Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162377

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-22

    (29 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerGlide Pro™ Midline Catheter is inserted into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide Pro™ Midline Catheter is suitable for use with power injectors.

    Device Description

    The PowerGlide Pro™ Midline Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device is intended for short term use (<30 days) to sample blood and administer fluids intravenously, and employs an identical placement technique to the cited predicate device. The subject device consists of an introducer needle with a passive safety mechanism, guidewire, and single lumen catheter rated for power injection. The subject PowerGlide Pro™ Midline Catheter features a catheter with a reinforced tip to support patency of the catheter during aspiration or blood draw. The PowerGlide Pro™ Midline Catheter is offered in 18, 20, and 22 gauge sizes. The 18 and 20 gauge devices are offered in 8 cm or 10 cm lengths. The 22 gauge device is offered in only an 8 cm length.

    AI/ML Overview

    The provided document describes the PowerGlide Pro™ Midline Catheter, which is an intravascular catheter intended for short-term use (<30 days) to sample blood or administer fluids intravenously. The submission is a 510(k) premarket notification, indicating substantial equivalence to a previously cleared predicate device (K153280). The key modification to the new device is a "reinforced catheter tip."

    Here is an analysis of the acceptance criteria and the study performed, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document describes several performance tests conducted to demonstrate substantial equivalence of the modified PowerGlide Pro™ Midline Catheter to its predicate. The acceptance criteria are generally framed as "less than or equal to the predicate device," "not swell beyond twice the size of the labeled OD," "not collapse under a vacuum," "remains in the same orientation... and does not kink," and "be greater than maximum use pressure." The reported performance is summarized as the device "met all predetermined acceptance criteria." Specific numerical values for criteria or performance are not explicitly provided in this summary for most tests, but rather stated as meeting the internal standards or predicate performance.

    Performance TestAcceptance Criteria (as stated or implied)Reported Device Performance
    Catheter Tip Adhesion Break ForceForce required to push catheter off needle after conditioning shall be less than or equal to the predicate device. (Bard internal standards and procedures)Met all predetermined acceptance criteria
    Catheter Swelling ODCatheter does not swell beyond twice the size of the labeled OD during power injection. (Bard internal standards and procedures)Met all predetermined acceptance criteria
    Catheter CollapseCatheter tip will not collapse under a vacuum during aspiration. (Bard internal standards and procedures, FDA guidance)Met all predetermined acceptance criteria
    Tip Location During and After Power InjectionCatheter tip remains in the same orientation during power injection (tip pointing in direction of venous flow) and does not kink following power injection when conducted at the maximum indicated flow rate. (Bard internal standards and procedures)Met all predetermined acceptance criteria
    Reinforced Tip Flexural Fatigue (Cyclic Kinking and Burst Test)Burst pressure must be greater than maximum use pressure when the tip is occluded, following cyclical kink conditioning. (Bard internal standards and procedures, ISO 10555-1:2013)Met all predetermined acceptance criteria
    Reinforced Tip Separation StrengthBurst pressure must be greater than maximum use pressure when the tip is occluded. (ISO 10555-1:2013)Met all predetermined acceptance criteria
    Biocompatibility (Hemolysis, Coagulation, Complement Activation)Tests conducted per ISO 10993-1:2009 and ISO 10993-4:2002 to demonstrate hemocompatibility. Specific criteria would be defined by these standards (e.g., non-hemolytic, no significant activation of coagulation/complement).Met all predetermined acceptance criteria derived from the verification tests related to biocompatibility evaluation

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify the sample sizes used for the test sets for any of the performance tests. It also does not provide information regarding the provenance of the data (country of origin, retrospective or prospective). These tests appear to be laboratory-based engineering performance tests rather than clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This type of information is not applicable to the studies described. The studies are engineering performance tests of a medical device, not clinical studies requiring expert interpretation or ground truth establishment in a medical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable to the studies described, as they are engineering performance tests and not clinical studies requiring adjudication of medical findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is an intravascular catheter, not an AI-powered diagnostic or assistive technology. No MRMC study or AI-related effectiveness study was performed or is relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a physical medical instrument, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the performance tests listed are the established physical and mechanical properties (e.g., force, swelling, flow rate, kink resistance, burst pressure) as defined by Bard's internal standards, FDA guidance, and international standards (ISO 10555-1:2013, ISO 10993 series). These are objective measurements against defined standards, not based on expert consensus, pathology, or outcomes data.

    8. The sample size for the training set:

    This information is not applicable. There is no "training set" as this is not an algorithm or AI system being developed or evaluated.

    9. How the ground truth for the training set was established:

    This information is not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1