Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190059

    Validate with FDA (Live)

    Device Name
    CEREC Guides
    Manufacturer
    Date Cleared
    2019-09-27

    (259 days)

    Product Code
    Regulation Number
    872.3980
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CEREC Guides are intended to support the dentist or oral surgeon when drilling for placement of dental implants. CEREC Guides are intended to be designed and fabricated using the Sirona Dental CAD/CAM System's CEREC Chairside software and CAM equipment, Galileos Implant dental implant planning software, and Calibra Universal Self-Adhesive Resin Cement.

    Device Description

    The CEREC Guide dental surgical guides are milled poly(methyl methacrylate) [PMMA] devices which are designed and fabricated utilizing the Sirona Dental CAD/CAM System with CEREC Chairside Software (K181520) and are intended to act only as dental implant placement templates. The CEREC Guides are designed in the Sirona Dental CAD/CAM System's CEREC Chairside Software utilizing, as an input, a completed dental implant treatment plan which is developed in the "Galileos Implant treatment planning software as cleared under premarket notification K093090.

    The CEREC Guide surgical guides are offered in two variants, the CEREC Guide 2 and CEREC Guide 3. Both the CEREC Guide 2 and CEREC Guide 3 are milled from the CEREC Guide Bloc milling blocks. The CEREC Guide Blocs are clear, PMMA blocks intended for milling utilizing the CEREC MC XL family of milling units. CEREC Guide Blocs are offered in two size variants, "medi" and "maxi".

    The CEREC Guide 2 surgical guides consist of the surgical guide body milled from the CEREC Guide Bloc PMMA milling blocks. The CEREC Guide 2 surgical guides are used in conjunction with CEREC Guide Drill Keys. CEREC Guide Drill Keys are stainless steel, reusable drill guide instruments which are offered in variants with internal diameters ranging from 2.0 mm to 4.85 mm.

    The CEREC Guide 3 surgical guides consist of the CAD/CAM surgical guide body milled from the CEREC Guide Bloc PMMA milling blocks. CEREC Guide 3 surgical guides feature integral titanium guide sleeves that are bonded the CEREC Guide 3 surgical guide body using Dentsply Sirona Calibra® Universal adhesive (K073173). The CEREC Guide 3 Guide Sleeves are single use and are offered in 4 size variants, with outer diameters ranging from 5.5 mm to 6.0 mm, and internal diameters ranging from 4.48 mm - 5.2 mm.

    AI/ML Overview

    The provided FDA 510(k) premarket notification for "CEREC Guides" focuses on demonstrating substantial equivalence to a predicate device ("SIMPLANT Guide," K170849) through non-clinical performance data. It does not involve a multi-reader multi-case (MRMC) study or artificial intelligence (AI) performance evaluation. Therefore, many of the requested elements for an AI-based device, such as ground truth establishment by experts, adjudication methods, and MRMC study details, are not applicable here.

    Here's a breakdown of the available information relevant to acceptance criteria and study proving device performance:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of acceptance criteria with specific numerical thresholds and corresponding performance results in the format requested, typical for AI/ML device submissions. Instead, it describes various non-clinical tests conducted to support substantial equivalence. The "acceptance criteria" are implied by the successful completion of these validation tests.

    Acceptance Criterion (Implied)Reported Device Performance and Study Type
    Software Validation: Conformity with IEC 62304 and FDA guidance for software in medical devices.Validation Study: Tested to ensure conformity with IEC 62304 ("Medical device software - Software lifecycle processes") and FDA guidance "Guidance for the Content of Premarket Submissions of Software Contained in Medical Devices (May, 2005)." This included validation of input data from Galileos Implant and calibration of the compatible milling unit. (Implied: Software functions correctly and safely).
    Accuracy of Milled Guides: Deviation of apical drill position.Validation Testing: Confirmed the accuracy of the milled CEREC Guide devices by comparing the surgical guide design in CEREC SW to the apical position determined from the implant treatment plan developed using Galileos Implant dental implant planning software. (Implied: Milled guides accurately reflect the treatment plan).
    Material Strength: Mechanical properties of the PMMA material.Material Strength Analysis: Based on requirements of ISO 10477 ("Dentistry - Polymer-based crown and bridge veneering materials"), published clinical literature on intra-oral surgical drilling and bite forces, and the PMMA material characteristics of CEREC Guides and PMMA reference devices. (Implied: Material is strong enough for intended use).
    Drill Sleeve Retention: Force required to push out bonded drill sleeves.Bench Test Data: Supported the requirement that the force needed to push out the bonded CEREC Guide 3 drill sleeve exceeds intra-surgical drilling forces reported in published literature. (Implied: Drill sleeves remain secure during use).
    Biocompatibility: Safety of materials in contact with the patient.Biocompatibility Testing: Conducted on the CEREC Guide Bloc PMMA material and the finished CEREC Guide 3 (incorporating PMMA, Guide Sleeves, and Calibra® Universal adhesive). This was done with reference to the June 2016 CDRH Guidance on ISO 10993-1. (Implied: Materials are biocompatible).
    High-Level Disinfection Process Validation: Effectiveness of disinfection.Validation Testing: Validated the high-level disinfection process recommended for single-use CEREC Guides prior to use, referencing the March 2015 CDRH/CBER Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling." (Implied: Disinfection method is effective).

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical bench and validation tests rather than a test set of patient data. Specific sample sizes for these engineering and material tests (e.g., number of guides milled for accuracy, number of drill sleeves tested for retention) are not detailed in this summary. The data provenance is from the manufacturer's internal testing as part of the regulatory submission process.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This submission relies on non-clinical engineering and performance testing, not on interpretative performance (like image reading) where expert ground truth is typically established. The "ground truth" for the device's function is its design specifications and physical properties, tested through controlled experiments.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There's no human adjudication process described, as it's not a study evaluating human interpretation or AI output against a consensus.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. This device, CEREC Guides, is a physical, milled surgical guide, not an AI or software-assisted diagnostic tool that would typically undergo an MRMC study to assess reader performance. The submission explicitly states: "No human clinical data was included in this premarket notification to support the substantial equivalence of the subject CEREC Guides."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. Its performance is evaluated based on its physical properties, manufacturing accuracy, and integration with existing approved software and hardware. The "device" itself is the physical CEREC Guide.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests is based on:

    • Design Specifications and Reference Standards: The intended design dimensions and functionalities derived from the Galileos Implant treatment planning software served as the reference for accuracy testing.
    • Established Material Properties and Biocompatibility Standards: ISO standards (e.g., ISO 10477, ISO 10993-1) and published literature for material strength and biocompatibility served as the benchmark.
    • Validated Disinfection Protocols: FDA and industry guidelines for high-level disinfection processes.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no AI/ML model, there is no training set or associated ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1