510(k) Data Aggregation

    K Number
    K191041

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-28

    (497 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orthodontic Fixation Screw [Smart Anchor Miniscrew] is intended for use as a temporary anchor for orthodontic treatment for use in patients aged 12 and older.

    Device Description

    The Orthodontic Fixation Screw [Smart Anchor Miniscrew] is an orthodontic screw used for straightening of irregular teeth. It is designed for indications such as malocclusion treatment, straightening of irregular teeth, improvement of intermaxillary space and occlusion, and maintenance after orthodontic treatment. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices and it is used temporarily, removed after orthodontic treatment has been completed. Orthodontic screw is designed to facilitate placement of orthodontic appliances such as wires, springs, and elastic ligatures. Orthodontic Fixation Screw [Smart Anchor Miniscrew] is made of Titanium Alloy (T)-6AL-4V), which meets ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility. The head of the screw is through hole head type. The head of the Orthodontic Fixation Screw [Smart Anchor Miniscrew] is designed to apply various orthodontic tools. There is a hole in the screw head through which a wire can be passed to fix the mandible and maxilla in orthodontic treatment. Also, the design accommodates the use of the screw with the orthodontic appliances (bracket, wire, and elastic band etc.). The Orthodontic Fixation Screw [Smart Anchor Miniscrew] is available in in diameters of 1.2, 1.4, 1.6, 1.8, and 2.0mm and lengths of 8 and 10mm. The Orthodontic Fixation Screws with a diameter of 1.4mm and greater are also available in a 6mm length version.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Orthodontic Fixation Screw [Smart Anchor Miniscrew]". This document describes the device and its equivalence to predicate devices, focusing on bench testing rather than clinical studies or AI-based performance.

    Therefore, the requested details regarding acceptance criteria for AI/ML performance, study types (MRMC, standalone), ground truth establishment, expert adjudication, and sample sizes for training/test sets are not present in this document. The document confirms that no clinical performance testing was necessary for the demonstration of substantial equivalence.

    The acceptance criteria provided in this document are primarily for the physical and mechanical properties of the orthodontic screw, demonstrating substantial equivalence to a predicate device.

    Here's what can be extracted from the provided text regarding acceptance criteria and the study:


    Acceptance Criteria and Device Performance (Based on Bench Testing for Substantial Equivalence)

    Acceptance Criteria CategorySpecific Test/ParameterAcceptance Criteria (Implicit from Equivalence Claim)Reported Device Performance (Implicit from Equivalence Claim)
    Material BiocompatibilityCytotoxicity (ISO 10993-5)Meets applicable requirement of ISO 10993Performed, meets standards for Titanium Alloy (ASTM F136)
    Hypersensitivity (ISO 10993-10)Meets applicable requirement of ISO 10993Performed, meets standards for Titanium Alloy (ASTM F136)
    Intracutaneous Reactivity (ISO 10993-10)Meets applicable requirement of ISO 10993Performed, meets standards for Titanium Alloy (ASTM F136)
    Acute Systemic Toxicity (ISO 10993-11)Meets applicable requirement of ISO 10993Performed, meets standards for Titanium Alloy (ASTM F136)
    Local Effect after Implantation (ISO 10993-6)Meets applicable requirement of ISO 10993Performed, meets standards for Titanium Alloy (ASTM F136)
    SterilizationMoist Heat Sterilization ValidationValidated per ISO 17665-1 and ISO 17665-2Validated per ISO 17665-1 and ISO 17665-2 (Non-Sterile, Steam sterilized by user)
    Mechanical PerformanceAxial Pullout Strength TestDemonstrated substantial equivalence to predicate (K161335)Comparative bench test conducted using ASTM 543-13; differences in diameter addressed with reference device. Performance not quantified but deemed equivalent.
    Torsional Strength TestDemonstrated substantial equivalence to predicate (K161335)Comparative bench test conducted using ASTM 543-13; performance not quantified but deemed equivalent.
    Insertion-Removal Torque TestDemonstrated substantial equivalence to predicate (K161335)Comparative bench test conducted using ASTM 543-13; performance not quantified but deemed equivalent.
    Fracture Load Test (for Ø 1.2mm)Demonstrated substantial equivalence to reference device (K161197) with 1.2mm diameterConducted for Ø 1.2mm, performance not quantified but deemed equivalent.
    Rotational Fracture Torque Test (for Ø 1.2mm)Demonstrated substantial equivalence to reference device (K161197) with 1.2mm diameterConducted for Ø 1.2mm, performance not quantified but deemed equivalent.
    Dimensional/VisualVisual and Dimension CheckConforms to manufacturer's declared design characteristicsTested and conformed.
    Risk ManagementRisk AnalysisConducted based on ISO 14971:2012Analysis conducted based on ISO 14971:2012.

    Study Details:

    The document describes non-clinical performance testing to demonstrate substantial equivalence, not a study involving AI/ML performance, human readers, or clinical outcomes.

    1. Sample size used for the test set and the data provenance: Not applicable in the context of AI/ML. For bench testing, samples of the device were used. No information on the number of devices tested is explicitly provided. The provenance of the data is from in-house bench testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical and biocompatibility tests is established by physical measurements and laboratory analyses against established standards (ASTM, ISO).
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No. The document explicitly states "No clinical performance testing was necessary for the demonstration of substantial equivalence."
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: No. This device is an orthodontic screw, not an algorithm.
    6. The type of ground truth used:
      • For biocompatibility: Laboratory test results against ISO 10993 standards.
      • For mechanical tests: Physical measurements (e.g., force, torque) against ASTM 543-13 and manufacturer's specifications, compared to predicate device performance.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    In summary, the provided FDA 510(k) clearance letter and summary pertain to a physical medical device (an orthodontic screw) and its demonstration of substantial equivalence through non-clinical (bench) testing, not through clinical or AI/ML performance studies. Therefore, many of the requested data points related to AI/ML device validation are not relevant to this document.