510(k) Data Aggregation

    K Number
    K160125

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-04-22

    (93 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CASCADIA lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy. Additionally, the CASCADIA lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. CASCADIA lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine.

    The CASCADIA cervical implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiguous levels from C2 to T1. These patients should be skeletally mature and have had six weeks of non-operative treatment. The CASCADIA cervical implants are also to be used with supplemental fixation; the hyperlordotic CASCADIA cervical implants (i.e., ≥ 10°) are required to be used with an anterior cervical plate as the form of supplemental fixation.

    Device Description

    The subject Cascadia Interbody System is comprised of additively manufactured implants made of titanium allov (ASTM standards F3001 and F136). These devices are available in a variety of different sizes and heights to match more closely the patient's anatomy. The purpose of the subject submission is to add cervical implants to the system and to expand the indications for use of the CASCADIA lumbar implants to include use as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    Function: The system functions as an intervertebral body fusion device to provide support and stabilization of the cervical segments of the spine.

    AI/ML Overview

    This document describes the Cascadia Interbody System, specifically focusing on its regulatory approval via a 510(k) submission (K160125). A 510(k) submission demonstrates substantial equivalence to legally marketed predicate devices, rather than proving that a device meets specific, pre-defined acceptance criteria through a study designed explicitly for that purpose with detailed performance metrics.

    Therefore, the information requested about acceptance criteria and a study proving those criteria are met for an AI/ML device is not directly applicable to this submission for a medical implant. This document does not describe an AI/ML device or a study with performance metrics in terms of sensitivity, specificity, or reader improvement.

    However, I can extract information related to the performance evaluation and comparative analysis that was performed for this device to gain FDA clearance.

    Here's a breakdown of the relevant information from the provided text, structured as closely as possible to your request for clarity, but with the understanding that it's for a medical implant, not an AI/ML diagnostic system:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission for a medical implant, there aren't "acceptance criteria" in the sense of specific clinical performance metrics like sensitivity/specificity for an AI/ML algorithm. Instead, the "performance evaluation" described is non-clinical and aims to demonstrate the mechanical and structural equivalence to predicate devices. The "acceptance criteria" can be inferred as successfully passing these non-clinical tests, showing substantial equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    Withstand Static Torsion (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Withstand Static Compression (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Withstand Dynamic Compression (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Withstand Dynamic Torsion (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Withstand Static Compression Shear (ASTM F2077) (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Withstand Dynamic Compression Shear (ASTM F2077) (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Resist Subsidence (ASTM F2267) (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Resist Expulsion (Worst Case Components)Performance evaluations conducted. Proposed implants were found to be substantially the same as predicate devices.
    Overall Conclusion"There are no significant differences between the Cascadia Interbody System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use." (This is the ultimate "acceptance" for a 510(k) submission).

    2. Sample Size for the Test Set and Data Provenance

    This information is not applicable. The "test set" in this context refers to the samples used for non-clinical performance evaluations (mechanical testing), not a clinical trial with human data. The document does not specify the exact number of components or test specimens used for the evaluations, nor does it refer to data provenance in terms of country of origin or retrospective/prospective acquisition, as would be relevant for clinical data.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. Ground truth, in the context of AI/ML, refers to clinically verified labels or outcomes. For this medical implant, "ground truth" is established by engineering standards and mechanical test results, not by expert clinical review. The document does not mention the involvement of experts for establishing "ground truth" in the AI/ML sense.


    4. Adjudication Method for the Test Set

    This information is not applicable for a medical implant's non-clinical performance evaluation. Adjudication methods (e.g., 2+1, 3+1) are used to resolve discrepancies among human readers or experts establishing ground truth for clinical data, which is not the type of evaluation described.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study is relevant for evaluating the clinical utility of diagnostic devices, particularly AI/ML algorithms, by comparing human reader performance with and without AI assistance. The Cascadia Interbody System is an implant, not a diagnostic tool or an AI/ML system.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This refers to the performance of an AI/ML algorithm in isolation. The Cascadia Interbody System is a physical medical implant, not an algorithm.


    7. The Type of Ground Truth Used

    For this medical implant, the "ground truth" for the non-clinical performance evaluations is engineering standards and mechanical test results (e.g., ASTM F2077, ASTM F2267). This demonstrates that the device meets intended mechanical properties and behaves similarly to predicate devices. It is not expert consensus, pathology, or outcomes data in the clinical diagnostic sense.


    8. The Sample Size for the Training Set

    This information is not applicable. "Training set" refers to data used to train an AI/ML algorithm. The Cascadia Interbody System is a physical implant and does not involve an AI/ML algorithm requiring a training set.


    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reasons as #8.