510(k) Data Aggregation

    K Number
    K202556

    Validate with FDA (Live)

    Date Cleared
    2020-10-30

    (57 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

    It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, clinics and similar care environment for clinical diagnosis of patients.

    Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.

    Device Description

    The HS30 is a general purpose, mobile, software controlled, diagnostic ultrasound system, Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, Elastoscan Mode or as a combination of these modes. The HS30 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS30 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    AI/ML Overview

    This document describes the Samsung Medison Co., LTD's HS30 Diagnostic Ultrasound System, which is a general purpose ultrasound system for various clinical applications. The document focuses on showing substantial equivalence to previously cleared predicate devices (HS30 Diagnostic Ultrasound System K182632 and HS40 Diagnostic Ultrasound System K200339).

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, the "acceptance criteria" are demonstrating substantial equivalence to the predicate devices in terms of intended use, imaging capabilities, technological characteristics, safety, and effectiveness. The reported device performance is presented as being conformant to various applicable medical device safety standards.

    Acceptance CriteriaReported Device Performance
    Intended Use consistency with predicate devicesConsistent with predicate devices (Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel applications).
    Imaging Capabilities consistency (Modes of Operation, Scanhead Types, Scanhead Frequency, Features)Largely consistent or expanded upon predicate devices. The HS30 (under review) explicitly lists "3D/4D mode" as a mode of operation, which was listed for the HS40 but not the primary HS30 predicate. Some transducer types and biopsy guides vary slightly but are generally comparable.
    Technological Characteristics (Acoustic Output Display, System Characteristics) consistencyConsistent with predicate devices (e.g., MI/TI output display, mobile cart, LED monitor, 256 gray shades, power requirements).
    Safety & Effectiveness Demonstrated by Compliance to StandardsConforms with applicable medical device safety standards including IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasonic diagnostic equipment safety), ISO 10993-1 (biocompatibility), ISO 14971 (risk management), and NEMA UD 2-2004 (acoustic output measurement).

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "The subject of this premarket submission, HS30 is not required clinical studies to support substantial equivalence."

    Therefore, no test set, sample size, or data provenance information from clinical studies is provided for the device for the purpose of this 510(k) submission. The evaluation was non-clinical, focusing on compliance with standards and comparison to predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    As no clinical studies were required, there is no mention of experts establishing a ground truth for a test set. The assessment relied on non-clinical evaluations and technical comparisons.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used for this submission.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission does not involve an AI-assisted diagnostic device or MRMC studies. The device is a diagnostic ultrasound system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a hardware and software diagnostic ultrasound system, not a standalone algorithm.

    7. The Type of Ground Truth Used

    For this 510(k) submission, the "ground truth" for proving substantial equivalence was primarily:

    • Compliance with recognized industry standards: IEC and ISO standards for safety, electromagnetic compatibility, acoustic output, and biocompatibility.
    • Comparison to legally marketed predicate devices: The functional specifications and intended uses of the HS30 (under review) were compared directly to already cleared HS30 (K182632) and HS40 (K200339) systems.

    8. The Sample Size for the Training Set

    Not applicable. Since no clinical studies were required for this submission, there is no mention of a training set for an algorithm development.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set was identified in this submission.