    Device Name
    InnovaMatrix FS
    Date Cleared
    2021-04-21

    (54 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InnovaMatrix™ FS is indicated for the management of wounds including: partial- and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled / undermined wounds, surgical wounds (donor sites/grafts, post Mohs surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, second-degree burns and skin tears) and draining wounds.

    The device is intended for one-time use.

    Device Description

    InnovaMatrix™ FS is a decellularized extracellular matrix (ECM) topical wound covering derived from porcine placental tissue. Triad processes the tissue into the ECM topical wound covering. InnovaMatrix™ FS is composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans. The wound dressing is provided in fenestrated sheets that are approximately 40-100 microns thick in sizes ranging from 2 x 2cm to 5 x 5cm. They are provided as single-use, sterile wound coverings.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called InnovaMatrix™ FS. The document describes the device, its indications for use, and a comparison to a predicate device. However, it does not contain specific acceptance criteria, a scientific study detailing device performance against those criteria, or information on specific statistical metrics (e.g., sensitivity, specificity, AUC) normally associated with AI/algorithm performance studies.

    Instead, this document focuses on demonstrating substantial equivalence to a previously cleared device (InnovaMatrix™ K193552) and mentions verification testing to ensure certain characteristics. The primary change in the InnovaMatrix™ FS is the addition of fenestrations.

    Therefore, many of the requested details about acceptance criteria, study design, expert involvement, and ground truth for AI performance are not available in this regulatory submission.

    Here's what can be extracted based on the provided text, while explicitly noting what is not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    Device Function/CharacteristicAcceptance CriteriaReported Device Performance (Summary)
    Material CompositionImplied: Same as predicate (InnovaMatrix™)Composed of collagen, elastin, laminin, fibronectin, hyaluronic acid and sulfated glycosaminoglycans (same as predicate)
    Manufacturing ProcessImplied: Same as predicateUndergoes the same manufacturing processing steps as predicate
    BiocompatibilityImplied: Same as predicateExisting biocompatibility data for InnovaMatrix™ (predicate) is applicable.
    Shelf LifeImplied: Same as predicateExisting shelf life data for InnovaMatrix™ (predicate) is applicable.
    SterilizationImplied: Same as predicateExisting sterilization data for InnovaMatrix™ (predicate) is applicable. Device is E-Beam-sterilized.
    Human Repeat Insult Patch TestingImplied: Same as predicateExisting data for InnovaMatrix™ (predicate) is applicable.
    Human Skin Prick TestingImplied: Same as predicateExisting data for InnovaMatrix™ (predicate) is applicable.
    BioburdenWithin specified values determined by risk assessmentVerification testing confirmed bioburden was within specified values.
    EndotoxinWithin specified values determined by risk assessmentVerification testing confirmed endotoxin was within specified values.
    Water Absorption CapacityWithin specified values determined by risk assessmentVerification testing confirmed water absorption capacity was within specified values.
    Fenestrations (Design Feature)Meet pre-determined acceptance criterion for verification activity.Design Controls demonstrates that the modified device met the pre-determined acceptance criterion for the verification activity.

    Missing Information: Specific numerical values or ranges for "specified values" (e.g., for bioburden, endotoxin, water absorption capacity) are not provided in this document. The "acceptance criterion for the verification activity" for fenestrations is also not detailed.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes verification testing and design controls for the modified device (InnovaMatrix™ FS), not a clinical or AI performance study with a "test set" in the typical sense.

    • Sample Size for Test Set: Not applicable in the context of an AI/algorithm performance study. The document mentions "verification testing" for characteristics like bioburden, endotoxin, and water absorption, but does not provide sample sizes for these tests.
    • Data Provenance: Not applicable for an AI/algorithm performance study. The data for verification testing would likely be from laboratory tests performed by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is not an AI/algorithm performance study that relies on expert interpretation to establish ground truth for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of an adjudication method as this is not an AI/algorithm performance study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This submission is for a collagen wound dressing, not an AI software device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone (algorithm only) performance study was not done. InnovaMatrix™ FS is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to established laboratory standards and specifications for material characteristics (bioburden, endotoxin, water absorption capacity) and manufacturing/design output (fenestrations). It is not pathology, outcomes data, or expert consensus in an observational or diagnostic sense. The ground truth for proving substantial equivalence relies on comparison to the predicate device and established safety/performance parameters for similar products.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no "training set" or corresponding ground truth establishment methodology.