Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200121

    Validate with FDA (Live)

    Date Cleared
    2020-04-18

    (88 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Route 92 Medical 088 Access System, 110 cm, is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system.

    Device Description

    The Route 92 Medical 088 Access System, 110 cm, is comprised of a Support Catheter and a Delivery Catheter. The distal portion of the Support Catheter is a single-lumen, variable stiffness catheter. Like the predicate device, the proximal portion is a stainless-steel control wire. The Delivery Catheter is a hubbed, single-lumen variable stiffness catheter. Both catheters are hydrophilically coated. The Route 92 Medical Access System, 110 cm, is intended for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovascular system. The devices are provided sterile and non-pyrogenic and are intended for single use only.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Route 92 Medical 088 Access System, 110 cm:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines performance criteria in two main sections: Biocompatibility Testing and Performance Testing. For both, the acceptance criterion is implicitly "PASS" meaning the device met pre-determined criteria for each test. Specific quantitative acceptance criteria are not detailed in this summary.

    Test CategorySpecific TestAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityCytotoxicity – ISO MEM ElutionNon-cytotoxicNon-cytotoxic
    Sensitization – ISO Guinea Pig MaximizationNo sensitization responseNo sensitization response
    Irritation – ISO Intracutaneous ReactivityRequirements metRequirements met
    Acute Systemic Toxicity – ISO Acute Systemic InjectionRequirements metRequirements met
    Pyrogen – Material Mediated PyrogenNon-pyrogenicNon-pyrogenic
    Hemocompatibility – Complement ActivationNo adverse effects expectedNo adverse effects expected
    Hemocompatibility – Partial Thromboplastin TimeNon-activatorNon-activator
    Hemocompatibility – ASTM HemolysisNon-hemolyticNon-hemolytic
    Hemocompatibility – ThromboresistanceSimilar to control devicesSimilar to control devices
    Performance TestingDimensional VerificationConformance to specificationsPASS
    Luer IntegrityMet pre-determined criteriaPASS
    RHV SealingMet pre-determined criteriaPASS
    Tensile StrengthMet pre-determined criteriaPASS
    Kink ResistanceMet pre-determined criteriaPASS
    Torsion ResistanceMet pre-determined criteriaPASS
    Tip FlexibilityMet pre-determined criteriaPASS
    Air LeakageMet pre-determined criteriaPASS
    Liquid Leakage / Static BurstMet pre-determined criteriaPASS
    Dynamic BurstMaintained mechanical integrityPASS
    Hydrophilic Coating IntegrityMet pre-determined criteriaPASS
    Particulate RecoveryMet pre-determined criteriaPASS
    Simulated Use TestingMet pre-determined criteriaPASS
    RadiopacityMet pre-determined criteriaPASS
    Contrast DeliveryMet pre-determined criteriaPASS
    In Vivo ThromboresistanceMet pre-determined criteriaPASS
    Packaging IntegrityMet pre-determined criteriaPASS

    2. Sample Sizes Used for the Test Set and Data Provenance:

    The document states "All samples met the pre-determined acceptance criteria" for various performance tests, but does not specify the exact sample size for any of the individual tests. The provenance of the data (e.g., country of origin, retrospective or prospective) is also not provided. Many tests mention adherence to ISO standards, implying controlled laboratory conditions, but details about the test setting are not available. The Biocompatibility and Performance tests are "Non-Clinical Testing."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not provided as the tests conducted are primarily non-clinical, mechanical, and material-based, not involving human interpretation of medical images or patient outcomes that would typically require expert ground truth establishment in that manner. For the "Simulated Use Testing," "Radiopacity," and "Contrast Delivery," and "In Vivo Thromboresistance" which involve models and animals, the experts involved in conducting and evaluating these tests are not specified.

    4. Adjudication Method for the Test Set:

    An adjudication method is not applicable and therefore not mentioned, as the tests are primarily objective physical and chemical characterizations, rather than assessments requiring human judgment or consensus from multiple reviewers for a "ground truth" determination.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not conducted. This document describes a 510(k) submission for a medical device (catheter system), not an AI or imaging diagnostic tool. Therefore, human reader performance with or without AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

    No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm or software.

    7. The Type of Ground Truth Used:

    For the non-clinical tests, the "ground truth" is defined by the specifications and acceptable ranges established by industry standards (e.g., ISO) and the manufacturer's pre-determined acceptance criteria. For "Simulated Use Testing," "Radiopacity," "Contrast Delivery," and "In Vivo Thromboresistance," the ground truth would be based on direct observation and measurement within the respective models (neurovascular model, animal model), demonstrating the device's functional integrity and performance as intended. No human pathology, outcomes data, or expert consensus in the medical diagnostic sense is mentioned as ground truth.

    8. The Sample Size for the Training Set:

    This information is not applicable as this is a physical medical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as this is a physical medical device, not an AI model requiring a training set and associated ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1