    Device Name
    Stethee Pro
    Manufacturer
    Date Cleared
    2017-10-30

    (91 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stethee™ Pro 1 is an electronic stethoscope intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.

    Device Description

    The M3DICINE Stethee™ Pro 1 is an electronic stethoscope that picks up sounds of the heart, arteries, veins, lungs and other internal organs. Stethee™ Pro 1 provides capability for auscultation in frequencies from 20-2000Hz.

    After amplification and filtering, the sounds are transferred to the user's ears via any Bluetooth connected headphones. The Stethee™ Pro 1 includes features that permit it to record and store sounds on a peripheral smart device (e.g., mobile phone) by transferring the sounds as an audio buffer to the smart device via a Bluetooth® connection. The smart device makes use of a downloaded companion stand-alone Mobile Medical Application (MMA), Stethee™ App, to record, manage and replay the captured patient sounds on the smart device.

    Stethee™ App is also able to display the audio file as a phonocardiograph (21 CFR 870.2390) on the user's smart device and can use the audio file to calculate and display a simple Heart Rate. Stethee™ App does not direct or influence the performance of Stethee™ Pro 1.

    The Stethee™ Pro 1 device user interface includes a simple push down mechanism ("control knob") to start recording. It has a light-emitting diode (LED) display ring which indicates the device's status state (ON or OFF, charge status, and Bluetooth® synchronization status).

    The Stethee™ Pro 1 operates on a rechargeable certified 3.7V lithium ion polymer battery.

    The Stethee™ Pro 1 incorporates embedded software. The embedded software controls all the various features found in the Stethee™ Pro 1, such as sound capture, digital signal processing, volume control, LED display ring, and Bluetooth® transfer.

    AI/ML Overview

    The document describes the Stethee™ Pro 1 electronic stethoscope and its equivalence to a predicate device, the 3M™ Littmann® Model 3200. However, it does not provide acceptance criteria in terms of specific performance metrics with numerical targets or a detailed study proving the device meets those specific acceptance criteria.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device through comparative performance testing.

    Here's an attempt to answer your questions based on the provided text, highlighting what information is available and what is not:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the format of defined thresholds for different performance metrics. It rather presents a comparison of technical characteristics between the Stethee™ Pro 1 and the predicate device. The implied "acceptance criteria" is that the Stethee™ Pro 1 performs substantially equivalent to the predicate device in these characteristics.

    FeatureImplied Acceptance Criteria (Substantial Equivalence to Predicate)Stethee™ Pro 1 Reported Performance
    ClassificationElectronic Stethoscope, Class II, DQDElectronic Stethoscope, Class II, DQD
    Intended UseMedical diagnostic purposes, detection/amplification of internal sounds, used on any person undergoing physical assessmentSame as predicate
    ContraindicationsNoneNone
    Pickup SensorConvert sound waves to digital signalMicrophone (-22 dB sensitivity), Sampling Rate: 16kHz, Bit Rate: 16 bits
    Frequency ResponseResponsive in audible ranges, comparable to predicate's Bell/Diaphragm/Extended RangeResponsive from 20-2000Hz
    Maximum Sound LevelAmplifies up to 24XAmplifies up to 24X
    Volume ControlYes, in 8 stepsYes, in 8 steps
    Power SourceBattery-poweredRechargeable Lithium-ion polymer battery
    Low Battery IndicatorYesYes (LED color change)
    Operating Controls & IndicatorsON/OFF, status indicationON/OFF features by electronics (tapping mechanism), LEDs for status (ON/OFF, charging, Bluetooth® synchronization)
    Heart Rate: Detection & DisplayYes, displayed on device or peripheralYes, using companion Stethee™ App (MMA)
    Heart Rate: Minimum Audio Sample Requirements5 seconds recording0.3-2 second recording
    Heart Rate: MethodologyContinuous real-time calculation, updated every few secondsContinuous real-time calculation after initial 0.3-2 second sampling, updates after each heart beat
    Heart Rate: Range of Detection & Accuracy30-199 BPM with ±10% consistency or ±5 bpm30-200 BPM with allowable readout error rate of no greater than ±10 % (i.e., 10% consistency) of the input rate or ±5 bpm
    Heart Rate: Handling of inconsistent soundsInconsistent sounds indicated by "-"Inconsistent sounds displayed on Stethee™ App GUI as "-"
    Direct ListeningReal-time sound listening capabilityReal-time listening using a Bluetooth® enabled headset
    Recording and PlaybackCapability to record and playback soundsWith Stethee™ App (not on device itself)
    VisualizationSounds visualized as phonocardiograph using softwareSounds visualized as phonocardiograph using companion software Stethee™ App
    Wireless Technology & Peripheral Platform CompatibilityBluetooth® at 2.4 GHz, compatible with devicesBluetooth® at 2.4 GHz, compatible with Android and iOS devices
    Ambient & Frictional Noise Reduction TechnologyYesYes

    2. Sample size used for the test set and the data provenance

    The document states: "M3DICINE submitted performance testing information in this 510(k) demonstrating that the Stethee™ Pro 1 can perform over its intended range of operation (20-2000Hz). Comparative performance testing submitted in the 510(k) demonstrated the two devices perform in a substantially equivalent manner."

    However, no specific sample sizes for test sets or data provenance (e.g., country of origin, retrospective/prospective) are mentioned for this comparative performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The testing described is comparative performance testing against a predicate device, not necessarily a diagnostic study requiring expert ground truth for classification.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was described or performed. The device is an electronic stethoscope with an accompanying app that can display phonocardiographs and heart rate. It's not an AI-assisted diagnostic tool in the sense of improving human reader performance on complex image interpretation. The comparison is primarily on technical specifications and basic functionality.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The document describes the performance of the Stethee™ Pro 1 device itself, which includes embedded software for sound capture, digital signal processing, volume control, and Bluetooth transfer. The heart rate calculation by the Stethee™ App can be considered an "algorithm only" function in terms of its calculation process.

    • Heart Rate Calculation: The Stethee™ App "calculates and display a simple Heart Rate."
      • Minimum Audio Sample Requirements: Requires initial 0.3-2 second recording (predicate requires 5 seconds).
      • Methodology: Performs continuous real-time calculation after initial 0.3-2 second sampling and updates heart rate display after each heart beat (predicate updates every 2 seconds after initial 5 seconds).
      • Range of Detection and Accuracy: 30-200 BPM with an allowable readout error rate of no greater than ±10 % or ±5 bpm (predicate: 30-199 BPM with ±10% consistency).
    • Visualization: "Sounds can be visualized as phonocardiograph using companion software Stethee™ App." This is a standalone function of the software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the heart rate accuracy, the implied ground truth would be a validated heart rate measurement, likely from an ECG or a highly accurate reference device. However, the exact type of ground truth used for performance testing is not explicitly stated. For the broader "substantial equivalence" claim, the predicate device itself serves as the reference for comparison.

    8. The sample size for the training set

    The document does not mention a training set or any machine learning/AI model training in the context of diagnostic capabilities beyond basic heart rate calculation. The "embedded software" handles signal processing and device control but isn't described as a learnable AI model.

    9. How the ground truth for the training set was established

    As no training set is mentioned in the context of the device's diagnostic claims, this information is not applicable/provided.