    Intended Use

    ASAHI PTCA Guide Wires:

    ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature.

    ASAHI Peripheral Guide Wires:

    ASAHI Peripheral Guide Wires are intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.

    ASAHI Guide Wire Extension:

    The ASAHI Guide Wire Extension accessory is intended for extension of the working length of an already introduced guide wire when exchanging over-the-wire interventional devices during an angioplasty procedure.

    Device Description

    This modification details a change relating to the removal of perfluorooctanoic acid (PFOA) used by suppliers in the manufacture of PTFE (polytetrafluoroethylene) coating used on the ASAHI family of guidewires including ASAHI PTCA guide wires and ASAHI peripheral guidewires. All of these wires are steerable guide wires, constructed with a core wire with coil assembly. The distal ends of the guide wires have a radiopaque tip to achieve visibility. Various coatings are applied to the distal and proximal portions of the guidewires.

    This submission covers the addition of an alternate type of PTFE coating material. The designs and performance of the guidewires are unchanged.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for ASAHI PTCA Guide Wires, ASAHI Peripheral Guide Wires, and ASAHI Guide Wire Extension. This type of submission is for medical devices that are substantially equivalent to legally marketed predicate devices, meaning they do not require extensive novel clinical trials to demonstrate safety and effectiveness as a brand new device would.

    The document does not describe a study involving an AI/Machine Learning device, nor does it pertain to a device that would typically have "accuracy," "sensitivity," or "specificity" as acceptance criteria in the traditional sense of diagnostic or prognostic AI. Instead, it describes acceptance criteria for a physical medical device (guidewires) based on material changes (specifically, a change in the PTFE coating to remove PFOA) and proves it meets those criteria through non-clinical (bench) testing and biocompatibility assessments.

    Therefore, I cannot provide the information requested in points 2-9, nor can I create a table with "accuracy," "sensitivity," and "specificity" as these metrics are not relevant to this type of device and study.

    However, I can extract the acceptance criteria and the "reported device performance" as presented in the non-clinical testing section for this specific medical device.

    Acceptance Criteria and Reported Device Performance (Non-Clinical Testing)

    TestAcceptance Criteria (Implied by Test Method Summary and Intent)Reported Device Performance (Results/Conclusions)
    Coating AdherenceIntegrity of the coated core wire maintained after pretreatment and manipulation in excess of expected clinical use.Test results confirmed that the coating adhesion was maintained during simulated clinical use in all test articles.
    Coating Integrity & Particulate CharacterizationCoating integrity maintained and particulate generation characterized during simulated use (advancing/retracting through a clinically relevant model).This testing characterized the coating integrity and particulate generation during simulated use. The test results confirmed that the coating integrity was maintained during simulated clinical use of the test articles.

    Biocompatibility Testing (All acceptance criteria are "Passed" based on the results)

    TestAcceptance Criteria (Implied by Test Method Summary and Intent - typically "Pass" based on recognized standards)Reported Device Performance (Results/Conclusions)
    CytotoxicityNo evidence of causing cell lysis or toxicity.Passed. No evidence of causing cell lysis or toxicity.
    Intracutaneous IrritationDifference between test article and control less than 1.0.Passed. Difference between test article and control less than 1.0.
    SensitizationNo evidence of causing delayed dermal contact sensitization.Passed. No evidence of causing delayed dermal contact sensitization.
    Systemic ToxicityNo mortality or evidence of systemic toxicity.Passed. No mortality or evidence of systemic toxicity.
    Material Mediated PyrogenicityTest article judged as non-pyrogenic.Passed. Test article judged as non-pyrogenic.
    Hemocompatibility - HemolysisHemolytic index within acceptable limits (e.g., < 2.0% for non-hemolytic).Passed. Hemolytic index was 0.0%.
    Hemocompatibility - Complement Activation (SC5b-9, C3a)Not considered a potential activator of the complement system.Passed. Test article is not considered to be a potential activator of the complement system.
    Hemocompatibility - PTT"Considered activator" or "not considered activator" as per standard (context needed for specific acceptance value).Passed. Test article is considered activator. (Note: This might mean it activates PTT, which could be an expected or acceptable characteristic depending on the material and intended use, or simply that the test was completed and the result recorded).
    Hemocompatibility – in vivo ThromboresistanceConsidered thromboresistant.Passed. Test article is considered thromboresistant.

    Regarding the other requested information (points 2-9):

    This document describes a 510(k) submission for a physical medical device (guidewires) with a material change, not an AI/Machine Learning device. Therefore, the following points are not applicable in the context of this document:

    • Sample size used for the test set and the data provenance: Not an AI device using data sets in this manner. Testing involved physical samples of guidewires.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Ground truth for this device is established through internationally recognized physical and biological testing standards (e.g., ISO, ASTM, USP).
    • Adjudication method for the test set: Not applicable for physical/biocompatibility testing.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable, as this is not an AI diagnostic/prognostic device for human interpretation.
    • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable to a physical guidewire.
    • The type of ground truth used: For physical/biocompatibility testing, the "ground truth" is defined by the established and validated test methods and passing criteria of the relevant standards (e.g., ISO 10993 series for biocompatibility).
    • The sample size for the training set: Not applicable to a physical guidewire.
    • How the ground truth for the training set was established: Not applicable to a physical guidewire.