Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203597

    Validate with FDA (Live)

    Device Name
    Cholesterol2
    Date Cleared
    2022-06-30

    (568 days)

    Product Code
    Regulation Number
    862.1175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cholesterol2 assay is used for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. The Cholesterol2 assay is to be used as an aid in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    Device Description

    The Cholesterol2 assay is an automated clinical chemistry assay for the quantitation of cholesterol in human serum or plasma on the ARCHITECT c System. Cholesterol esters are enzymatically hydrolyzed by cholesterol esterase to cholesterol and free fatty acids. Free cholesterol, including that originally present, is then oxidized by cholesterol oxidase to cholest-4-ene-3-one and hydrogen peroxide. The hydrogen peroxide oxidatively couples with N,N-Bis(4-sulfobutyl)-3-methylaniline (TODB) and 4-aminoantipyrine to form a chromophore (quinoneimine dye) which is quantitated at 604 nm.

    AI/ML Overview

    The provided text describes the Abbott Cholesterol2 assay, an in vitro diagnostic device for quantifying cholesterol in human serum or plasma.

    Here's an analysis of the acceptance criteria and study data:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for each performance characteristic. Instead, it presents the results of various studies as proof of device performance. The table below summarizes the reported performance, which implicitly serves as the "met" criteria.

    Performance CharacteristicReported Device Performance (Cholesterol2)
    Analytical Measuring Interval (AMI)5-748 mg/dL
    Extended Measuring Interval (EMI)748-2992 mg/dL
    Reportable Interval2-2992 mg/dL
    Precision
    Control Level 1 (251 mg/dL)SD: 1.9 mg/dL (Within-Run), 2.6-3.1 mg/dL (Within-Laboratory); %CV: 0.7% (Within-Run), 1.0-1.2% (Within-Laboratory)
    Control Level 2 (106 mg/dL)SD: 1.0 mg/dL (Within-Run), 1.3-1.7 mg/dL (Within-Laboratory); %CV: 1.0% (Within-Run), 1.2-1.6% (Within-Laboratory)
    Panel A (21 mg/dL)SD: 0.6 mg/dL (Within-Run), 0.7-0.8 mg/dL (Within-Laboratory); %CV: 3.0% (Within-Run), 3.2-4.1% (Within-Laboratory)
    Panel B (237 mg/dL)SD: 2.8 mg/dL (Within-Run), 3.7-4.9 mg/dL (Within-Laboratory); %CV: 1.2% (Within-Run), 1.5-2.0% (Within-Laboratory)
    Panel C (718 mg/dL)SD: 6.4 mg/dL (Within-Run), 4.6-6.9 mg/dL (Within-Laboratory); %CV: 0.9% (Within-Run), 0.7-1.0% (Within-Laboratory)
    Limit of Blank (LoB)0 mg/dL
    Limit of Detection (LoD)2 mg/dL
    Limit of Quantitation (LoQ)5 mg/dL (at 20% CV maximum allowable precision)
    LinearityLinear across the analytical measuring interval of 5 to 748 mg/dL
    Interference (Endogenous)
    Conjugated Bilirubin (7 mg/dL)No significant interference (within ± 10%)
    Unconjugated Bilirubin (11 mg/dL)No significant interference (within ± 10%)
    Hemoglobin (1000 mg/dL)No significant interference (within ± 10%)
    Total Protein (15 g/dL)No significant interference (within ± 10%)
    Conjugated Bilirubin (40 mg/dL)Interference: -39% at 150 mg/dL analyte, -31% at 220 mg/dL analyte
    Unconjugated Bilirubin (16 mg/dL)Interference: -11% at 150 mg/dL analyte
    Interference (Exogenous)
    Acetaminophen (160 mg/L)No significant interference
    Acetylcysteine (150 mg/L)No significant interference
    Acetylsalicylic acid (30 mg/L)No significant interference
    Aminoantipyrine (40 mg/L)No significant interference
    Ampicillin-Na (80 mg/L)No significant interference
    Biotin (4250 ng/mL)No significant interference
    Ca-dobesilate (60 mg/L)No significant interference
    Cefotaxime (53 mg/dL)No significant interference
    Cefoxitin (6600 mg/L)No significant interference
    Cyclosporine (2 mg/L)No significant interference
    Desacetylcefotaxime (6 mg/dL)No significant interference
    Dipyrone (100 mg/L)No significant interference
    Dobutamine (0.2 mg/dL)No significant interference
    Doxycycline (20 mg/L)No significant interference
    Ibuprofen (220 mg/L)No significant interference
    Intralipid (1050 mg/dL)No significant interference
    Levodopa (8 mg/L)No significant interference
    Methotrexate (140 mg/dL)No significant interference
    Metronidazole (130 mg/L)No significant interference
    Methylaminoantipyrine (40 mg/L)No significant interference
    Methyldopa (20 mg/L)No significant interference
    N-Acetyl-p-benzoquinone (NAPQI) (20 mg/L)No significant interference
    Phenylbutazone (330 mg/L)No significant interference
    Phenytoin (6 mg/dL)No significant interference
    Rifampicin (50 mg/L)No significant interference
    Sodium heparin (4 U/mL)No significant interference
    Sulpiride (15 mg/L)No significant interference
    Theophylline (60 mg/L)No significant interference
    Ascorbic acid (60 mg/L)Interference: -10% at 150 mg/dL analyte
    Intralipid (2000 mg/dL)Interference: -27% at 150 mg/dL analyte, -22% at 220 mg/dL analyte
    Methyldopa (30 mg/L)Interference: -14% at 150 mg/dL analyte
    Method Comparison (vs. Predicate)
    Serum (n=138)Correlation Coefficient: 1.00; Intercept: 0.41; Slope: 0.98 (Range: 7-684 mg/dL)
    Tube Type SuitabilityAcceptable for Serum, Serum separator, Lithium heparin, Lithium heparin separator, Sodium heparin tubes.
    Dilution VerificationAutomated dilution protocol (1:5.97) and manual dilution procedure (1:4) evaluated. (Performance details not provided in summary).
    TraceabilityTraceable to National Reference System for Cholesterol (Abell-Kendall reference method in a CDC-Certified CRMLN).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Study: 2 controls and 3 human serum panels were tested. Each sample was tested in duplicate, twice per day for 20 days. This means 80 measurements per sample (2 duplicates x 2 times/day x 20 days).
    • Lower Limits of Measurement Study: n ≥ 60 replicates for LoB, LoD, and LoQ determinations. They used low-analyte level samples and zero-analyte samples.
    • Linearity Study: The number of samples for the linearity study is not explicitly stated, but it covered the range of 5 to 748 mg/dL.
    • Interference Studies: Each endogenous substance was tested at 2 analyte levels (approximately 150 mg/dL and 220 mg/dL). Exogenous substances were tested at various specified interferent levels. The number of samples for each interferent is not provided.
    • Method Comparison Study: 138 serum samples were used.
    • Tube Type Study: Samples were collected from a minimum of 40 donors.
    • Dilution Verification: 8 human serum samples.

    Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the context of medical device regulatory submission, these are typically prospective studies conducted in a controlled laboratory environment. The "human serum panels" and "human serum samples" imply human-derived biological samples.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is an in vitro diagnostic (IVD) chemistry assay. The concept of "experts establishing ground truth" as it applies to image interpretation or clinical diagnosis by medical professionals (like radiologists) does not directly apply here in the same way.

    For IVDs like this, the "ground truth" or reference values are established through:

    • Reference methods: The Cholesterol2 reagent is certified to be traceable to the National Reference System for cholesterol, against the Abell-Kendall reference method in a CDC-Certified Cholesterol Reference Method Laboratory Network (CRMLN). The Abell-Kendall method is considered the gold standard for cholesterol measurement.
    • Analytically Validated Methods: For values outside the AMI but within the EMI, samples were "value assigned using the analytically validated method."
    • Known concentrations: For studies like linearity, spiked samples with known concentrations are used.

    Therefore, the "experts" are the methodologists and laboratory professionals overseeing and validating the reference methods and the analytical validation processes. No specific number or qualification of clinical experts (e.g., radiologists) is relevant for establishing the ground truth for a quantitative chemistry assay.

    This is a standalone performance evaluation of the assay itself, demonstrating its analytical accuracy, precision, and robustness.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The concept of "adjudication" (e.g., 2+1, 3+1 where multiple human readers agree or have a tie-breaker by an expert) is not applicable to this type of quantitative diagnostic assay. The results are numerical values generated by the automated instrument and reagents. Deviations or discrepancies would be resolved through re-testing, quality control, or investigation into instrument or reagent issues, rather than human expert adjudication of a "diagnosis."


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging device or an AI assistant for human readers. Its output is a quantitative measurement of cholesterol, not an interpretation that requires human "reading" or decision support.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone performance evaluation of the device (Cholesterol2 assay on the ARCHITECT c8000 System) was done. The studies described (reportable interval, precision, lower limits of measurement, linearity, interference, method comparison, tube type, dilution verification) all evaluate the analytical performance of the assay and instrument directly, without human interpretation as part of the primary outcome measure. The output is a numerical concentration of cholesterol.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for this quantitative assay primarily relies on:

    • Reference Methods: Specifically, the Abell-Kendall reference method, which is considered the gold standard for cholesterol measurement and is used in CDC-Certified Cholesterol Reference Method Laboratory Networks (CRMLN). The device's traceability to this method is explicitly stated.
    • Analytically Validated Methods: For verifying values in the extended measuring interval.
    • Known Spiked Concentrations: For studies such as linearity and dilution verification, where samples are prepared with precisely known concentrations.

    This is an analytical ground truth, not a clinical ground truth derived from pathology or patient outcomes.


    8. The sample size for the training set

    This document does not describe a typical "training set" in the context of machine learning or AI. This is a chemistry assay that uses reagents and enzymatic reactions, not an algorithm that is "trained" on data. Therefore, the concept of a training set as used in AI development is not applicable here. The assay's analytical characteristics are determined through standard laboratory validation studies.


    9. How the ground truth for the training set was established

    As explained above, there is no "training set" in the AI sense for this device. The analytical accuracy and reliability are established through comparisons to certified reference methods and known standard concentrations, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1