Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180355

    Validate with FDA (Live)

    Date Cleared
    2019-01-10

    (336 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Advocate Non-Contact Infrared Thermometer is a non-sterile, reusable, handheld device. It can be used by consumers in homecare environment and doctors in clinic as reference. It is intended for measuring human body temperature of all ranges of people by detecting infrared heat from the forehead.

    Device Description

    The Advocate Non-Contact Infrared Thermometer model EF001A measures temperatures of people by detecting the infrared energy radiated directly from the forehead without physical contact. The device is composed of a Probe of metals with infrared sensor inside to detect the infrared energy, an LCD Display, a SCAN button to start measuring temperatures, a Power switch button to switch on or off the device, a Body/Ambient button to switch between two measuring modes, and an Enclosure of ABS.

    The device has the following features: one-second measuring time, measuring Body or Ambient temperature, 12-memory recalls, 9F/℃ unit switchable, over range message (Hi/Lo), low battery indication, auto display for the last reading when power is on, and auto shut-off when the device is idle for 60 seconds.

    AI/ML Overview

    The device under consideration is the Advocate Non-Contact Infrared Thermometer, Model EF001A. The provided information details its acceptance criteria and the studies conducted to prove it meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for infrared thermometers is accuracy, as defined by standards such as ASTM E1965-98.

    Acceptance CriteriaReported Device PerformanceComments
    Measuring Accuracy
    93.2°F - 94.8°F (34.0°C - 34.8°C)±0.5°F / 0.3°CComplies with ASTM E1965-98 (2016)
    95.0°F - 107.6°F (35.0°C - 42.0°C)±0.4°F / 0.2°CComplies with ASTM E1965-98 (2016)
    107.8°F - 109.4°F (42.1°C - 43.0°C)±0.5°F / 0.3°CComplies with ASTM E1965-98 (2016)
    Electrical SafetyComplies with IEC 60601-1:2005/A1:2012Pass
    EMC TestingComplies with IEC 60601-1-2:2014 & FCC 47 CFR Part 18Pass
    Performance TestingComplies with ASTM E1965-98(2016) & ISO 80601-2-56:2017Pass
    Biocompatibility TestingComplies with ISO 10993-5:2009 (Cytotoxicity) & ISO 10993-10:2010 (Skin Irritation, Sensitization)Pass
    Software Validation and VerificationComplies with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005"Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The primary study mentioned is a "Clinical test."

    • Sample Size: 120 subjects, categorized into:
      • 40 infants (0-1 year)
      • 40 children (1-5 years)
      • 40 adults (>5 years)
    • Data Provenance: Not explicitly stated from which country or location the subjects were sourced. The study is described as a "Clinical test" which typically implies prospective data collection, but it is not definitively stated whether it was prospective or retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The document refers to "Clinical Test" and compliance with standards like ASTM E1965-98, which outlines methods for determining clinical accuracy. However, it does not specify the number or qualifications of experts involved in establishing the ground truth measurements against which the device was compared.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. The clinical test section describes the assessment criteria for clinical accuracy but does not detail any adjudication methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The device is a thermometer, and its performance is evaluated against established temperature measurement standards, not through comparison with human readers or AI assistance in interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone thermometer. The performance tests ("Performance testing" and "Clinical test") demonstrate the device's accuracy in measuring temperature without human intervention other than the act of taking the measurement. Therefore, the reported performance is inherently standalone performance.

    7. The Type of Ground Truth Used

    The ground truth for the clinical accuracy test was established by assessing the device's measurements against the requirements of specific standards. For temperature measurement, the ground truth is typically a highly accurate reference thermometer or a standardized method as outlined in the referenced clinical accuracy standards. In this case, the "Clinical test" methods and criteria were assessed to meet the requirements of clinical accuracy per the referenced standard ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometer for Intermittent Determination of Patient temperature. This standard itself would describe the method for establishing ground truth, likely involving comparative measurements with an accepted standard.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. As a non-AI/ML device (a digital infrared thermometer), there is no "training set" in the context of algorithms that learn from data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1