Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160557

    Validate with FDA (Live)

    Date Cleared
    2016-07-06

    (128 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over the counter device is for adults aged 18 years and older with wrist circumference ranging from 5.3 inches to 7.7 inches (approx. 135 mm to 195 mm) and for home use.

    HLI S8VA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth-enabled device. Also, users could simply synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    Device Description

    HL158VA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 5.3 inches to 7.7 inches (approx.13.5 cm to 19.5 cm) and for home use.

    The device will display a symbol wor or to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.

    Besides, HL158VA has a built-in Wrist Position Guide. By holding START/STOP button, the display will illuminate with different icons that are designed to guide you move your wrist. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data. Also synchronize the current date and time, and check the battery status of blood pressure monitor by means of DailyChek® application software with the paired Bluetooth-enabled device.

    AI/ML Overview

    The document discusses the Food and Drug Administration's (FDA) review of the Full Automatic (NIBP) Blood Pressure Monitor, Model HL158VA. It determined the device to be substantially equivalent to previously marketed predicate devices.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The primary clinical performance acceptance criteria for the blood pressure monitor are related to its accuracy, as per ISO 81060-2.

    Acceptance Criteria (from ISO 81060-2:2013)Reported Device Performance (HL158VA)
    Mean difference limits for blood pressure measurementsFulfilled (results of clinical investigation)
    Standard deviation limits for blood pressure measurementsFulfilled (results of clinical investigation)
    Pressure Accuracy: ± 3mmHgAchieved
    Pulse Accuracy: ± 5%Achieved

    Additional Technical Specifications:

    FeaturePredicate Device (HL158BT, K093824)Subject Device (HL158VA)Acceptance/Performance
    Method of MeasurementOscillometricOscillometricSame as predicate, accepted
    Measurement TypeDuring deflationDuring inflationModified, but validated clinically
    Range of measurement (Pressure)0~300 mmHgRated Range of Cuff Pressure: 0300mmHg, Rated Range of Determination: 40280mmHgAccepted
    Range of measurement (Pulse)40~199 Beats/minute40~199 Beats/minuteSame as predicate, accepted
    InflationAutomaticAutomaticSame as predicate, accepted
    DeflationAutomatic (Rapid Exhaust Valve)Automatic (Passive Exhaust Valve)Modified, but presumably validated
    Pressure Changed Rate2-5 mmHg/sec. (from 150mmHg to 90mmHg)2-5 mmHg/sec. (from 90mmHg to 150mmHg)Modified, but acceptable.
    DisplayLiquid Crystal DigitalLiquid Crystal DigitalSame as predicate, accepted
    Memory2*60, total 1202*120, total 240Improved, accepted
    Number of Push Button7 keys6 keysModified, accepted
    Cuff size (Wrist circumference)135 ~ 195 mm (5.3~7.7 inches)135 ~ 195 mm (5.3~7.7 inches)Same as predicate, accepted
    Risk Category IndicatorYesYesSame as predicate, accepted
    Irregular Heartbeat DetectorYesYesSame as predicate, accepted
    Data Link FunctionYes (Via USB cable)Yes (Via Bluetooth)Modified, validated non-clinically
    Talking FunctionYesNoRemoved, accepted
    Wrist Position GuideNoYesAdded, validated

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: A "group of 100 subjects" was used for the clinical investigation.
    • Data Provenance: The document does not specify the country of origin. The study was a prospective clinical investigation, designed to evaluate the device against a known standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide details on the number or qualifications of experts used to establish the "ground truth" (reference blood pressure measurements). In ISO 81060-2 clinical validation, this typically involves trained observers taking auscultatory measurements using a mercury sphygmomanometer in a standardized manner.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not explicitly describe an adjudication method for the test set. For ISO 81060-2 clinical validation, measurements from multiple trained observers are often compared to minimize individual observer bias, but a formal "adjudication" in the sense of resolving discrepancies is not typically described in summary documents like this.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is a standalone blood pressure monitor and does not involve human readers analyzing medical images with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was conducted. The clinical investigation directly evaluated the accuracy of the device (HL158VA) against the ISO 81060-2 standard, which assesses the device's ability to measure blood pressure without human intervention influencing the measurement itself (though a human operates the device).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical validation was based on reference blood pressure measurements taken by trained observers, as specified by the requirements of the ISO 81060-2 standard for non-invasive sphygmomanometers. This is a form of "expert measurement" or "reference method."

    8. The sample size for the training set

    The document does not mention a training set sample size. For medical devices like this, which measure physiological parameters using established principles (oscillometric method), a "training set" in the machine learning sense is not typically used or reported for regulatory submissions; rather, the device's algorithms are developed and then validated through clinical studies.

    9. How the ground truth for the training set was established

    As no training set (in the machine learning sense) is mentioned, there's no information on how its ground truth was established. The device's underlying algorithm is based on the well-established oscillometric method, and its accuracy is validated directly on a test set (the 100 subjects) rather than through a separate training/validation split typical of AI systems.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1