510(k) Data Aggregation

    K Number
    K162538

    Validate with FDA (Live)

    Date Cleared
    2016-11-17

    (66 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM) is for in vitro diagnostic use in the verification of calibration and reportable range of the ADVIA Centaur® Insuin assay.

    The ADVIA Centaur® Insulin (IRI) Calibrator is for in vitro diagnostic use in calibrating the ADVIA Centaur® systems Insulin assays.

    Device Description

    ADVIA Centaur® Insulin (IRI) Master Curve Material is a set of ten vials of material containing various levels of insulin in buffered saline with casein, potassium thiocyanate (3.89%), sodium azide (<0.1%), and preservatives. Each set contains ten levels (MCM1-10); ready-to-use 1.0 mL per level. MCM1 contains no analyte. The IRI MCMs assigned values are lot-specific of target values: 0.0, 2.5, 4.5, 10.0, 20.0, 39.0, 79.0, 158, 225, and 330 mU/L.

    ADVIA Centaur® Insulin (IRI) Calibrator is a set of four vials (two vials of low calibrator and two vials of high calibrator) comprised of buffered saline, with casein, potassium thiocyanate, sodium azide and preservative. Only the high calibrator contains insulin.

    AI/ML Overview

    The provided text describes modifications to the "ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM)" and "ADVIA Centaur® Insulin (IRI) Calibrator" and outlines the studies conducted to demonstrate their substantial equivalence to predicate devices.

    Here's an analysis of the acceptance criteria and the study information, extracting what is available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in a table format with corresponding performance metrics for the modified devices. Instead, it states that "Verification testing was conducted and results show that the modified devices meet pre-determined acceptance criteria." and that a "method comparison study demonstrates that human clinical sample results are equivalent with both calibrators (modified and unmodified)."

    However, based on the context, the implicit acceptance criteria and reported performance would be:

    Acceptance Criteria CategoryImplicit Acceptance Criteria (based on "same" characteristics and equivalence claims)Reported Device Performance (based on "same" characteristics and equivalence claims)
    ADVIA Centaur® Insulin (IRI) Master Curve Material (MCM)
    Intended UseSame as predicateSame as predicate
    InstrumentSame as predicate (ADVIA Centaur Systems)Same as predicate (ADVIA Centaur Systems)
    FormSame as predicate (Liquid)Same as predicate (Liquid)
    MatrixSame as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)Same as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)
    Number of LevelsSame as predicate (10)Same as predicate (10)
    Target ConcentrationsSame as predicate (0.0, 2.5, 4.5, 10.0, 20.0, 39.0, 79.0, 158, 225, 330 mU/L)Same as predicate (0.0, 2.5, 4.5, 10.0, 20.0, 39.0, 79.0, 158, 225, 330 mU/L)
    Fill VolumeSame as predicate (1.0 mL)Same as predicate (1.0 mL)
    Shelf Life/StabilitySame as predicate (12 months at 2-8 degrees C)Same as predicate (12 months at 2-8 degrees C)
    On board StabilitySame as predicate (8 hours on board)Same as predicate (8 hours on board)
    ADVIA Centaur® Insulin (IRI) Calibrator
    Intended UseSame as predicateSame as predicate
    InstrumentSame as predicate (ADVIA Centaur Systems)Same as predicate (ADVIA Centaur Systems)
    FormSame as predicate (Liquid)Same as predicate (Liquid)
    MatrixSame as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)Same as predicate (Buffered saline with casein, potassium thiocyanate, sodium azide and preservatives)
    Number of LevelsSame as predicate (2)Same as predicate (2)
    Target ConcentrationsSame as predicate (Low: <0.5 mU/L, 135 – 165 mU/L)Same as predicate (Low: <0.5 mU/L, 135 – 165 mU/L)
    Fill VolumeSame as predicate (1.0 mL)Same as predicate (1.0 mL)
    Shelf Life/StabilitySame as predicate (18 months at 2-8 degrees C)Same as predicate (18 months at 2-8 degrees C)
    On board StabilitySame as predicate (8 hours)Same as predicate (8 hours)
    Overall PerformanceEquivalent human clinical sample results with modified and unmodified calibratorsAchieved equivalence in human clinical sample results.
    OtherMeets pre-determined acceptance criteria for verification testingModified devices meet pre-determined acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: Not explicitly stated. The document mentions "human clinical sample results" were used in a method comparison study, but the number of samples is not provided.
    • Data provenance: "human clinical sample results" are mentioned, suggesting human patient samples. The country of origin and whether the data is retrospective or prospective are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The study concerns assay calibrators and master curve materials, where "ground truth" would typically refer to the accurately assigned values of the analytes in the calibrators/controls, usually established through highly accurate reference methods or certified reference materials, not expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable in this context. Adjudication methods like 2+1 or 3+1 are typically used for studies involving human interpretation of images or clinical data, where consensus among experts establishes ground truth. This is a study of in-vitro diagnostic device performance (calibrators and master curve materials), not human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not a study involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is an in-vitro diagnostic calibrator/master curve material, not an algorithm. The performance evaluation is inherently "standalone" in the sense that the device itself is being tested for its ability to correctly verify calibration and calibrate assays. The study of "method comparison" compares the results obtained using the modified calibrators/materials versus the unmodified ones on human clinical samples, which is a standalone performance assessment of the materials.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the calibrator values and master curve material values is inherent in their design and manufacturing process, where specific concentrations of insulin are prepared and assigned. For the "method comparison study," the ground truth is implied to be the measurements obtained using the predicate (unmodified) calibrators/materials, with which the modified devices' performance is compared for equivalence. It's about demonstrating that measurements of "human clinical sample results" using the new materials are equivalent to those obtained using the established predicate materials.

    8. The sample size for the training set:

    Not applicable. This device is an in-vitro diagnostic calibrator/master curve material, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.