510(k) Data Aggregation

    K Number
    K192061

    Validate with FDA (Live)

    Date Cleared
    2019-11-21

    (112 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

    The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

    This CT system can be used for low dose lung cancer screening in high risk populations .*

    • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

    Scan&GO:

    The in-room scan application is a planning and information system designed to perform the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.

    The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

    Device Description

    The subject device SOMATOM go.Platform with SOMARIS/10 syngo CT VA30 are Computed Tomography Xray Systems which feature one continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go.Platform with Software SOMARIS/10 syngo CT VA30 produces CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

    The Scan&GO mobile workflow is an optional planning and information software designed to perform the necessary functions required for planning and controlling of the SOMATOM go.Platform CT scanners. Scan&GO can be operated on a Siemens provided tablet or a commercially available tablet that meets certain minimum technical requirements. It allows users to work in the scanner and the patient.

    AI/ML Overview

    The provided text describes acceptance criteria and testing for the Siemens SOMATOM go.Platform CT Scanners with software version SOMARIS/10 syngo CT VA30, and the Scan&GO mobile medical application.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details various non-clinical tests conducted, with statements of the test results meeting the acceptance criteria. However, it does not present a single consolidated table of specific, quantifiable acceptance criteria alongside reported performance values for those criteria. Instead, it offers narrative summaries of the testing and its outcomes, indicating successful verification and validation.

    Below is a table constructed from the provided text, outlining the features tested and the reported performance (which is generally stated as "met acceptance criteria" or "similar/improved performance").

    Feature/TestAcceptance Criteria (Implicit)Reported Device Performance
    Non-Clinical Performance Testing:
    kV and Filter independent CaScorePerformance of special kernel variants Artifical120 and eDDensity and mDDensity similar or improved within accuracy limits compared to initial release versions.The test results show that performance of special kernel variants Artifical120 and eDDensity and mDDensity is similar or improved within the limits of accuracy of the test compared to the respective initial release versions. In conclusion, the features DirectDensity and Calcium Scoring at any kV have been enabled for the release SOMARIS/10 VA30.
    Recon&GO - Spectral ReconDeviations between cleared image processing algorithms in Inline DE and new realization "Spectral Recon" should be extremely small and not impact diagnostic performance.Deviations between the already cleared image processing algorithms in Inline DE and the new technical realization "Spectral Recon" are extremely small and are not expected to have any impact on the diagnostic performance. Residual deviations are a consequence of rounding differences and slight differences in implementation.
    TwinSpiral Dual Energy / TwinSpiral DEProvide CT-images of diagnostic quality, similar to conventional 120kV images in terms of CT-values and image noise at same radiation dose. Iodine CNR at same radiation dose comparable between Mixed images and 120kV images.Based on these results it can be stated that the TwinSpiral Dual Energy CT scan mode provides CT-images of diagnostic quality, which are similar to conventional 120kV images in terms of CT-values and image noise at same radiation dose. The mixed images show a slight reduction in the iodine CT-value, but at the same time image noise at same dose is also lower. So in combination the iodine CNR at same radiation dose is comparable between Mixed images and 120kV images.
    Flex 4D Spiral - Neuro/BodyScanned volume in agreement with planned scan range; irradiated range markers in agreement with exposed area on film.Scan ranges with the new Flex4D Spiral feature can be freely selected within the limits mandated by the scan mode and protocol. The scanned volume was found to be in agreement with the planned scan range for a variety of different tested scan modes, scan lengths and scanners. Radiochromic film placed in the isocenter for a variety of scan ranges showed that the irradiated range markers displayed by the scanner acquisition software during the planning of the respective F4DS scans were in good agreement with the exposed area on the film.
    DirectDensityAbility to provide images that can be shown as relative mass density or relative electron density.The conducted test performed demonstrated the subject device's ability to show relative mass or relative electron density images.
    HD FoVProvide visualization of anatomies outside the standard field of view; image quality standards for radiotherapy applications met.Phantom testing conducted to assess the subject device ability to provide visualization of anatomies outside the standard field of view and that the image quality standards for radiotherapy applications are met.
    Contrast media protocolAll Factory Contrast Protocols within limits prescribed by approved labeling of Ultravist®.All Factory Contrast Protocols are within the limits as prescribed by the approved labeling of Ultravist®. (no protocol for coronary CTA)
    InjectorCouplingCorrectness of contrast injection parameters transferred between CT device and supported injection devices verified.Correctness of the contrast injection parameters transferred between the CT device and the supported injection devices has been verified.
    Direct i4DAbility to acquire data for a full breathing cycle at every position even if respiratory rate changes, avoiding interpolation artifacts compared to conventional 4DCT.The test results show that with Direct i4D it is possible to acquire data for a full breathing cycle at every position of the patient even if the respiratory rate changes during the data acquisition. Compared to the conventional 4DCT scan mode interpolation artifacts (which occur because not for every position a complete breathing cycle could be acquired) can successfully be avoided with Direct i4D.
    Check&GOHelpful in aiding user to reduce instances where image quality may be compromised (for metal detection and contrast determination).The "Check&GO feature can be proven helpful in aiding the user to reduce instances where the image quality may be compromised." (For metal detection and automatic contrast state determination).
    Siemens Direct Laser (RTP Laser)Unit tested against general requirements, mechanics, connectors, function requirements, and integral light markers (IEC 60601-2-44).RTP-Laser Electronics - Test specification (Unit) Version 00 and Report - General Requirements - Mechanics, Connectors - Function requirements Attachment 12 to Report CN19-003-AU01-S01-TR31 - Test for the new RTP Laser Unit 10830876 - Integral Light Markers For Patient Marking (IEC 60601-2-44) were successfully demonstrated.
    Wireless Coexistence TestingSafe operation of wireless components in combination with applicable system functionality, ensuring coexistence with other devices.Testing for co-existence considered for following scenarios: Co-Channel Testing, Adjacent Channel Testing, RF Interference Testing, Separation Distance/Location Testing. Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.
    System Test (Workflow, User Manual, Legal/Regulatory)All acceptance criteria defined for these tests must be met.All tests performed meet the pre-determined acceptance criteria.
    System Integration Test (Functional, Image Quality, DICOM)All acceptance criteria defined for these tests must be met.All tests performed meet the pre-determined acceptance criteria.
    Subsystem Integration Test (Functional, DICOM)All acceptance criteria defined for these tests must be met.All tests performed meet the pre-determined acceptance criteria.

    2. Sample size used for the test set and the data provenance

    • Check&GO Testing:

      • Sample size: 500 CT-series from 100 patients.
      • Data provenance: Not explicitly stated, but clinical datasets were used ("clinical datasets from 100 patients"). It's specified as a "bench test," which implies it was likely retrospective from an existing data archive. Country of origin is not mentioned.
    • Other Non-Clinical Testing (Phantom, Integration, Functional): The document frequently refers to "phantom images," "test levels," "development activities," and "bench tests." No specific sample sizes for these tests (e.g., number of phantom scans) or data provenance are provided beyond the general descriptions of the tests themselves, which are stated as having been conducted "during product development."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Check&GO Testing:

      • Ground Truth Establishment: The datasets were "manually annotated with a detailed GT contrast-state (None, Low, InhomogeneousLow, Standard, InhomogeneousHigh, High)."
      • Number & Qualifications of Experts: Not specified.
    • Other Tests: For other tests, such as those involving image quality or physical measurements (e.g., Flex 4D Spiral, DirectDensity), the ground truth is typically derived from physical measurements, reference standards (e.g., known phantom properties), or established technical specifications, rather than expert consensus on clinical interpretation. The document does not mention the use of experts to establish ground truth for these tests. The indication for the new "Kidney Stones" feature notes: "Only a well-trained radiologist can make the final diagnosis under consideration of all available information," suggesting the involvement of radiologists in the clinical context, but not for ground truth establishment specifically for the device's technical validation.

    4. Adjudication method for the test set

    • The document does not explicitly describe an adjudication method (like 2+1, 3+1, etc.) for any of the test sets. For the Check&GO test, ground truth was "manually annotated," implying a single process for ground truth establishment rather than a consensus/adjudication method among multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study involving human readers and AI assistance is reported for this device in the provided text. The device itself is a CT scanner system and its associated software, not explicitly an AI-assisted diagnostic tool for interpretation in collaboration with human readers. The Check&GO feature is described as "aiding the user," but no study on human performance improvement is included.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The non-clinical performance testing, particularly phantom studies and specific feature evaluations like "kV and Filter independent CaScore," "Recon&GO - Spectral Recon," "TwinSpiral Dual Energy," "Flex 4D Spiral," "DirectDensity," "HD FoV," and "InjectorCoupling," can be considered standalone algorithm/device performance evaluations. These tests assess the technical output and accuracy of the device and its software features independent of human interpretation or interaction during the measurement process. The Check&GO feature's "Bench Test" also evaluates the algorithm's performance against annotated ground truth.

    7. The type of ground truth used

    • Check&GO: Expert annotation of "detailed GT contrast-state" (None, Low, InhomogeneousLow, Standard, InhomogeneousHigh, High) for 500 CT series.
    • Other Feature Tests (e.g., CaScore, Spectral Recon, Flex 4D Spiral, DirectDensity, HD FoV, TwinSpiral DE): Primarily derived from physical phantom measurements, comparison to established technical specifications, or reference images/algorithms (e.g., comparing to initial release versions or conventional 120kV images).
    • National Lung Screening Trial (NLST): Outcomes data from a large clinical trial (N Engl J Med 2011; 365:395-409) is cited to support the "low dose lung cancer screening" indication for use, not for direct ground truth establishment during this device's specific validation, but rather as supportive clinical literature for the screening concept itself.

    8. The sample size for the training set

    • The document does not specify any training set sizes. The studies described are primarily for verification and validation, not for training machine learning models. The Check&GO feature describes "500 CT-series from 100 patients were used for the testing of the algorithm," but this is explicitly called "testing," not training.

    9. How the ground truth for the training set was established

    • Since no training set details (size or establishment method) are provided, this information is not available in the document.