Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160004

    Validate with FDA (Live)

    Device Name
    Crosstella OTW
    Manufacturer
    Date Cleared
    2016-04-29

    (116 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Crosstella OTW PTA Balloon Dilatation Catheter is intended to dilate stenoral, ilio-femoral, popliteal, infrapoplited and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

    Device Description

    Crosstella OTW PTA Balloon Dilatation Catheter (Crosstella OTW) is an over-the-wire type of balloon dilation catheter, which consists of an outer tube, inner tube, balloon, radiopaque markers, manifold, outer reinforcement tube and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the manifold at the proximal end. Crosstella OTW has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the inner tube to the guidewire port of the manifold for the introduction of a guidewire.

    The maximum compatible diameter of a guidewire used together with Crosstella OTW in a PTA procedure is 0.018 inches. Additionally, the compatible size of a sheath introducer is 4 The nominal inflated balloon diameters range from 2.0 mm to 6.0 mm with balloon or 5 Fr. working lengths of 20 mm to 200 mm. Two different catheter working lengths will be provided: 900 and 1500 mm.

    AI/ML Overview

    Here is an analysis of the provided FDA 510(k) summary for the Crosstella OTW PTA Balloon Dilatation Catheter, focusing on acceptance criteria and study details.

    Please note: This document describes a medical device (a balloon dilatation catheter) and its clearance based on substantial equivalence to a predicate device. The nature of the studies described are non-clinical bench and biocompatibility tests, not clinical trials involving human subjects to assess diagnostic accuracy or treatment efficacy in the way an AI/ML algorithm might be studied. Therefore, many of the requested points regarding AI/ML specific study design (like ground truth, expert adjudication, MRMC studies, training set details) are not applicable to this type of device submission.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Crosstella OTW PTA Balloon Dilatation Catheter were based on bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters' dated September 8, 2010," and biocompatibility testing in accordance with "Draft Guidance for Industry and Food and Drug Administration Staff; Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"" (dated April 23, 2013).

    The document states, "The Crosstella OTW met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs."

    Below is a table summarizing the types of tests performed and the implicit reported performance against the acceptance criteria. Specific numerical values for the acceptance criteria and performance are not provided in this summary document, but the overall conclusion is that the device met these criteria.

    Acceptance Criteria CategorySpecific Tests PerformedReported Device Performance
    Performance TestingDimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Radiopacity, Coating Integrity, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), Transportation and Shelf Life Testing."The results from these tests demonstrate that the technological characteristics and performance criteria of the Crosstella OTW are adequate for the intended use of the device and that the device can perform in a manner equivalent to devices currently on the market with the same intended use." This indicates that the device met the performance specifications derived from the FDA guidance and applicable standards for each test listed.
    Biocompatibility TestingCytotoxicity, Sensitization, Intracutaneous reactivity (irritation), Systemic toxicity (acute), Pyrogenicity, Hemocompatibility (thrombogenicity, hemolysis, and immunology), and Genotoxicity (bacterial gene mutation assay, in vitro mammalian genotoxicity assay, and in vivo cytogenetics assay)."The results from these tests demonstrate that the Crosstella OTW is biocompatible for its intended use similar to the predicate device." This means all biocompatibility tests passed and showed the device materials are safe for patient contact.
    Overall Conclusion"The Crosstella OTW met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." This is the ultimate statement of meeting the acceptance criteria.

    Detailed Study Information (Specific to a Medical Device requiring Bench/Biocompatibility Testing)

    Given that this is a 510(k) submission for a physical medical device (a catheter) and not an AI/ML algorithm, many of the requested points are not directly applicable. However, I will address them in the context of this device type where relevant, or state "Not Applicable" (N/A) if they pertain exclusively to AI/ML software.

    1. Sample size used for the test set and the data provenance:

      • Sample Size (Test Set): The specific number of individual catheters or material samples tested for each benchmark test is not explicitly stated in this summary. However, device manufacturers typically test multiple units/lots to ensure reproducibility and statistical validity according to industry standards and FDA guidance.
      • Data Provenance: The tests would have been performed in a laboratory setting by Kaneka Corporation or a contracted testing facility. This is therefore prospective testing of newly manufactured devices under controlled conditions, not retrospective analysis of patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A. For bench and biocompatibility testing of a physical device, "ground truth" is established by direct measurement of physical properties against engineering specifications or by observing biological reactions in controlled in vitro or in vivo (animal) models. It does not involve expert interpretation of medical images or data from human patients. The "experts" would be engineers, chemists, microbiologists, etc., performing the tests and analyzing the results against predefined acceptance criteria.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations of clinical data, typically in diagnostic studies. For device performance testing, results are quantitative or qualitative assessments against objective criteria. Discrepancies would be resolved through re-testing, investigation of methodology, or review by senior technical personnel, not clinical adjudication.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A. This type of study is specific to AI/ML diagnostic tools. This submission is for a physical intervention device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A. This is a physical device, not an algorithm. Bench tests are "standalone" in the sense that they assess the device's inherent properties and performance without human clinical interaction beyond the operation of test equipment.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A in the clinical sense. For this device, the "ground truth" for performance is defined by:

      • Established engineering specifications and industry standards: E.g., a balloon's rated burst pressure is compared against a specified pressure.
      • Biological responses in controlled environments: E.g., cell viability in cytotoxicity tests, or lack of adverse reaction in in vivo (animal, if applicable) biocompatibility tests.
      • Comparison to predicate device characteristics: Demonstrating that the new device performs "similarly" or "equivalently" to the predicate device across various parameters.
    7. The sample size for the training set: N/A. There is no "training set" as this is not an AI/ML algorithm. Device design and manufacturing processes are refined through iterative development, but this is distinct from machine learning training.

    8. How the ground truth for the training set was established: N/A. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1