Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211225

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-28

    (96 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MIS Ti-base abutment is a titanium base placed onto MIS dental implants to provide support for customized cement-retained or screw retained single or multiple-unit restorations. It is used with a digitally designed mesostructure. MIS Ti-base and the mesostructure make up a two- piece abutment used in conjunction with MIS dental implants, to be placed in the upper or lower jaw arches, in order to restore masticatory function.

    Narrow platform Ti-bases are indicated for use only in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws.

    MIS short implants are to be used only with straight abutments. Mesostructures for use with the MIS Tibase abutment are to be made from inCoris ZI, IPS e.max® CAD Abutment or VITA ENAMIC® (IS), designed and manufactured using Sirona CEREC SW version 4.6.1 software.

    MIS Ti-base abutments are intended for use with the following MIS implants:

    • C1 conical connection implant system
    • V3 conical connection implant system
    • SEVEN internal hex implant system
    • M4 internal hex implant system
    • Lance+ internal hex implant system
    Device Description

    The subject MIS Ti-base abutments are endosseous dental implant abutments intended to be connected to MIS dental implants and used to support CAD/CAM customized cement-retained or screw retained single or multiple-unit restorations.

    MIS Ti-base abutments consist of a titanium base and a prosthetic screw, both made of TI-6AI-4V ELI complying with ASTM F136. The prosthetic screw tightens the finished CAD/CAM abutment to the dental implant.

    MIS Ti-base abutments are the bottom-half/base of a two-piece custom ceramic-titanium abutment consisting of a ceramic coping/mesostructure and a titanium base.

    The top-half custom ceramic coping/mesostructure or crown is intended to be fabricated from Sirona inCoris ZI zirconium oxide ceramic block, IPS e.max® CAD ceramic block, or from IPS e.max® CAD ceramic block or VITA ENAMIC® (IS) ceramic block and designed and milled using Sirona chairside Dental CAD/CAM System, with software version: CEREC SW version 4.6.1. The mesostructure design will be subject to the Sirona system controls, such as: A maximum angulation of 20° and minimum wall thickness of 0.5mm for inCoris ZI and e.max materials and 0.8mm for VITA ENAMIC material.

    It is not permitted to reduce the Ti-base's diameter, shorten the Ti-base or modify its implant-abutment connection and emergence profile in any way.

    The subject, pre-fabricated titanium base abutment is designed with interface compatibility to specific MIS dental implant systems. The subject MIS Ti-base abutments are MIS connection and internal hex connection Ti-base abutments, and their connection is compatible with MIS conical connection C1 and V3 implants, and MIS SEVEN, M4 and Lance+ internal hex implants, which are not subject to this submission and were previously cleared.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the MIS Ti-base abutment. This is an FDA submission for devices that are "substantially equivalent" to predicate devices, meaning they have the same intended use and similar technological characteristics. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are typically demonstrating this substantial equivalence through non-clinical performance data, rather than a clinical trial with specific performance metrics like sensitivity or specificity for an AI algorithm.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission for a physical medical device (dental abutment) and not an AI algorithm, the "acceptance criteria" are related to demonstrating that the device performs as safely and effectively as a legally marketed predicate device. The performance is assessed through non-clinical testing against established standards.

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance (Summary of Non-clinical Testing)
    BiocompatibilityThe subject device is manufactured using identical methods, facility, and raw material as the predicate (K191152). Patient contact duration and type are the same. IPS e.max® and VITA ENAMIC® (IS) mesostructure materials are cleared under K191382 and K153645 respectively, and no modifications to these materials are included. No new biocompatibility testing was required.
    Fatigue Testing (Mechanical Performance)Withstand 2,000,000 cycles without failure at a substantially equivalent load to the cited predicates, as per ISO 14801:2016. Worst-case abutments (narrowest, 20° mesostructure, various materials) were tested.
    Sterilization TestingValidated steam sterilization parameters (gravity displacement and pre-vacuum) for inCoris ZI mesostructure according to ANSI/AAMI/ISO 17665-1:2006 and ANSI/AAMI/ISO 17665-2:2009. For IPS e.max and VITA ENAMIC (IS) mesostructures, recommended parameters are based on predicate clearances K191382 and K153645.
    Software Verification and Validation (CAD/CAM Integration)Software verification and validation testing for the abutment design library demonstrated that restrictions prevent design of the mesostructure component outside of design limitations. The encrypted design library was validated to ensure established limitations are locked and cannot be modified. (For CEREC SW version 4.6.1).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a "test set" in the context of clinical data for AI. For the non-clinical fatigue testing, "worst case abutments" were chosen. The exact number of samples tested for each worst-case configuration (e.g., number of narrowest abutments, number of specific mesostructures) is not explicitly stated but implied to be sufficient for ISO 14801:2016 compliance.
    • Data Provenance: Not applicable in the context of clinical or retrospective data for an AI algorithm. The performance data comes from laboratory non-clinical tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This is not applicable as this is a non-clinical device submission for a physical component, not an AI algorithm requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    • Not applicable for a non-clinical device submission.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical dental abutment. The software mentioned (CEREC SW version 4.6.1) is for design and milling, not a standalone AI diagnostic algorithm.

    7. The Type of Ground Truth Used

    • For fatigue testing, the "ground truth" is defined by the mechanical strength and durability requirements of the ISO 14801:2016 standard, ensuring the device can withstand chewing forces.
    • For biocompatibility, the "ground truth" is adherence to ISO 10993-1 and prior clearances of materials used.
    • For sterilization, the "ground truth" is effective sterilization as demonstrated by ANSI/AAMI/ISO 17665 standards.
    • For software verification, the "ground truth" is the established design limitations and the software's ability to enforce them.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI algorithm that requires a training set. The software mentioned (CEREC SW) is a CAD/CAM design software, not a machine learning model developed with a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable as there is no training set for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1