in the context of the device's intended use.

    2. Sample size used for the test set and the data provenance

    The document specifies "Sample Size 125" for the ASTM D5151-19 (Freedom from Holes) test, which is based on ISO 2859-1, GI Level II, and Lot Size 3000. For other tests like dimensions, physical properties, and biocompatibility, specific sample sizes are not explicitly stated in this summary, but the results are reported as average values, indicating that multiple samples were tested.

    The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated for each test. However, the manufacturer is "Dezhou Purism Medical Technology Co., Ltd." located in China, suggesting the testing was likely conducted in or overseen from China. The studies are non-clinical (laboratory-based) tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The tests performed are objective, standardized laboratory tests according to ASTM and ISO standards for medical gloves. They do not involve expert interpretation or ground truth establishment in the way clinical studies with human assessors would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As these are objective laboratory tests, there is no adjudication method involved. The results are quantitative measurements against predefined criteria.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a nitrile examination glove, not an AI-powered diagnostic tool, and therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical glove. The performance evaluation is based on its material properties and manufacturing quality, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance tests conducted (dimensions, physical properties, freedom from holes, residual powder, biocompatibility) is established by adherence to recognized international standards (ASTM D6319-19, ASTM D5151-19, ASTM D6124, ISO 10993-10:2010, ISO 10993-5:2009, ISO 10993-11:2017) and their specified acceptance criteria. These standards define the measurable properties that a glove must meet for its intended use.

    8. The sample size for the training set

    Not applicable. This device is a physical product and does not involve AI or machine learning, thus there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1