Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202995

    Validate with FDA (Live)

    Date Cleared
    2021-02-19

    (141 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CareView 3600RF detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purposic procedures. This product is not intended for mammography applications.

    Device Description

    The CareView 3600RF detector is a class of radiography X-ray flat panel detector that has an imaging area of 867.5mm×433.1mm. The detector communicates by a wired connection (Giga-bit Ethernet communication mode).

    The detector functions by intercepting X-ray photons. Then the scintillator emits visible spectrum photons that illuminate an array of photo detectors (a-Si) that create electrical signals. The electrical signals are then digitally converted to display an image on the monitor.

    The detector should be connected to a computer and X-ray generator to digitize X-ray images and transfer radiography diagnostics. The x-ray generator, an essential part of a full x-ray system, is not part of the subject medical device.

    AI/ML Overview

    The information provided indicates that the CareRay Digital Medical Technology Co., Ltd. CareView 3600RF X-ray Flat Panel Detector is seeking market clearance through a 510(k) submission, asserting substantial equivalence to the iRay Technology (Shanghai) Ltd. Mars1417V-PSI Wireless Digital Flat Panel Detector (K161730). Since this is a 510(k) submission for a traditional medical device (X-ray Flat Panel Detector) and not a software-as-a-medical-device (SaMD) or AI/ML-driven device with clinical performance endpoints based on diagnostic accuracy, the acceptance criteria and study detailed primarily focus on technical performance metrics rather than clinical efficacy involving human readers or AI algorithms.

    Here's a breakdown of the available information regarding acceptance criteria and performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an X-ray Flat Panel Detector, the "acceptance criteria" and "reported device performance" are typically defined by various technical specifications and physical performance metrics, often compared against a predicate device and industry standards. The submission highlights several of these:

    Performance MetricAcceptance Criteria (Predicate Device K161730)Reported Device Performance (CareView 3600RF)
    X-ray AbsorberGOS ScintillatorGOS Scintillator
    Readout MechanismThin Film TransistorThin Film Transistor
    Image Matrix Size2304 x 2800 pixels5632 x 2816 pixels
    Pixel Pitch150 µm154 µm
    Effective Imaging Area355 mm x 434 mm867.5 mm x 433.1 mm
    Grayscale14 bit16 bit
    Spatial Resolution3.4 lp/mm3.4 lp/mm
    MTF (@ 0.5 lp/mm)0.75 (75%)75%
    MTF (@ 1 lp/mm)(Not specified explicitly, but generally expected to be comparable based on spatial resolution)50%
    DQE (@ RQA5, 3.2 µGy, 0.5 lp/mm)0.27 (27%)28%
    DQE (@ RQA5, 3.2 µGy, 1 lp/mm)(Not specified explicitly)20%
    CommunicationsWired: Gigabit Ethernet, Wireless: IEEE 802.11a/b/g/n (2.4 GHz/ 5 Glz)Gigabit Ethernet (Wired only)
    Imaging PlateCarbon Fiber PlateCarbon Fiber Plate
    CoolingAir coolingAir cooling
    Operation Temperature+5 ~ +35°C+10 ~ +40°C
    Operation Humidity30 ~ 75% RH (Non-Condensing)30 ~ 75% RH (Non-Condensing)
    Operation Atmospheric Pressure700 ~ 1060 hPa700 ~ 1060 hPa
    Operation AltitudeMax. 3000 metersMax. 3000 meters
    Storage/Transport Temperature-20 ~ +55°C-20 ~ +55°C
    Storage/Transport Humidity10 ~ 90% RH (Non-Condensing)10 ~ 90% RH (Non-Condensing)
    Storage/Transport Atmospheric Pressure700 ~ 1060 hPa700 ~ 1060 hPa
    Storage/Transport Altitude(Not specified explicitly for predicate)Max. 3000 meters
    Electrical Safety & EMCCompliance with IEC/ES 60601-1, IEC/EN 60601-1-2All test results satisfactory

    The acceptance criteria here are implicitly drawn from the predicate device's characteristics and the relevant performance standards for X-ray detectors. The "reported device performance" are the measurements taken from the CareView 3600RF. The statement "All test results are satisfactory" for electrical safety and EMC indicates compliance with relevant standards as the acceptance criteria.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical studies that evaluated the device's technical performance. These studies typically use physical phantoms or test objects rather than patient data. Therefore, the concept of "sample size" in terms of patient images or clinical cases, and "data provenance" (country of origin, retrospective/prospective) as it applies to clinical studies, is not directly applicable here. The "test set" would consist of various physical measurements performed on the detector itself, following standardized protocols (e.g., those for DQE, MTF, NPS).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the evaluation is based on technical, physical performance characteristics of the X-ray detector (e.g., DQE, MTF), not on diagnostic interpretation of medical images by human experts. The "ground truth" for these technical metrics is established by physical measurement methods and calibrated equipment, compared against defined industry standards.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used in clinical studies where expert disagreement on ground truth needs resolution.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was performed. This submission is for an X-ray Flat Panel Detector, which is a hardware component for image acquisition, not an AI-driven image analysis or diagnostic assistance software. Therefore, the concept of "human readers improving with AI vs without AI assistance" does not apply to this device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    There is no indication of a standalone algorithm performance study. The device is a hardware detector, and the software mentioned (CareRay software, APIs) is for controlling the detector and transferring images, not for independent diagnostic interpretation or AI-driven analysis.

    7. The type of ground truth used

    For the non-clinical studies mentioned (Detective quantum efficiency (DQE), Modulation transfer function (MTF), Noise power spectrum (NPS), Spatial resolution, etc.), the "ground truth" is based on objective physical measurements and standardized phantom evaluations. These measurements assess the intrinsic performance characteristics of the detector itself, following recognized international and national standards for medical imaging devices.

    8. The sample size for the training set

    This information is not applicable. The CareView 3600RF is an X-ray Flat Panel Detector, a hardware device. It is not an AI/ML algorithm that requires a "training set" of data in the conventional sense of machine learning.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1