Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201541

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-04

    (178 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    2 - 60
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FlowTriever2 Catheter is indicated for:

    • · The non-surgical removal of emboli and thrombi from peripheral blood vessels.
    • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

    The FlowTriever2 Catheter is intended for use in the peripheral vasculature.

    Device Description

    The FlowTriever Retrieval/Aspiration System is a single-use over-the-wire catheter-based system for the minimally invasive treatment of thromboemboli in the peripheral vasculature. The system is comprised of two main components packaged separately:
    • Triever Catheters (available in 3 sizes: 16, 20, and 24 Fr)
    • FlowTriever2 Catheter

    The FlowTriever2 Catheter is inserted through the Triever Catheter and advanced to the thrombus. Self-expanding disk(s) are deployed to engage thrombus by retracting the outer delivery catheter. The FlowTriever2 Catheter is retracted into the Triever Catheter to capture the targeted thrombus. Additional clot may be removed by aspiration with the provided 60 cc VacLok Vacuum syringe. After the procedure is complete, the Triever Catheter and FlowTriever2 Catheter are removed from the patient.

    AI/ML Overview

    The provided text does not describe a study involving an AI/Machine Learning device or any associated acceptance criteria for such a device. The document is an FDA 510(k) clearance letter for the FlowTriever Retrieval/Aspiration System, a medical device used for the non-surgical removal of emboli and thrombi from peripheral blood vessels.

    The acceptance criteria and study described in the document relate to the mechanical and functional performance of the medical device itself, not to the performance of an AI algorithm.

    Therefore, I cannot provide the requested information regarding AI device acceptance criteria, study details, sample sizes, expert involvement, or MRMC studies. The document only mentions "Non-Clinical Testing" and "Animal testing" that met predetermined acceptance criteria for the physical device.

    Here's what can be extracted from the document regarding the device's acceptance:

    The "Non-Clinical Testing" section describes the verification and validation tests performed to support the substantial equivalence of the modified FlowTriever Retrieval/Aspiration System. These tests demonstrated compliance with relevant product specifications and met predetermined acceptance criteria.

    A table of the acceptance criteria (tests performed) and the reported device performance (met criteria) is as follows:

    Acceptance Criteria (Test Performed)Reported Device Performance
    Visual & Dimensional InspectionMet predetermined acceptance criteria (implied by "all acceptance criteria were met")
    Guidewire Compatibility VerificationMet predetermined acceptance criteria
    Radial Expansion ForceMet predetermined acceptance criteria
    Deployment ForceMet predetermined acceptance criteria
    Retraction ForceMet predetermined acceptance criteria
    Element Visual Inspection; Post Retraction CyclesMet predetermined acceptance criteria
    Leak and Vacuum Test; Post Retraction CyclesMet predetermined acceptance criteria
    Kink ResistanceMet predetermined acceptance criteria
    Clot Analog Burden Removal ValidationMet predetermined acceptance criteria
    Chronic Clot Analog Burden Removal ValidationMet predetermined acceptance criteria
    Simulated Use, Track and TensileMet predetermined acceptance criteria
    Liquid Leakage under PressureMet predetermined acceptance criteria
    Burst PressureMet predetermined acceptance criteria
    Simulated Use, TorqueMet predetermined acceptance criteria
    Corrosion ResistanceMet predetermined acceptance criteria
    Particulate testingMet predetermined acceptance criteria

    Regarding the other requested information for an AI study:

    • Sample sized used for the test set and the data provenance: Not applicable. The document discusses in vitro and animal testing for a physical device, not an AI test set.
    • Number of experts used to establish the ground truth... and qualifications: Not applicable. Ground truth for a physical device's mechanical performance is established through engineering specifications and quantifiable measurements, not expert consensus on image interpretation.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done: No, not for an AI device. The document states "Clinical testing was not required for the determination of substantial equivalence" for this physical device.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used: For this physical device, the ground truth is based on engineering specifications, physical measurements, and performance benchmarks in controlled environments (e.g., in vitro and animal models).
    • The sample size for the training set: Not applicable for this type of device submission.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1