Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160643

    Validate with FDA (Live)

    Date Cleared
    2016-10-18

    (225 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENROUTE 0.014" Guidewire is intended for use in the peripheral vasculature.

    Device Description

    The ENROUTE .014” guidewire is a disposable medical device designed for single use only. It consists of a 95cm PTFE coated 0.014" diameter stainless steel core wire, one end of which is reduced in diameter over approximately 9.5cm in a progressive fashion through a centreless grinding operation. The profile of this reduced section affords the product an area of reduced stiffness. The very distal tip section is flattened to further reduce stiffness and enable the tip to be shaped.

    The distal section is covered with a 5cm platinum tungsten spring coil. This provides for greater visibility on x-ray equipment (radiopacity). A hydrophilic coating is applied to the distal section to enhance lubricity. The product is available in straight configuration

    AI/ML Overview

    This document describes the 510(k) premarket notification for the ENROUTE 0.014" Guidewire, which is intended for use in the peripheral vasculature. The device demonstrates substantial equivalence to predicate devices (Brivant Guidewire K060551 and Nitrex Nitinol Guidewire K040345) based on technological characteristics, performance testing, and intended use.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical "acceptance criteria" in a table format for each test. Instead, it states that "The performance testing assessment supports that the biocompatibility, shelf life, and functional specifications of the proposed ENROUTE 0.014" Guidewire device were met. The ENROUTE 0.014" Guidewire device test data supports the claims of substantial equivalence to the predicate devices." This implies that the device's performance results were within acceptable ranges to demonstrate substantial equivalence to the predicate devices. The types of tests performed are listed, and for some, the new device had specific data generated ("X") or leveraged data from a predicate ("X").

    TestAcceptance Criteria (Implied: Met Predicate Performance/Specifications)Reported Device Performance (Implied: Met Criteria)
    Tensile StrengthEquivalent to predicate deviceMet
    Torque StrengthEquivalent to predicate deviceMet
    Dimensional measurements (Overall length)Within range of predicate deviceMet
    Torque ResponseEquivalent to predicate deviceMet
    Catheter CompatibilityCompatible with equivalent interventional devicesMet
    Coating Adherence/Coating IntegrityEquivalent to predicate deviceMet
    Particulate TestingEquivalent to predicate deviceMet
    Tip Stiffness/FlexibilityEquivalent to predicate deviceMet
    RadiopacityVisible on x-ray equipment (radiopaque)Met
    BiocompatibilityIn compliance with ISO 10993-1:2009 and G95-1Met
    Shelf LifeEstablishedMet

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "in vitro bench tests" but does not specify the sample sizes used for each test. The data provenance (country of origin, retrospective/prospective) is not explicitly stated, but the submission is from "Lake Region Medical" in "Galway, Ireland."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is a medical device submission, not an AI/software claim. Therefore, the concept of "ground truth established by experts" as typically understood in AI studies (e.g., radiologists labeling images) does not apply here. The "acceptance criteria" and "performance" are based on engineering specifications and bench testing, not expert consensus on diagnostic interpretations.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable as this is a medical device bench testing study, not an expert-driven AI study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable as this is a medical device bench testing study, not an AI comparative effectiveness study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is based on engineering specifications, FDA guidance documents (e.g., "Coronary and Cerebrovascular Guidewire Guidance, Jan 1995"), and established performance characteristics of the predicate devices. For biocompatibility, the ground truth is adherence to internationally recognized standards like ISO 10993-1:2009.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device, not an AI model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1