510(k) Data Aggregation

    K Number
    K181703

    Validate with FDA (Live)

    Date Cleared
    2018-12-28

    (184 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX Implants can be placed with immediate function on singletooth, bar and bridges applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.

    Straumann® BLX SRAs and Antomic Abutments: Prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Temporary components can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. Final abutments may be placed into occlusion when the implant is fully osseointegrated. BLX Temporary Abutments have a maximum duration of usage of 180 days.

    Device Description

    The Straumann BLX Implants are fully tapered implants manufactured utilizing the Roxolid material and are finished with SLActive® surface. The connection is identified as conical fitting with Torx style engaging feature. There are two available prosthetic platforms identified as RB (Regular Base) and WB (Wide Base). The subject devices have a RB platform (a "small top/head") with a Torx-style internal connection, which is identical for all the implant lengths. They are provided sterile by gamma radiation and are available in various sizes.

    Straumann® BLX SRAs are manufactured from TAN and are intended to be placed on BLX Straumann dental implants to provide support for permanent screw-retained single or multi-unit restorations. They are provided sterile by gamma radiation and are available in various sizes and angulations.

    Straumann® BLX Anatomic Abutments are manufactured from TAN and are intended to be placed on BLX Straumann dental implants to provide support for permanent cement-retained single or multi-unit restorations. They are provided non-sterile and are available in various sizes and angulations. They are delivered with the corresponding basal screw. Anatomic Abutments are provided with pre-defined mucosa margins and with a size that allows for further modification by the customer via hand grinding tools.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Straumann® BLX Line Extension - Implants, SRAs and Anatomic Abutments. This submission focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than performing a de novo study to establish new acceptance criteria. Therefore, the information requested in your prompt (Acceptance Criteria and study design for proving it) is not explicitly present in the provided document in the way one might find it for a novel device undergoing clinical trials.

    However, based on the performance testing section, we can infer some aspects that act as "acceptance criteria" for demonstrating substantial equivalence to the predicate devices. The study conducted is primarily a series of bench tests and validation studies rather than a clinical trial or a study designed to prove new acceptance criteria.

    Here's an attempt to structure the answer based on the provided text, highlighting what is implicitly or explicitly stated about "acceptance" in the context of a 510(k) submission:


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission demonstrating substantial equivalence to predicate devices, explicit "acceptance criteria" as pass/fail thresholds for a novel claim are not presented. Instead, the performance tests aimed to show that the subject device performs equivalently to the predicate devices and meets relevant industry standards. The "performance" column here reflects the outcome of those equivalence demonstrations.

    Acceptance Criterion (Inferred from Equivalence Testing)Reported Device Performance
    Sterilization Validation: Sterility Assurance Level (SAL) of 10-6 in accordance with ISO 11137-1:2006.Validated to SAL of 10-6 using the over kill bioburden method in accordance with ISO 11137-2:2013 for BLX implants and SRAs.
    Shelf Life: Packaging and materials prevent adverse effects over time.Shelf life of 5 years demonstrated as packaging is equivalent to predicate/reference devices, and materials are not adversely affected by time.
    End-User Sterilization (Anatomic Abutments): Validation according to ISO 17665-1 and ISO 17665-2.Recommended sterilization method has been validated for BLX Anatomic Abutments.
    Pyrogenicity: Meets pyrogen limit specifications (e.g., LAL Endotoxin Analysis with testing limit of 20 EU/device).Device meets pyrogen limit specifications as determined by LAL Endotoxin Analysis with a testing limit of 20 EU/device (for blood contacting and implanted devices).
    Biocompatibility: Equivalent to predicate/reference devices per ISO 10993-1:2009 and FDA Guidance.Biological assessment shows the subject devices have equivalent body contact, duration, material formulation, and sterilization methods compared to predicate/reference devices. No new testing performed due to equivalence.
    Dynamic Fatigue & Static Strength: Equivalent to predicate/reference devices as per FDA guidance for root-form endosseous dental implants.Dynamic fatigue and static strength tests were performed on both RB and WB platforms (worst-case considerations) and shown to be equivalent to the reference devices.
    Insertion Torque: Allows for suitable implant insertion torques.Insertion torque tests were conducted on worst-case BLX implants (Ø3.75 mm) and demonstrated that the implants and related cutting instruments allow reaching suitable implant insertion torques.

    2. Sample size used for the test set and the data provenance

    • Sterilization Validation: The document states that the validation was in accordance with ISO 11137-1:2006 and ISO 11137-2:2013, specifically mentioning the "over kill bioburden method." These standards outline protocols for sample size determination, but the exact sample sizes used for the sterilization validation are not specified in this document.
    • Bench Testing (Dynamic Fatigue, Static Strength, Insertion Torque): The document mentions that "worst-case considerations" were used for testing various platforms (RB and WB) and implant sizes (e.g., Ø3.75 mm for insertion torque). Specific sample sizes for these bench tests are not provided.
    • Data Provenance: All data appears to be prospective bench testing and validation studies conducted by or for the manufacturer (Institut Straumann AG / Straumann USA, LLC). No country of origin for the data is specified beyond it being conducted by the applicant's entity.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The studies described are engineering and materials science tests (sterilization, biocompatibility, mechanical strength), not studies involving human interpretation or clinical outcomes that would require "experts to establish ground truth." The ground truth for these tests is based on established scientific and engineering principles, international standards (ISO), and FDA guidance documents.

    4. Adjudication method for the test set

    This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are relevant for studies that involve human interpretation and potential disagreements, such as image reading studies. The described studies are objective engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. The device is an endosseous dental implant system, not an AI-powered diagnostic tool or image analysis software. Therefore, an MRMC study or evaluation of AI assistance is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided. As stated above, this is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used

    The "ground truth" for the various tests described is derived from:

    • International Standards: e.g., ISO 11137-1, ISO 11137-2, ISO 17665-1, ISO 17665-2, ISO 10993-1.
    • FDA Guidance Documents: Such as "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile."
    • Engineering Principles: For mechanical testing like fatigue, strength, and torque.
    • Predicate Device Performance: The underlying "ground truth" for a 510(k) submission is that the new device is "substantially equivalent" to predicate devices, meaning it performs as safely and effectively without raising new questions of safety or effectiveness.

    8. The sample size for the training set

    This information is not applicable and not provided. There is no "training set" as this is not a machine learning or AI-driven device.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. There is no "training set" for this device.