    Device Name
    Qitexio
    Manufacturer
    Date Cleared
    2022-03-14

    (447 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Qitexio Luer Lock Syringes are manual devices used to inject fluids into the body. They are also used for delivery of Lipiodol (Ethiodized Oil) Injection.

    Device Description

    Medex, a subsidiary of Guerbet Group, designed and manufactures the Qitexio® Luer Lock Syringe, a disposable, handheld syringe available in 3 different volumes (1, 3 and 20 mL). Oitexio® Syringes are standard syringes which are verified to endure the potentially damaging chemical effects of Lipiodol® (Ethiodized Oil), an oil-based contrast media manufactured by Guerbet.

    Qitexio® Luer Lock Syringes are manual devices used to inject fluids into the body. Qitexio® Syringes are also used for delivery of Lipiodol® (Ethiodized Oil) Injection.

    They are single-use devices, sterilized by gamma radiation.

    Standard five-piece piston syringe constructed with a clear hollow barrel into which is inserted a closely fitting movable plunger, O-Ring, rotative finger flange, and backstop. Fitting offered with male luer lock.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Qitexio Luer Lock Syringes, a medical device. The information provided heavily focuses on proving the substantial equivalence of the new device to a predicate device, the Merit Syringe (K173601).

    The acceptance criteria for this type of device (a piston syringe) are primarily defined by conformity to established FDA recognized consensus standards and a battery of performance, biocompatibility, and sterilization tests. The document does not describe an AI/ML-driven device or study. Therefore, some of the requested information regarding AI/ML study design, such as MRMC studies, ground truth establishment for training sets, and expert adjudication, are not applicable to this submission.

    Here's a breakdown of the acceptance criteria and proof, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    For the Qitexio Luer Lock Syringes, acceptance criteria are generally met by demonstrating conformity to recognized consensus standards and by passing specific performance tests. The document repeatedly states that the device "complies with" or "met the established acceptance criteria" for these standards and tests. Specific numerical performance results are not provided in this summary, as is common for 510(k) summaries which focus on demonstrating equivalence rather than presenting raw data.

    Acceptance Criteria CategorySpecific Criteria (Standard/Test)Reported Device Performance
    Performance TestingISO 7886-1 Sterile Hypodermic Syringes For Single Use - Part 1: Syringes For Manual UseComplies; specific for Lipiodol® (Ethiodized Oil) Injection delivery.
    ISO 80369-7 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applicationsComplies.
    Functional performance testing (with preconditioning: double sterilization, accelerated/real-time aging)Met established criteria; performs as intended.
    BiocompatibilityISO 10993 Biological Evaluation of Medical Devices (Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Hemolysis, Material-Mediated Pyrogenicity, Chemical Characterization)Non-toxic, biocompatible for intended use; all tests passed.
    USP <788> Particulate Matter in Injections (Method 1 Light Obscuration Particle Count)Met USP <788> 1B acceptance criteria (for ≤ 100 mL volume).
    SterilizationISO 11137-1:2006 (Sterilization, Validation, and Routine Control of a Sterilization Process for Medical Devices)Validated sterilization cycle; SAL 10-6 achieved; dose 25-50 kGy.
    ISO 11137-2:2013 (Establishing the Sterilization Dose)Determined 25 kGy requirement.
    USP 39 -NF 34 and European Pharmacopoeia (9th Edition) requirements for Material-mediated PyrogenicityConfirmed non-pyrogenicity (Rabbit Pyrogen study).
    Limulus Amebocyte Lysate (LAL) test for Bacterial endotoxinsMet criteria: < 20 endotoxin units. Device non-pyrogenic.
    Package IntegrityASTM F 1886/F 1886 M (Visual Inspection)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ISO 11607-1 (Peel Characteristics, Peel Test)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ASTM F 88/F 88 M (Peel Test)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ASTM F 1929 (Dye Test)Passed (after max dose sterilization).
    ASTM F 2096 (Bubble test)Passed (after max dose + ISTA 3A Simulated Transportation).
    Shelf-LifeDevice functionality and package integrity after 3 years aging (real-time or accelerated)Acceptable results supporting a 3-year shelf-life.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of syringes) used for each individual test (e.g., performance, biocompatibility). It mentions "samples" or "representative device samples" were used.

    • Data Provenance: Not explicitly stated regarding country of origin for the test data. However, the manufacturer is Medex, a subsidiary of Guerbet Group, located in France. The testing was conducted to comply with international standards (ISO, ASTM, USP, European Pharmacopoeia) recognized by the FDA for market clearance in the US.
    • Retrospective or Prospective: The testing described is prospective, conducted specifically to gain regulatory clearance for the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not Applicable. This submission is for a physical medical device (Piston Syringe), not an AI/ML diagnostic tool requiring expert interpretation as ground truth for a test set. The "ground truth" for this device's performance is established by objective measurements against engineering and biological standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. As above, this is not an AI/ML diagnostic tool requiring human adjudication of subjective interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. This type of study is typically performed for imaging or diagnostic AI/ML devices where the AI assists human readers. It is not relevant for a Piston Syringe.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This is not an algorithm or software device. The "performance" is that of the physical syringe itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance and safety is established through:

    • Conformity to recognized consensus standards (e.g., ISO 7886-1 for syringe function, ISO 10993 for biocompatibility, ISO 11137 for sterilization).
    • Objective bench testing measurements (e.g., particulate counts, endotoxin levels, peel force, functional performance like fluid injection).
    • Biological assay results (e.g., cytotoxicity, irritation, pyrogenicity).
    • Material characterization to ensure compatibility with intended fluids (e.g. Lipiodol).

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/ML device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As this is not an AI/ML device, there is no training set or associated ground truth establishment. Device design and manufacturing processes are validated through engineering principles and quality system regulations (21 CFR Part 820).