510(k) Data Aggregation

    K Number
    K181888

    Validate with FDA (Live)

    Device Name
    Osstell Beacon
    Manufacturer
    Date Cleared
    2019-08-22

    (405 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Osstell Beacon is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.

    Device Description

    The Osstell Beacon is a modification of the Osstell ISQ (K082523). The system is designed to measure implant stability in the oral cavity and maxillofacial region. Similar to K082523, the Osstell Beacon is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The system involves the use of a Smartpeg (aluminum rod) attached to the implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement tip on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the implant and is derived from the resonance frequency value obtained from the Smartpeg.

    AI/ML Overview

    The provided document, a 510(k) summary for the Osstell Beacon device, does not contain detailed information about specific acceptance criteria or a dedicated study proving performance against such criteria. The document focuses on demonstrating substantial equivalence to a predicate device (Osstell ISQ K082523) by detailing technological characteristics rather than presenting a performance study with acceptance criteria.

    However, based on the nonclinical testing section, it can be inferred that performance testing was conducted to confirm compliance to design specifications. Since the document claims the "Beacon performs as intended" and lists "Accuracy: ±2 ISQ units" under its specifications, we can reasonably extrapolate this accuracy as an implicit acceptance criterion.

    Here's a breakdown of the requested information, with disclaimers about missing details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the information, an explicit table of acceptance criteria and a detailed performance study are not provided in this 510(k) summary. However, an inferred acceptance criterion for accuracy exists:

    Acceptance Criteria (Inferred)Reported Device Performance
    Accuracy: ±2 ISQ unitsAccuracy: ±2 ISQ units

    Note: This is an inferred acceptance criterion based on the device's stated accuracy in its specifications and the claim that performance testing confirmed compliance to design specifications. The document does not explicitly state this as an acceptance criterion from a study protocol.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Performance testing was conducted to confirm compliance to the design specifications." However, it does not provide any specific sample size for a test set or details about the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not mention the use of experts to establish ground truth for any test set. The device likely provides direct measurements rather than relying on expert interpretation for its core function.

    4. Adjudication Method for the Test Set

    Since no involvement of experts in establishing ground truth or a test set is explicitly described, there is no mention of an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document states: "Clinical studies were not required to validate the modifications in the Osstell Beacon." This implies that no MRMC comparative effectiveness study was conducted. The device is a measurement tool, and the submission focuses on its equivalence to a predicate device rather than human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The document indicates that "Performance testing was conducted to confirm compliance to the design specifications." The device operates as an "Implant Stability Analyzer" that provides an "Implant Stability Quotient (ISQ)" value directly. This implies a standalone performance evaluation of the algorithm's ability to accurately measure the resonance frequency and calculate the ISQ value. The "Accuracy: ±2 ISQ units" is a standalone performance metric.

    7. Type of Ground Truth Used

    The type of ground truth used for the performance testing is not explicitly stated but can be inferred as a reference measurement of resonance frequency or implant stability against which the device's calculated ISQ value would be compared. For a measurement device, this typically involves using a highly accurate, calibrated system or a known physical property (e.g., a standard with a known resonance frequency).

    8. Sample Size for the Training Set

    The document focuses on substantial equivalence and nonclinical testing. It does not mention any training set size because the device's underlying technology (Resonance Frequency Analysis and ISQ software algorithm) is based on established physical principles and is inherited from its predicate device (Osstell ISQ K082523), rather than requiring a machine learning model that needs a training set.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set being used in the context of machine learning, there is no information on how its ground truth might have been established. The device relies on a physical measurement and algorithm, not a trainable model.