Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171423

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-27

    (73 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nasal Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (at least 4 years and older and weighing >15 kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.

    Device Description

    The Nasal Alar SpO2 Sensor is a disposable, single patient use Pulse Oximetry sensor designed to attach to the patient's nasal alar region - the fleshy region at the side of the nose. Skin contact and adhesive free sensor retention is via soft silicone rubber cushions encapsulating the optical components. The Nasal Alar SpO2 Sensor with its associated patient cable terminates in a DB-9 connector compatible with monitors employing Nellcor SpO2 technology. The sensor utilizes red and IR LED light sources of 660 nm and 890 nm respectively along with a silicon photodiode detector to detect changes in oxygen saturation in the blood. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the study that proves the device meets those criteria, according to your defined points.

    Device: Nasal Alar SpO2 Sensor (Premarket Notification K171423 for Xhale, Inc.)
    Purpose of this 510(k): Expansion of indications to change patient population from a weight of >30 kg to at least 4 years and older and weighing >15 kg. The sensor itself is unchanged from the predicate device (K143216).


    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance acceptance criteria are inferred from the predicate device's performance, as the device itself is unchanged and the study is focused on the expanded patient population.

    Acceptance Criteria (from predicate)Reported Device Performance (for new patient population)
    SpO2 Accuracy (ARMS): 70-100% ± 3%Within ARMS specification of ± 3% for the range of 70-100%
    Pulse Rate (BPM): 30-240 bpm ± 3 bpmNo specific study data provided for BPM; stated as "not changed" from predicate, implying performance is identical.
    IEC 60601-1 (Electrical Safety): YesYes (Identical to predicate)
    IEC 60601-1-2 (EMC): YesYes (Identical to predicate)
    ISO 80601-2-61 (Pulse Oximetry Standards):Yes (Identical to predicate for mechanical strength, storage/operating temp/humidity, fluid ingress)
    Mechanical strength: YesYes (Identical)
    Storage and Operating Temperature and Humidity: YesYes (Identical)
    Fluid ingress: YesYes (Identical)
    Pulse rate accuracy: YesYes (Identical)
    Inter-device reliability and accuracy: YesYes (Identical)
    Biocompatibility (ISO 10993-1): Materials passed for prolonged duration, mucosal contact, external communicating tissueMaterials are identical and previously tested/accepted under K122996.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a number of volunteer subjects, but the study type is a "controlled desaturation study with healthy volunteers."
    • Data Provenance: The study was a "controlled desaturation study," which is typically prospective clinical data. No country of origin is specified, but given the FDA submission, it's highly likely to be conducted in the USA or follow international standards accepted by the FDA. It is a prospective clinical study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable for this type of device and study. The ground truth for SpO2 accuracy in desaturation studies is established by simultaneous arterial blood gas measurements, not expert human readers/reviewers.

    4. Adjudication Method for the Test Set

    • Not applicable. The ground truth (arterial blood gas measurements) does not require expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging AI systems where human interpretation is a key component to be improved. This device is a sensor measuring a physiological parameter.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    • Yes, performance was effectively assessed in a "standalone" manner, as the device's accuracy was compared against the objective reference standard (arterial blood gas measurements) in the desaturation study. This is an algorithm-only performance assessment in the context of a pulse oximeter sensor.

    7. The Type of Ground Truth Used

    • Objective Physiological Measurement: The ground truth for SpO2 accuracy was established through a "controlled desaturation study" which, by standard practice in pulse oximetry validation (e.g., as per ISO 80601-2-61), involves simultaneously measuring arterial blood gas (ABG) values (specifically co-oximetry for actual functional oxygen saturation in blood samples) alongside the device's SpO2 readings as subjects' oxygen levels are lowered.

    8. The Sample Size for the Training Set

    • Not applicable. This is a 510(k) submission for a physical sensor, not an AI/ML algorithm that requires a "training set" in the computational sense. The device itself (the sensor) is unchanged from the predicate. The "study" here is a clinical validation study rather than a machine learning training/testing paradigm.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" in the context of an AI/ML algorithm here.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1