510(k) Data Aggregation

    K Number
    K190508

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-11

    (285 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVO 450 series is the air-powered dental handpiece which is used by trained dental professionals for removal of carious material, cavities, crown preparations and as a surgical tool for impacted third molar removal and periodontal procedures.

    Device Description

    The EVO 450 series High Speed Handpieces are similar to other high-speed dental handpieces currently on the US dental market in design, function, and intended use. The devices are air-powered handpieces that are reusable, ergonomically shaped, and are provided both with and without a fiber optic light system. Several models could be connected to couplings of manufacturers including TTBIO, KaVo, NSK Mach, Sirona, and Star. Others could be connected to hose connectors type 2 or type 3 according to ISO 9168:2009 Dental handpieces-Hose connectors for air driven dental handpieces. The EVO 450 series High Speed Handpieces will be supplied with water, air and light through the tube and the coupling of a dental treatment unit. For proper operation, make sure that the supply of cooling air is dry, clean and uncontaminated. Make sure that air is not emitted in the direction of the bur. After each patient use, disinfect the tubing waterlines including the handpiece waterlines by flushing with cleaning water to prevent cross-contamination according to handpiece waterlines disinfection procedure of main dental unit.

    AI/ML Overview

    This looks like a submission for a 510(k) premarket notification for a medical device. The document states that the EVO 450 series High Speed Handpieces are substantially equivalent to the predicate device, Maxima PRO2 45L (K141576). It does not contain information about the acceptance criteria or a study that proves the device meets the acceptance criteria in the format you requested, specifically regarding AI/algorithm performance.

    The provided document describes a medical device (dental handpiece), not an AI/ML powered device. As such, the concept of "acceptance criteria" and "study proving the device meets the acceptance criteria" as typically applied to AI/ML devices (e.g., performance metrics like sensitivity, specificity, AUC compared to a pre-defined threshold) is not present here.

    Instead, this document focuses on demonstrating substantial equivalence to a predicate device, which is the regulatory pathway for this type of medical device. This involves comparing the new device to a legally marketed predicate device in terms of:

    • Intended Use: Both devices are high-speed, air-powered dental handpieces for specific dental procedures.
    • Technological Characteristics: Comparison of physical attributes, materials, operating principles, and performance specifications.
    • Safety and Effectiveness: Demonstrated through compliance with recognized standards and non-clinical testing.

    Here's an analysis of the provided information in relation to your questions, noting the context of a non-AI medical device:

    1. A table of acceptance criteria and the reported device performance

      • Acceptance Criteria (Implicit for Substantial Equivalence): The implicit acceptance criteria here are that the proposed device (EVO 450 series) is substantially equivalent to the predicate device (Maxima PRO2 45L) in terms of safety and effectiveness, based on similar intended use and technological characteristics. This is demonstrated through direct comparison and adherence to standards.
      • Reported Device Performance: The "Summary Table of Substantial Equivalence" on pages 4 and 5 directly compares the proposed device's characteristics against the predicate device. This serves as the reported performance, showing that the device's specifications (e.g., speed, operation pressure, materials) are comparable or within acceptable ranges relative to the predicate and relevant standards.

      Table of Comparison (derived from the document):

      Descriptive InformationProposed Device (EVO 450 series)Predicate Device (Maxima PRO2 45L K141576)Comparison/Comment
      Indication for useRemoval of carious material, cavities, crown preparations, surgical tool for impacted third molar removal and periodontal procedures.Removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures.Substantially similar.
      Principle of operationAir-powered turbine, cooling water, optional light system.Air-powered turbine, cooling water, light system.Identical mechanism.
      Water portsFourOneDiffers but does not raise new safety or effectiveness questions.
      Fiber opticsWith and without fiber-optic glass rod built-in light systemWith built-in light systemOption for both, similar technology.
      Light intensity on input sideN/AApprox. 25000 LuxNot directly comparable, but presence of light system is noted.
      Head DimensionsΦ10.5×H12.5mmΦ12.5×H14.6mmSlight dimensional difference, but within a functional range.
      Type of chuckPush button locking chuckIdenticalIdentical.
      Materials of body of the deviceStainless steelStainless steelIdentical.
      Bur retention forceUp to 24 N-cmIdenticalIdentical (implied, as no difference noted).
      Operation pressure2.0-3.0 bar2.75±0.1 barOverlapping range.
      Speed360,000~440,000 rpm380,000~420,000 rpmOverlapping range.
      Conformance with standards for shankØ1.59mm~1.60 mmIdenticalIdentical (implied, as no difference noted).
      Head angle45-degreeIdenticalIdentical.
      Compliance to StandardsISO 14457:2017, ISO 9168:2009ISO 14457:2012, ISO 9168:2009Both comply with relevant standards (new device with updated version for one).
      Sterilization MethodSteam Autoclave (Moist Heat)Steam Autoclave (Moist Heat)Identical.
      Sterilization StandardISO 17665-1:2006ISO 17665-1:2006Identical.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • This information is not applicable and not provided in the context of a 510(k) submission for this type of device. The "testing" refers to non-clinical bench testing and adherence to standards, not a patient-based test set or clinical trial data collection.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable. This device is not an AI/ML diagnostic tool requiring expert ground truth for image or data interpretation. Performance is assessed through engineering specifications and compliance with standards.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. There is no "test set" in the sense of clinical cases or data requiring adjudication.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This document is for a dental handpiece, not an AI-assisted diagnostic or therapeutic device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable, as this is not an AI/ML algorithm.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

      • Not applicable in the context of an AI/ML algorithm. For this device, "ground truth" relates to engineering specifications verified through testing and adherence to established industry standards (e.g., ISO for dental handpieces, sterilization, biocompatibility).
    8. The sample size for the training set

      • Not applicable, as this is not an AI/ML algorithm development.
    9. How the ground truth for the training set was established

      • Not applicable, as this is not an AI/ML algorithm development.

    In summary: The provided document is a 510(k) summary for a conventional medical device (a dental handpiece). It focuses on demonstrating substantial equivalence to a predicate device through comparison of features and adherence to non-clinical performance standards. It does not involve AI/ML technology, and therefore, the questions related to AI/ML specific acceptance criteria, study design, and ground truth establishment are not addressed or applicable in this context.