510(k) Data Aggregation

    K Number
    K190525

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-03

    (60 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

    For children aged 17 and under, the balloon catheter system is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

    The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT intended to irrigate from within a target sinus for therapeutic procedures nor to facilitate diagnostic procedures with TruDi™ NAV Wire.

    Device Description

    The RELIEVA ULTIRRA® Sinus Balloon Catheter is a flexible catheter that is intended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon and permits the passage of an electromagnetic navigable sinus guidewire or sinus illumination system to facilitate access to the target sinus ostia. A hypotube is incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Simulated Use TestingMet all performance acceptance criteria
    Dimensional SpecificationsMet all performance acceptance criteria
    Balloon Burst PressureMet all performance acceptance criteria
    Joint Separation ForceMet all performance acceptance criteria
    Deflation TimeMet all performance acceptance criteria
    Balloon Cycle FatigueMet all performance acceptance criteria
    Shelf LifeEstablished per ASTM F1980-07, ASTM F88/F88M-09, ISTA 2A, and ASTM F2096-11 requirements.
    BiocompatibilityBiocompatible per ISO 10993-1.
    Sterilization Assurance Level (SAL)10^-6, validated per AAMI/ANSI/ISO 11135:2014 (overkill/half-cycle approach in a fixed chamber).
    Ethylene Oxide ResidualsMet ISO 10993-7:2008 requirements.
    Packaging Shelf LifeEstablished per ASTM F1980-07.

    2. Sample Size for Test Set and Data Provenance

    The document states, "Clinical data was not necessary for the RELIEVA ULTIRRA® Sinus Balloon Catheter." This indicates that no clinical test set was used, and therefore no sample size or data provenance for a clinical test set is applicable. The evaluation was based on non-clinical performance testing.

    3. Number of Experts and Qualifications for Ground Truth

    Since no clinical data or human-in-the-loop performance was performed for this 510(k) submission, the concept of "experts used to establish ground truth" is not applicable in the context of evaluating diagnostic accuracy or clinical decision-making. The ground truth for the non-clinical performance tests would be the established engineering and materials standards and specifications. Engineers and technicians with expertise in medical device testing would have established and verified these.

    4. Adjudication Method for Test Set

    Not applicable, as no clinical test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed. The submission explicitly states "Clinical data was not necessary."

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is a manual surgical instrument (catheter), not an algorithm or AI system. The performance studies conducted were non-clinical, evaluating the physical and mechanical properties of the device.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical performance data was based on established engineering standards, material specifications, and regulatory requirements (e.g., ISO, ASTM standards) for medical device performance (e.g., burst pressure, fatigue, biocompatibility, sterilization).

    8. Sample Size for Training Set

    Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The device's design and manufacturing processes are developed and refined through engineering principles and iterative testing, not through data-driven training sets as understood in AI systems.

    9. How Ground Truth for Training Set Was Established

    Not applicable, as there is no training set in this context. The "ground truth" for the device's development and manufacturing would be its intended function, design specifications, and relevant industry standards.