510(k) Data Aggregation

    K Number
    K161432

    Validate with FDA (Live)

    Date Cleared
    2017-02-08

    (260 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The myNeWT system is indicted for patients who would benefit from Negative Pressure Wound Therapy as it may promote wound healing by removing low to moderate levels (< 37.5 ml/day) of exudate and infectious materials.

    Appropriate wound types include:

    • Chronic wounds
    • Acute, Sub-acute and dehisced wounds
    • Traumatic wounds
    • Partial thickness burns
    • Pressure or venous insufficiency ulcers
    • Diabetic ulcers
    • Flaps and grafts
    • Closed incision sites
    Device Description

    The myNeWT Negative Pressure Wound Therapy System is a portable, body worn, Single Use system for patients who have hard to heal wounds. Therapy is accomplished by the electronically controlled pump unit delivering a continuous negative pressure at -80 mmHg ±20mmHg to the wound surface, which draws the exudate into a flexible exudate canister, creating an environment that promotes wound healing via the removal of low to moderate amounts (<37.5ml/day) of exudate and infectious materials from the wound. When the canister is full it can be replaced without disturbing the wound bed or the wound dressing. This device is programmed to run for 7 days after the batteries are inserted, after which the device will not operate and will need to be replaced.

    The kit consists of :-

    • A battery operated vacuum pump, supplied with 2 x AA size Lithium batteries,
    • a flexible exudate canister with attached tubing,
    • a sterile wound interface kit.
    • a carrying bag with straps for the pump and canister.
    • User Manual
    • Quick Reference Guide Healthcare Professional
    • Quick Reference Guide - Patient

    It must be used with a commercially available wound dressing as selected by the Healthcare Professional based on the wound requirements or the physician's prescription. These dressings are NOT provided as part of the kit. A list of recommended dressings is provided in the user manual. The dressing must be applied and the myNeWT System set up by a Healthcare Professional skilled in wound care in accordance with the user manual.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the myNeWT Negative Pressure Wound Therapy System, structured according to your request:

    Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than a detailed report of a new clinical study with specific performance metrics against pre-defined acceptance criteria for a novel device. Therefore, much of the requested information regarding detailed acceptance criteria, specific performance metrics, sample sizes for test/training sets, expert adjudication methods, MRMC studies, standalone performance, and ground truth establishment in a clinical context is generally not present in this type of regulatory submission. The studies reported are primarily non-clinical (electrical, biocompatibility, mechanical, etc.) and aim to show the device meets design specifications and is safe and effective when compared to existing devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) summary focused on substantial equivalence rather than a de novo clinical trial, specific numerically defined "acceptance criteria" for a primary endpoint and "reported device performance" in a clinical human context are not explicitly stated in a tabular format. The document generally states that the device "passed all the verification and validation testing," indicating it met its internal design specifications and relevant standards.

    Here's an interpretation based on the non-clinical tests conducted, where the "acceptance criteria" are implied by compliance with standards and successful completion of tests:

    Acceptance Criteria Category (Implied)Reported Device Performance
    Electrical SafetyPassed AAMI ES 60601-1:2005/R 2012, IEC 60601-1-6: Ed 3.1 2013-10, ANSI/AAMI HA 60601-1-11:2011, and IEC 62366-1:2015.
    Electromagnetic CompatibilityPassed IEC 60601-1-2: Ed 4.0 2014-02 and IEC 60601-1-11: Ed 2.0 2015-01.
    Software Verification & ValidationConducted and documentation provided as recommended by FDA guidance.
    BiocompatibilityPassed ISO10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-10 (Primary skin Irritation).
    SterilityPassed Sterility Validation using Ethylene Oxide to ANSI/AAMI/ISO 11135:2014.
    Sterile PackagingPassed BS EN ISO 11607-1:2009 + A1:2014, including ISTA 2A distribution testing and accelerated/real-time aging.
    Mechanical PerformancePassed tests for compressor life, valve performance, battery life, pump noise, display visibility, canister strength, exudate containment, cleaning, and tube/connector mis-connection.
    Therapy PerformancePassed tests for vacuum levels (at pump and wound bed) and exudate removal performance.
    Stability/Shelf LifePassed stability testing to support shelf life claims.
    UsabilityDeveloped using a usability process conforming to IEC 62366-1:2015 and FDA Guidance.
    Distribution TestingPassed ISTA Procedure 2A (2011) for final packed kit.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable in the context of human clinical data for this 510(k) submission. The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence of the myNeWT System to the predicate devices."
    • Data Provenance: The reported "test" data primarily comes from bench testing of the device components and system to engineering specifications and international standards. This is not human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not provided as the "test set" primarily refers to non-clinical engineering and laboratory tests, not a clinical dataset requiring expert ground truth.

    4. Adjudication Method for the Test Set

    • Not applicable, as no clinical test set requiring adjudication by experts is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This information is not applicable. The device is a Negative Pressure Wound Therapy System, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical device (a pump system), not an AI algorithm. Its performance is inherent in its mechanical and software operation.

    7. The type of ground truth used

    • For the non-clinical tests, the "ground truth" is defined by established engineering specifications, international performance standards (e.g., ISO, IEC, AAMI), and documented test protocols. For example, for vacuum levels, the "ground truth" would be the target vacuum range as measured by calibrated equipment.

    8. The Sample Size for the Training Set

    • Not applicable, as this is not an AI/machine learning device requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as this is not an AI/machine learning device.