Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163031

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-27

    (269 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bunegin-Albin Air Aspiration Set is intended to aspirate venous air emboli.

    Device Description

    The Bunegin-Albin Air Aspiration Set is comprised of an air aspiration catheter, wire guide, entry access needle, dilator, syringe, stopcock, and scalpel. The air aspiration catheter is manufactured from radiopaque extruded polyethylene tubing and is designed with a pre-molded winged fitting. The catheter is 5.8 French in size with an endhole diameter of 0.035 inches. The wire guide is manufactured from stainless steel coils, a safety wire, and a mandril wire. The access needle is manufactured from stainless steel. The dilator is manufactured from polyethylene and has a pre-molded proximal hub. The set also includes a 12 cc syringe, a low pressure three way stopcock, and a scalpel. The Bunegin-Albin Air Aspiration Set is sterilized by ethylene oxide and intended for one-time use.

    AI/ML Overview

    This documentation outlines the acceptance criteria and the results of various studies conducted for the Bunegin-Albin Air Aspiration Set to demonstrate its performance and substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Performance (Zero Time)
    Catheter Shaft TensilePeak load of catheter shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Sideport TensilePeak load of catheter sideport section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Hub-to-Shaft TensilePeak load of catheter hub-to-shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Air AspirationNo air shall enter the hub (Annex D of BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Liquid LeakageNo part of the catheter shall leak liquid (Annex C of BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Kink LengthCatheter shall not kink (Annex B of BS EN 13868)Acceptance criterion met.
    Wire Guide CorrosionNo evidence of corrosion that could affect functional performance (Annex B of BS EN ISO 11070)Acceptance criterion met.
    Wire Guide FlexNo signs of defects or damage, including flaking or material loss (Annex G of BS EN ISO 11070)Acceptance criterion met.
    Wire Guide FractureWire guide shall not fracture when wound around an approximate former (Annex F of BS EN ISO 11070)Acceptance criterion met.
    Wire Guide TensilePeak load to failure ≥ 10 N (Annex H of BS EN ISO 11070)Acceptance criterion met.
    Wire Guide Rotations to FailureCharacterized for the wire guide.The number of rotations to failure was characterized. (Specific values not provided in this document but implied to be sufficient for meeting overall acceptance related to design input requirements.)
    Wire Guide Tip DeflectionCharacterized for the wire guide.The tip deflection side loads were characterized. (Specific values not provided in this document but implied to be sufficient for meeting overall acceptance related to design input requirements.)
    Dilator Shaft TensilePeak load of dilator shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
    Dilator Hub-to-Shaft TensilePeak load of dilator hub-to-shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
    Dimensional, Compatibility, Surface AnalysisAll measurements and dimensional requirements within listed tolerances. Set components compatible. External surface of effective length free of extraneous matter.Acceptance criterion met.
    Particulate TestingMeet USP 788 thresholds for small volume injections.Acceptance criterion met.
    Resistance to Overriding TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
    Separation Force TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
    Unscrewing Torque TestingTesting conducted in accordance with ISO 594-2.Acceptance criterion met.
    Performance (3-Year Accelerated Aging)
    Catheter Shaft TensilePeak load of catheter shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Sideport TensilePeak load of catheter sideport section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Hub-to-Shaft TensilePeak load of catheter hub-to-shaft section ≥ 15 N (BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Air AspirationNo air shall enter the hub (Annex D of BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Liquid LeakageNo part of the catheter shall leak liquid (Annex C of BS EN ISO 10555-1)Acceptance criterion met.
    Catheter Kink LengthCatheter shall not kink (based on a 20 mm kink radius) (Annex B of BS EN 13868)Acceptance criterion met.
    Dilator Shaft TensilePeak load of dilator shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
    Dilator Hub-to-Shaft TensilePeak load of dilator hub-to-shaft section ≥ 15 N (BS EN ISO 11070)Acceptance criterion met.
    Animal Testing
    Acute performance of aspiration catheter in animal modelCatheter shall receive a grading of "adequate" or "good" in each of the performance parameters, and it shall be able to remove air from the target sites.Acceptance criterion met.
    Biocompatibility
    Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Hemocompatibility (Hemolysis, Complement Activation, In Vivo Thrombogenicity)Testing performed per ISO 10993-1 and FDA guidance to ensure biocompatibility.Acceptance criterion met.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample sizes for each physical and chemical test (e.g., Catheter Shaft Tensile, Wire Guide Corrosion). Instead, the document refers to compliance with international standards (e.g., BS EN ISO 10555-1, BS EN ISO 11070, ISO 594-2, BS EN 13868, USP 788) which would implicitly define sample sizes for such tests.

    For the Animal Testing, a specific sample size is not stated, but the testing evaluated the "acute performance" in an animal model. The provenance would be experimental ("prospective" if considered in the context of device development) rather than clinical human data. The country of origin for the animal study is not specified in the provided text.

    The data for all engineering performance and biocompatibility tests is based on prospective testing conducted specifically for this device submission to demonstrate compliance with standards and design input requirements. The country of origin of this testing data is not explicitly stated, but it is for a device manufactured by Cook Incorporated, presumably in the US or one of their international facilities.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For the engineering performance tests (tensile strength, kink resistance, leakage, etc.) and particulate testing, "ground truth" is established by adherence to quantitative specifications defined by international standards (e.g., ≥ 15 N, no air entry, no liquid leak, USP 788 thresholds). Therefore, these do not rely on human expert consensus in the same way clinical image interpretation might. The "experts" in this context would be the technicians and engineers performing and verifying the tests against the defined standards.

    For Animal Testing, the grading of "adequate" or "good" for performance parameters and ability to remove air would imply evaluation by experts, likely veterinary surgeons or researchers with expertise in animal models and medical device performance. The number and specific qualifications of these experts are not provided.

    For Biocompatibility testing, "ground truth" is established by the results of standardized tests per ISO 10993-1 and FDA guidance. Experts would be toxicologists and other scientific professionals interpreting these results, but their number and specific qualifications are not detailed.

    4. Adjudication Method for the Test Set

    For the physical, chemical, and mechanical performance tests, the adjudication method is direct comparison of measured values against predefined numerical or qualitative acceptance criteria outlined in international standards. There is no mention of a human consensus or specific adjudication for these quantitative evaluations beyond meeting the standard.

    For the Animal Testing, the "grading of 'adequate' or 'good'" suggests an assessment, but the specific adjudication method (e.g., single expert, 2+1 consensus) is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The device is a physical medical instrument (an air aspiration set), not an AI or imaging diagnostic device that would typically involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable as this is a physical medical device. The "performance" studies are standalone in the sense that they evaluate the device's physical properties and functionality without human intervention being part of the device's performance mechanism, but rather as the operators and evaluators of the device.

    7. Type of Ground Truth Used

    The ground truth for the various tests is established through a combination of:

    • International Standards and Specifications: For most physical, mechanical, and chemical tests (tensile strength, kink length, leakage, particulate counts, etc.), the ground truth is defined by the acceptance criteria set forth in specific ISO and BS EN standards (e.g., BS EN ISO 10555-1, BS EN ISO 11070, ISO 594-2, BS EN 13868, USP 788 thresholds).
    • Empirical Observation/Expert Assessment: For animal testing, the ground truth is based on the observed "adequate" or "good" performance and the ability to remove air from target sites, likely evaluated by trained personnel.
    • Biocompatibility Standards: For biocompatibility, the ground truth is compliance with the requirements of ISO 10993-1 and FDA guidance, demonstrated through specific laboratory test results.

    8. Sample Size for the Training Set

    Not applicable in the context of device performance testing for substantial equivalence. There is no mention of a "training set" as this device does not involve machine learning or AI algorithms requiring data for training. The studies described are for verification and validation of the manufactured product.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1