510(k) Data Aggregation

    K Number
    K191608

    Validate with FDA (Live)

    Date Cleared
    2019-07-16

    (29 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Angiographic catheters are designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures. Angiographic catheters with marker bands may also be used for anatomical measurements.

    Device Description

    The Impress Angiographic Catheter with Hydrophilic Coating is an intravascular diagnostic catheter intended for the administration of contrast for conducting fluoroscopic studies. The catheter is available in a variety of 5F braided configurations including lengths from 40 cm to 125 cm, with or without marker bands and with hydrophilic coating. The device is available in variety of tip shapes to cater to variation in physician preference and patient anatomy. The device is intended for single use only and is supplied in a sterile configuration.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the Impress Angiographic Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel efficacy or conducting extensive clinical trials as might be seen for entirely new technologies.

    Therefore, the acceptance criteria and study described are primarily focused on verifying that the new device performs comparably to the predicate and meets established safety standards for its type. The document does not describe an AI device, so information related to human readers, AI assistance, effect size, or standalone algorithm performance is not applicable.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for a physical medical device (angiographic catheter), the "acceptance criteria" are not performance metrics in the way one might think for software or AI (e.g., sensitivity, specificity). Instead, they are related to established international and internal standards for medical device design, manufacturing, and safety. The reported device performance is that it met these standards and was found substantially equivalent.

    Acceptance Criteria (Standards & Tests)Reported Device Performance
    Biocompatibility Testing:The Impress Angiographic Catheter met the requirements for biocompatibility based on FDA Blue Book Memorandum #G95-1 and ISO 10993-1.
    - CytotoxicityPerformed and met standards.
    - HemolysisPerformed and met standards.
    Bench Testing:The Impress Angiographic Catheter met the specified requirements for all bench tests.
    - Marker band radiopacity, position, width, integrityPerformed and met standards.
    - Tensile (shaft at marker band and tip to shaft bond)Performed and met standards.
    - Shaft ID/OD, Catheter lengthPerformed and met standards.
    - Catheter stiffness, Kink resistancePerformed and met standards.
    - Hydrophilic coating lubricity and coating integrity friction testingPerformed and met standards.
    - Tip length, Burst pressure rating, Flow ratePerformed and met standards.
    Design Validation:The Impress Angiographic Catheter met the specified requirements for all design validation tests.
    - Marker band size, shape, location, surface profile, radiopacityPerformed and met standards.
    - Distal Tip Inspection, Guidewire compatibilityPerformed and met standards.
    - Catheter push-ability, track-ability, torque-ability, withdrawalPerformed and met standards.
    - Kink resistance, Catheter stiffness, Curve retentionPerformed and met standards.
    International Consensus Standards:The device was tested and found compliant with these standards.
    - ISO 10555-1:2013 (Intravascular catheters - General requirements)Met
    - ISO 594-1:1986, ISO 594-2:1998 (Conical Fittings with 6% Luer Taper)Met
    - ASTM F640-12 (Radiopacity for Medical Use)Met
    - ANSI/AAMI/ISO 11135:2014 (Sterilization - Ethylene oxide)Met
    - ISO 10993-1, -4, -5, -10, -11 (Biological Evaluation of Medical Devices)Met
    - United States Pharmacopeia 41-NF36:2018, <151> Pyrogen TestMet
    - ASTM F756-13 (Hemolytic Properties of Materials)Met
    Overall Substantial Equivalence to K093004 Impress Angiographic CatheterThe subject device is substantially equivalent based on identical indications for use, basic performance and safety profile, principle of operation, fundamental design principles, materials, and manufacturing technology. The addition of a marker band did not raise new questions of safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The text describes various bench and design validation tests for the physical properties and functionality of the catheter. It does not refer to a "test set" of patient data or studies in the typical sense of evaluating AI or diagnostic software.

    • Sample Size: Not explicitly stated as a single "sample size." The testing involves components and finished devices, and the number of units tested for each specific criterion (e.g., tensile strength, burst pressure) would be determined by internal quality control and standard methods specified in the cited ISO/ASTM standards.
    • Data Provenance: The data provenance is from laboratory testing performed by Merit Medical Systems, Inc. (the manufacturer). This is not retrospective or prospective patient data, but rather controlled engineering and biological compatibility testing of the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the document describes the testing of a physical medical device. There is no "ground truth" derived from expert consensus on medical images or patient outcomes for this type of submission. The ground truth refers to the objective measurements against established engineering and biological standards.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., reading medical images). Here, the testing involves objective measurements and adherence to specified limits within the cited standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC study was not done. This device is a physical angiographic catheter, not an AI-assisted diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone algorithm study was not done. This device is a physical catheter, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" used for this device's evaluation is based on:

    • Established engineering standards: Specifications for physical properties like dimensions, strength, flow rate, stiffness, etc., as outlined in ISO and ASTM standards.
    • Biocompatibility standards: Criteria defined by ISO 10993 series and FDA guidance for evaluating biological response (e.g., cytotoxicity, hemolysis).
    • Predicate device's performance: The previous version of the Impress Angiographic Catheter (K093004) served as the benchmark for demonstrating substantial equivalence. The new device must perform at least as well and not introduce new safety concerns.

    8. Sample Size for the Training Set

    There is no training set as this is not an AI/machine learning device. The device itself is manufactured based on established designs and processes and then verified through testing.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set mentioned in the context of this device's submission.