510(k) Data Aggregation

    K Number
    K171645

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-25

    (50 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CoMatryx Collagen Wound Dressing may be used in the management of partial and full thickness wounds, pressure (stage I-IV) and venous ulcers, ulcers caused by mixed vascular etiologies, venous stasis and diabetic ulcers, 1st and 2nd degree burns, cuts, abrasions, and surgical wounds.

    Device Description

    Collagen Wound Dressing, a native bovine Type I Collagen which, when applied to a wound surface, absorbs wound fluid and maintains a moist wound environment in the management of wound healing. The CoMatryx Collagen Wound Dressing is provided sterile in the following patient ready configurations:

    • 10 gram Bottle
    • 1 gram Vial
    • 1 gram pouch
    AI/ML Overview

    Acceptance Criteria and Device Performance Study for CoMatryx Collagen Wound Dressing

    This document analyzes the acceptance criteria and the study performed to demonstrate the substantial equivalence of the CoMatryx Collagen Wound Dressing, as described in the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's acceptance criteria are based on demonstrating substantial equivalence to the predicate device, CollaTek Powder (now Medifil II Particles). The performance data primarily relies on direct comparison of characteristics and functional testing against specified parameters.

    ParameterAcceptance Criteria (Predicate Reference / Specific Value)Reported Device Performance (CoMatryx Collagen Wound Dressing)Test Methods (Standards)
    Indications for UseSame as predicateSame as predicate510k Comparison
    Product CodeKGNKGN510K
    Rx StatusYesYes510K
    Physical ShapePowderPowder510K
    ColorWhite to off-whiteWhite to off-whiteVisual
    Single UseSingle use onlySingle use only510K
    Package Sizes1 gram vial, 1 gram pouch, 10 grams bottle1 gram vial, 1 gram pouch, 10 grams bottle510K
    Water Absorption CapacityNLT 20 times its weight in USP Purified WaterNLT 20 times its weight in USP Purified WaterStrukmyer Procedure #HOMA-AFTP5R
    Particle SizePowderPowderStrukmyer Procedure #HOMA-AFTPRK
    Product TypeCollagenCollagen510K
    Product DescriptionType I Bovine CollagenType I Bovine Collagen510K
    Animal TissueDermisDermis510K
    SterileYesYes510K
    Sterilization MethodE-BeamE-Beam510K
    Product ClassificationUnclassifiedUnclassifiedFDA
    Moisture Content≤ 17%≤ 17%Strukmyer Procedure #HOMA-9YQRG8
    pH≥ 2.5≥ 2.5Strukmyer Procedure #HOMA-9Z9HUM
    InfraRed Spectroscopy ScanCharacteristic of Material, comparable to predicateCharacteristic of Material, comparable to predicateStrukmyer Procedure #HOMA-A2CJEJ
    SDS-PAGE AnalysisCharacteristic of Material, comparable to predicateCharacteristic of Material, comparable to predicateAlamo Labs, Inc. Report#3480-01
    BiocompatibilityMeets ISO 10993 standards (excluding cytotoxicity for in-vitro)Initial cytotoxicity indicated effect, but wound healing study found no impairment.Cytotoxicity, Sensitization Guinea Pig, Intracutaneous Irritation Test, Acute Systemic Toxicity, Wound Healing Study ( porcine model)

    2. Sample Size for the Test Set and Data Provenance

    The document does not explicitly state a specific "test set" in the context of a dataset for an AI device. Instead, the "testing" involves a series of laboratory and animal studies to demonstrate the device's characteristics and safety.

    • Sample Size for Biocompatibility Tests: The document does not specify the exact number of samples used for each in vitro biocompatibility test (Cytotoxicity, Sensitization Guinea Pig, Intracutaneous Irritation Test, Acute Systemic Toxicity).
    • Sample Size for Physical/Chemical Characterization: The number of samples for each physical and chemical characteristic (e.g., Water Absorption Capacity, Moisture Content, pH, IR Spectroscopy, SDS-PAGE) is not explicitly stated. These are typically performed on a statistically relevant number of production lots.
    • Sample Size for Wound Healing Study: The document refers to "a Wound Healing Study" conducted on "porcine wounds." However, the exact number of porcine subjects or wounds used in this study is not provided.
    • Data Provenance: The studies appear to be prospective in nature, conducted specifically for the 510(k) submission. The location of the studies is not explicitly mentioned, but "Alamo Labs, Inc." is named for the SDS-PAGE analysis, suggesting a US-based laboratory.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This section is not applicable as the device is a wound dressing, not an AI/diagnostic device that requires expert-established ground truth for a test set. The "ground truth" for this type of device is established through laboratory test methods and animal studies demonstrating physical, chemical, and biological properties.

    4. Adjudication Method

    This section is not applicable as there is no mention of a human-read diagnostic test set requiring adjudication in this 510(k) submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This section is not applicable. The device is a wound dressing, not a diagnostic AI system, so an MRMC comparative effectiveness study is not relevant to its regulatory approval.

    6. Standalone (Algorithm Only) Performance Study

    This section is not applicable. The device is a physical wound dressing and does not involve an algorithm or AI component.

    7. Type of Ground Truth Used

    The "ground truth" for the CoMatryx Collagen Wound Dressing is established through:

    • Comparative Analysis: Direct comparison of the device's physical, chemical, and functional characteristics against a legally marketed predicate device (CollaTek Powder/Medifil II Particles).
    • Laboratory Testing: Adherence to established scientific and regulatory standards for material characterization (e.g., IR Spectroscopy, SDS-PAGE, pH, moisture content, water absorption).
    • Biocompatibility Testing: Adherence to ISO 10993 standards.
    • Animal Outcomes Data: The "Wound Healing Study" on porcine wounds provides outcomes data for the cellular response to the dressing and its impact on the healing process, particularly in addressing initial in vitro cytotoxicity concerns.

    8. Sample Size for the Training Set

    This section is not applicable as the device is a physical wound dressing and does not involve a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reason as point 8.