Search Results
Found 1 results
510(k) Data Aggregation
(6 days)
The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.
The PathLoc-L MIS Spinal System is intended for the following indications:
• Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, lordosis)
- Tumor
- Pseudarthrosis
- · Failed previous fusion
PathLoc-L MIS Spinal System consists of poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made of titanium alloy (ASTM F136).
The provided text is a 510(k) Summary for the PathLoc-L MIS Spinal System. It describes a medical device, which is a pedicle screw spinal system used to stabilize spinal segments. The submission is for adding new components and changing part numbers to an already cleared device (K161766).
Based on the document, this is not a study that proves a device meets acceptance criteria for an AI/CADe system. This is a submission for a physical medical device (spinal implant), and the "performance testing" section refers to mechanical tests conducted on the physical components of the spinal system, not the performance of an AI algorithm.
Therefore, most of the requested information regarding AI/CADe systems (e.g., sample sizes for test/training sets, ground truth establishment, MRMC studies, standalone algorithm performance, expert qualifications, adjudication methods) is not applicable to this document.
However, I can extract the relevant "acceptance criteria" and "device performance" in the context of this physical device's mechanical testing:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Test) | Reported Device Performance (as stated in the document) |
|---|---|
| ASTM F1717 Static Compression Test | "The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static compression test" (Implicitly, the device met the requirements of this standard as it was found "as safe, as effective and perform as well as the predicate device".) |
| ASTM F1717 Static Tension Test | "The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static tension test" |
| ASTM F1717 Static Torsion Test | "The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static torsion test" |
| ASTM F1717 Dynamic Fatigue Test | "The PathLoc-L MIS Spinal System was tested according to ASTM F1717... and dynamic fatigue test" |
| ASTM F543 Axial Pullout Test | "The PathLoc-L MIS Spinal System was tested... under ASTM F543, axial pullout test" |
Note on "Acceptance Criteria" and "Reported Performance": The document states that the performance testing was conducted according to specific ASTM standards. For a 510(k) submission, meeting these recognized standards for mechanical properties means the device is considered as safe and effective as a predicate device that also meets these standards. The document doesn't provide specific numerical results of these tests, but rather states that the tests were performed and the device (including the new components) performs as well as the predicate.
Given this is a physical implant device, the following points are not directly extractable or applicable as they pertain to AI/CADe systems:
- 2. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/CADe system.
- 3. Number of experts used to establish the ground truth... and the qualifications of those experts: Not applicable. Ground truth for an AI system is not relevant here.
- 4. Adjudication method: Not applicable.
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- 7. The type of ground truth used: Not applicable.
- 8. The sample size for the training set: Not applicable.
- 9. How the ground truth for the training set was established: Not applicable.
In summary, the provided document details the regulatory submission for a physical spinal implant system, focusing on its mechanical properties and equivalence to a predicate device, rather than the performance of a diagnostic AI algorithm.