510(k) Data Aggregation

    K Number
    K160984

    Validate with FDA (Live)

    Date Cleared
    2016-07-20

    (104 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B5 S is a mobile digital X-ray diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning operations in trauma surgery, orthopedics, spine surgery, and chest surgery. It is not intended to be used in interventional procedures.

    The B5 S permits a qualified doctor or technologist to take a range of diagnostic exposures of spinal column, chest, abdomen, extremities, and other body parts on the patients.

    Device Description

    The B5 S is a mobile digital X-ray diagnostic system, which is intended to generate X-ray fluoroscopic image of a patient. The application includes: real-time positioning and monitoring operations in trauma surgery, orthopedics, spine surgery, and chest surgery. It is not intended to be used in interventional procedures.

    There are two sets of X-ray tube assemblies and Image Intensifiers which are perpendicularly distributed on the G-Arm, acting as two sets of vertical X-ray source and receptor systems and providing fluoroscopy image of the patient. The two sets of X-ray tube assemblies and Image Intensifiers can operate simultaneously and separately.

    The B5 S is comprised of control unit and G-Arm, the control unit and G-Arm include below primary components.

    Control Unit Major Component: Viewing monitor (2), Exposure lamp (2), Control monitor (1), Keyboard (1), Footswitch subassembly (1), Printer(optional) (1).

    G-Arm Major Component: Image intensifier assembly (2), X-Ray tube assembly (2), Laser light indicator(optional) (2), Steering unit (optional) (1).

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the MultiScan G-Arm System, which is an Image-intensified fluoroscopic x-ray system. The document focuses on demonstrating substantial equivalence to a predicate device and provides information on non-clinical testing for safety and performance standards. It does not contain information about acceptance criteria for a device's diagnostic performance, a study to prove it meets those criteria, or details regarding ground truth establishment, expert-based evaluation, or AI integration.

    Therefore, many of the requested items cannot be extracted from this document.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be extracted. This document reports on compliance with safety and performance standards relevant to an x-ray system (e.g., electrical safety, EMC, radiation safety), not on diagnostic performance metrics like sensitivity, specificity, or AUC which would typically be associated with acceptance criteria for a device that provides diagnostic interpretations.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be extracted. This document describes non-clinical testing against standards. There is no "test set" of patient data in the context of diagnostic performance. The tests performed are on the device itself (e.g., electrical measurements, radiation output).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Cannot be extracted. There is no "ground truth" establishment in the context of diagnostic imaging outcomes mentioned in this document, as it focuses on the engineering and safety performance of the X-ray system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Cannot be extracted. Not applicable as there is no diagnostic test set or expert evaluation described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Cannot be extracted. This document describes an X-ray fluoroscopic imaging system, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI performance metrics are included.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Cannot be extracted. Not applicable as this is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Cannot be extracted. Not applicable as this document pertains to the physical and electrical performance of an X-ray system, not diagnostic accuracy.

    8. The sample size for the training set

    • Cannot be extracted. Not applicable as this is not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Cannot be extracted. Not applicable as this is not an AI algorithm.

    Summary of available information from the document:

    The provided document details the non-clinical testing performed on the MultiScan G-Arm System (B5 S) to demonstrate its safety and effectiveness for substantial equivalence to a predicate device (K151280).

    Acceptance Criteria and Reported Device Performance (Non-clinical Standards Compliance):

    The device was tested for compliance with several accepted safety and performance standards. The "acceptance criteria" in this context are the requirements of these standards, and the "reported device performance" is that the device "Pass"ed all these tests.

    Testing Item CategoryStandard ReferenceResult
    Electrical Safety & Mechanical StabilityAAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012Pass (for all 33 listed items, including power input, leakage currents, dielectric strength, instability tests, mechanical strength, etc.)
    Electromagnetic Compatibility (EMC)IEC 60601-1-2 Edition 3:2007-03Pass (for all 7 listed items, including mains terminal continuous disturbance voltage, radiated emission, electrostatic discharge, RF electromagnetic field immunity, fast transients, injected current, surges, variations of power frequency)
    Radiation Protection in Diagnostic X-rayIEC 60601-1-3 Edition 2.1 2013Pass (for all 3 listed items: Half-value layer, Leakage radiation in the loading state, Leakage radiation when not in the loading state)
    Specific Requirements for X-ray EquipmentIEC 60601-2-54 Edition 1.0 2009-06Pass (for all 4 listed items: Test for dosimetric information, Accuracy of X-ray tube voltage, Accuracy of X-ray tube current, Determining the attenuation of residual radiation)
    Digital Imaging CommunicationsDICOMMeets the provisions

    Study Description (Non-clinical Testing for Substantial Equivalence):

    The study was a non-clinical testing process designed to show conformance to recognized safety and performance standards. This is not a clinical study involving patient data for diagnostic accuracy.

    • Sample Size for Test Set / Data Provenance: Not applicable. The "tests" were conducted on the device hardware itself against predefined engineering and safety specifications, not on a set of patient cases.
    • Number of Experts / Qualifications / Adjudication: Not applicable. Evaluation was against technical standards, not diagnostic interpretation by human experts.
    • MRMC / Standalone / AI: Not applicable. The device is a traditional X-ray fluoroscopic system, not an AI-driven diagnostic tool.
    • Ground Truth Type: Not applicable in the diagnostic sense. The "ground truth" for the tests performed is the adherence to the specified technical parameters and safety limits defined by the IEC/AAMI standards.
    • Training Set Size / Ground Truth for Training Set: Not applicable. The device is a hardware system, not a machine learning model.